Timing and Energy Resolution of New Near-UV SiPMs Coupled to LaBr3:Ce for TOF-PET

被引:12
作者
Schmall, Jeffrey P. [1 ]
Wiener, Rony I. [1 ]
Surti, Suleman [1 ]
Ferri, Alessandro [2 ]
Gola, Alberto [2 ]
Tarolli, Alessandro [2 ]
Piemonte, Claudio [2 ]
Karp, Joel S. [1 ]
机构
[1] Univ Penn, Philadelphia, PA 19104 USA
[2] Fdn Bruno Kessler, I-38122 Trento, Italy
关键词
Coincidence timing resolution; Lanthanum Bromide; PET; silicon photomultiplier; time-of-flight PET; TIME-OF-FLIGHT; PERFORMANCE; IMPACT; DETECTOR;
D O I
10.1109/TNS.2014.2346579
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The high light output and fast decay time of LaBr3:Ce scintillation detectors leads to excellent timing performance. To realize the potential of timing resolution with LaBr3:Ce we have investigated the performance with SiPMs, which enable 1-to-1 coupling to individual crystals, and which have been optimized for the near-ultraviolet (NUV) scintillation light emission of LaBr3:Ce. Coincidence timing resolution (CTR) of 100 ps was measured for a 4 x 4 x 5 mm(3) LaBr3:30%Ce crystal directly coupled to a large-area 4 x 4 mm(2) NUV-SiPM. Results show very little dependence on temperature, in the range of to -20 degrees to 20 degrees C and bias voltage, from 2 V to 5 V over break-down. Optimal performance was achieved at an over-voltage (OV) range of 3 V - 5 V, at which high gain and high photon detection efficiency are achieved. Though saturation was evident at 511 keV, an energy resolution of 6.8%was measured after correcting for non-linearity. We also measured a CTR of 110 ps for a 4 x 4 x 5 mm(3) LaBr3:5%Ce crystal and 245 ps for a 4 x 4 x 30 mm(3) LaBr3:5%Ce crystal using the NUV-SIPM. The poorer timing measurement for the 30-mm long crystal is due mainly to a systematic shift in the time pick-off as a function of the depth-of-interaction. The excellent temperature stability, fast rise time, high gain, and low noise of the NUV-SiPM make it a practical and highly appealing photodetector for the readout of a LaBr3:Ce TOF-PET detector.
引用
收藏
页码:2426 / 2432
页数:7
相关论文
共 28 条
[1]   Physical Performance of the new hybrid PET/CT Discovery-690 [J].
Bettinardi, V. ;
Presotto, L. ;
Rapisarda, E. ;
Picchio, M. ;
Gianolli, L. ;
Gilardi, M. C. .
MEDICAL PHYSICS, 2011, 38 (10) :5394-5411
[2]  
BUDINGER TF, 1983, J NUCL MED, V24, P73
[3]   The imaging performance of a LaBr3-based PET scanner [J].
Daube-Witherspoon, M. E. ;
Surti, S. ;
Perkins, A. ;
Kyba, C. C. M. ;
Wiener, R. ;
Werner, M. E. ;
Kulp, R. ;
Karp, J. S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (01) :45-64
[4]  
Fakhri G. E., 2011, J. Nuclear Med., V52, P347
[5]   Performance of FBK high-density SiPM technology coupled to Ce:LYSO and Ce:GAGG for TOF-PET [J].
Ferri, Alessandro ;
Gola, Alberto ;
Serra, Nicola ;
Tarolli, Alessandro ;
Zorzi, Nicola ;
Piemonte, Claudio .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (04) :869-880
[6]   Analog Circuit for Timing Measurements With Large Area SiPMs Coupled to LYSO Crystals [J].
Gola, Alberto ;
Piemonte, Claudio ;
Tarolli, Alessandro .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (02) :1296-1302
[7]   The DLED Algorithm for Timing Measurements on Large Area SiPMs Coupled to Scintillators [J].
Gola, Alberto ;
Piemonte, Claudio ;
Tarolli, Alessandro .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (02) :358-365
[8]   Physical and clinical performance of the mCT time-of-flight PET/CT scanner [J].
Jakoby, B. W. ;
Bercier, Y. ;
Conti, M. ;
Casey, M. E. ;
Bendriem, B. ;
Townsend, D. W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (08) :2375-2389
[9]   Impact of Time-of-Flight on PET Tumor Detection [J].
Kadrmas, Dan J. ;
Casey, Michael E. ;
Conti, Maurizio ;
Jakoby, Bjoern W. ;
Lois, Cristina ;
Townsend, David W. .
JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (08) :1315-1323
[10]   Benefit of time-of-flight in PET: Experimental and clinical results [J].
Karp, Joel S. ;
Surti, Sulernan ;
Daube-Witherspoon, Margaret E. ;
Muehllehner, Gerd .
JOURNAL OF NUCLEAR MEDICINE, 2008, 49 (03) :462-470