Dual-camera DPIV for flow studies past artificial heart valves

被引:19
|
作者
Brucker, C
机构
[1] Aerodynamic Institute, RWTH Aachen, D-52062 Aachen
关键词
D O I
10.1007/s003480050077
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
A video-based digital-particle-image-velocimetry (DPIV) system with a time-delayed dual-camera recording was built for investigation of the unsteady flow structure in a pulsating flow behind artificial heart valves. The delay between the master and slave image is arbitrary by adjusting a phase-shift in the phase synchronization of the consumer video-cameras which represents a cheap and easy way to perform video-based cross-correlation DPIV at high velocities like that in heart valve flows. A calibration procedure was necessary to correct for misalignment of the image acquisition system with regard to translation and tilt of both image planes. With a framing rate of 50 Hz and continuous recording, a large number of PIV recordings at a rate of 50 Hz can be captured which enables to resolve the flow evolution in detail as demonstrated by a sequence showing the generation and sweep down of starting vortices behind the leaflets for the Bjork-Shiley-Monostrut (BSM) and Sorin-Bicarbon (SE) heart valves during the opening phase. Long-term recording with the video system over several hundreds of beating cycles offers statistical investigation of cycle-resolved fluctuations and evaluation of shear stresses. In general, this system is very suitable for video-based DPIV at higher velocities when a pulsed and synchronized illumination is not available or difficult to implement.
引用
收藏
页码:496 / 506
页数:11
相关论文
共 44 条
  • [1] Dual-camera DPIV for flow studies past artificial heart valves
    Ch. Brücker
    Experiments in Fluids, 1997, 22 : 496 - 506
  • [2] PARTICLE IMAGE VELOCIMETRY IN THE INVESTIGATION OF FLOW PAST ARTIFICIAL-HEART VALVES
    LIM, WL
    CHEW, YT
    CHEW, TC
    LOW, HT
    ANNALS OF BIOMEDICAL ENGINEERING, 1994, 22 (03) : 307 - 318
  • [3] FLOW DYNAMICS IN ARTIFICIAL HEART VALVES
    EFFERT, S
    ZEITSCHRIFT FUR KREISLAUFFORSCHUNG, 1970, 59 (07): : 648 - &
  • [4] DUAL-CAMERA STUDIES OF PULMONARY-FUNCTION WITH COMPUTER-PROCESSING OF DATA
    LOKEN, MK
    PAYNE, JT
    PONTO, RA
    KRONENBERG, R
    JOURNAL OF NUCLEAR MEDICINE, 1973, 14 (06) : 422 - 423
  • [5] STUDIES ON PSYCHOPATHOLOGY OF PATIENTS WITH ARTIFICIAL HEART VALVES
    SPEIDEL, H
    KALMAR, P
    KEREKJAR.MV
    KLINISCHE WOCHENSCHRIFT, 1970, 48 (20): : 1251 - &
  • [6] STUDIES ON PSYCHOPATHOLOGY OF PATIENTS IN ARTIFICIAL HEART VALVES
    SPEIDEL, H
    KALMAR, P
    KEREKJARTO, MV
    RAMB, W
    ZEITSCHRIFT FUR KREISLAUFFORSCHUNG, 1970, 59 (07): : 652 - +
  • [7] Artificial heart valves, flow acceleration and shear stresses
    Sakhaeimanesh, Ali A.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 13, 2009, 25 (13): : 170 - 173
  • [8] A flow chamber to study the performance of artificial heart valves
    Yin, Wei
    Ngwe, Ching
    Rubenstein, David A.
    FASEB JOURNAL, 2010, 24
  • [9] PULSATILE FLOW STUDIES OF PROSTHETIC HEART VALVES
    DAVEY, TB
    KAUFMAN, B
    SMELOFF, EA
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1966, 51 (02): : 264 - &
  • [10] Platelet activation and free emboli formation in flow past mechanical heart valves - numerical and in vivo studies
    Bluestein, D
    Yin, W
    Alemu, Y
    Jesty, J
    Saltman, AE
    Krukenkamp, IB
    Bernal, O
    Suresh, K
    SECOND JOINT EMBS-BMES CONFERENCE 2002, VOLS 1-3, CONFERENCE PROCEEDINGS: BIOENGINEERING - INTEGRATIVE METHODOLOGIES, NEW TECHNOLOGIES, 2002, : 1250 - 1251