Technical Note: The first live treatment on a 1.0 Tesla inline MRI-linac

被引:13
作者
Liney, Gary P. [1 ,2 ,3 ,4 ]
Jelen, Urszula [1 ]
Byrne, Hilary [5 ,6 ]
Dong, Bin [1 ]
Roberts, Tara L. [1 ,7 ]
Kuncic, Zdenka [5 ,6 ]
Keall, Paul [6 ]
机构
[1] Ingham Inst Appl Med Res, Dept Med Phys, Liverpool, NSW, Australia
[2] Liverpool Canc Therapy Ctr, Radiat Phys, Liverpool, NSW, Australia
[3] Univ New South Wales, Sch Med, Sydney, NSW, Australia
[4] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW, Australia
[5] Univ Sydney, Sch Phys, Fac Sci, Sydney, NSW, Australia
[6] Univ Sydney, ACRF ImageX Inst, Sydney Med Sch, Sydney, NSW, Australia
[7] Univ Western Sydney, Sch Med, Macarthur, NSW, Australia
基金
英国医学研究理事会;
关键词
MRI; MRI-Linac; radiotherapy; IMAGES; RADIOTHERAPY; SYSTEM;
D O I
10.1002/mp.13556
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose This work describes the first live imaging and radiation delivery performed on a prototype 1.0 T inline MRI-Linac system in a rat brain tumor model, which was conducted on 29 January 2019. Methods A human scale 1.0 T MRI-Linac was adapted to be suitable for animal studies via a specially constructed open 6-channel receiver radiofrequency (RF) coil. A Fischer rat injected with 9L glioma cells in the right hemisphere was imaged and irradiated at day 11 post surgery as part of a larger cohort survival study. The rat was anesthetized and positioned at the iscocenter of the MRI-Linac. Imaging was used to localize the brain and confirm the presence of a tumor following the administration of a gadolinium nanoparticle contrast agent. A single dose of 10 Gy was delivered using a 2.25 cm x 2.90 cm radiation field covering the whole brain and verified with radiosensitive film in situ. Real-time imaging was used throughout the irradiation period to monitor the animal and target position. Results The signal-to-noise ratio (SNR) measured in the rat brain was 38. Postcontrast imaging was able to demonstrate a tumor of 5 mm diameter in the upper right hemisphere of the brain approximately 45 min after administration of the nanoparticles. The radiation beam had no impact on SNR and images at the rate of 2 Hz were effective in monitoring both respiration and intrafractional motion. In vivo film dosimetry confirmed the intended dose delivery. The total procedure time was 35 min. Conclusions We have successfully used MRI guidance to localize and subsequently deliver a radiation field to the whole brain of a rat with a right hemispheric tumor. Real-time imaging during beam on was of sufficient quality to monitor breathing and perform exception gating of the treatment. This represents the first live use of a high field inline MRI-Linac.
引用
收藏
页码:3254 / 3258
页数:5
相关论文
共 15 条
[1]  
Billas I, 2018, 46 NPL IR
[2]   Effect of radiation induced current on the quality of MR images in an integrated linac-MR system [J].
Burke, Ben ;
Wachowicz, K. ;
Fallone, B. G. ;
Rathee, Satyapal .
MEDICAL PHYSICS, 2012, 39 (10) :6139-6147
[3]   Signal-to-noise measurements in magnitude images from NMR phased arrays [J].
Constantinides, CD ;
Atalar, E ;
McVeigh, ER .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (05) :852-857
[4]   First MR images obtained during megavoltage photon irradiation from a prototype integrated linac-MR system [J].
Fallone, B. G. ;
Murray, B. ;
Rathee, S. ;
Stanescu, T. ;
Steciw, S. ;
Vidakovic, S. ;
Blosser, E. ;
Tymofichuk, D. .
MEDICAL PHYSICS, 2009, 36 (06) :2084-2088
[5]   Characterization of the first RF coil dedicated to 1.5 T MR guided radiotherapy [J].
Hoogcarspel, Stan J. ;
Zijlema, Stefan E. ;
Tijssen, Rob H. N. ;
Kerkmeijer, Linda G. W. ;
Jurgenliemk-Schulz, Ina M. ;
Lagendijk, Jan J. W. ;
Raaymakers, Bas W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (02)
[6]   The Australian Magnetic Resonance Imaging-Linac Program [J].
Keall, Paul J. ;
Barton, Michael ;
Crozier, Stuart .
SEMINARS IN RADIATION ONCOLOGY, 2014, 24 (03) :203-206
[7]   Imaging performance of a dedicated radiation transparent RF coil on a 1.0 Tesla inline MRI-linac [J].
Liney, G. P. ;
Dong, B. ;
Weber, E. ;
Rai, R. ;
Destruel, A. ;
Garcia-Alvarez, R. ;
Manton, D. J. ;
Jelen, U. ;
Zhang, K. ;
Barton, M. ;
Keall, P. ;
Crozier, S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (13)
[8]   Technical Note: Experimental results from a prototype high-field inline MRI-linac [J].
Liney, G. P. ;
Dong, B. ;
Begg, J. ;
Vial, P. ;
Zhang, K. ;
Lee, F. ;
Walker, A. ;
Rai, R. ;
Causer, T. ;
Alnaghy, S. J. ;
Oborn, B. M. ;
Holloway, L. ;
Metcalfe, P. ;
Barton, M. ;
Crozier, S. ;
Keall, P. .
MEDICAL PHYSICS, 2016, 43 (09) :5188-5194
[9]   The ViewRay System: Magnetic Resonance-Guided and Controlled Radiotherapy [J].
Mutic, Sasa ;
Dempsey, James F. .
SEMINARS IN RADIATION ONCOLOGY, 2014, 24 (03) :196-199
[10]   Dose enhancement in radiotherapy of small lung tumors using inline magnetic fields: A Monte Carlo based planning study [J].
Oborn, B. M. ;
Ge, Y. ;
Hardcastle, N. ;
Metcalfe, P. E. ;
Keall, P. J. .
MEDICAL PHYSICS, 2016, 43 (01) :368-377