Optimization of a fully 3D single scatter simulation algorithm for 3D PET

被引:100
|
作者
Accorsi, R
Adam, LE
Werner, ME
Karp, JS
机构
[1] Childrens Hosp Philadelphia, Dept Radiol, Div Nucl Med, Philadelphia, PA 19104 USA
[2] Univ Penn, Sch Med, Dept Radiol, Philadelphia, PA 19104 USA
[3] Philips Med Syst, Philadelphia, PA 19104 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2004年 / 49卷 / 12期
关键词
D O I
10.1088/0031-9155/49/12/008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We describe a new implementation of a single scatter simulation (SSS) algorithm for the prediction and correction of scatter in 3D PET. In this implementation, out of field of view (FoV) scatter and activity, side shields and oblique tilts are explicitly modelled. Comparison of SSS predictions with Monte Carlo simulations and experimental data from uniform, line and cold-bar phantoms showed that the code is accurate for uniform as well as asymmetric objects and can model different energy resolution crystals and low level discriminator (LLD) settings. Absolute quantitation studies show that for most applications, the code provides a better scatter estimate than the tail-fitting scatter correction method currently in use at our institution. Several parameters such as the density of scatter points, the number of scatter distribution sampling points and the axial extent of the FoV were optimized to minimize execution time, with particular emphasis on patient studies. Development and optimization were carried out in the case of GSO-based scanners, which enjoy relatively good energy resolution. SSS estimates for scanners with lower energy resolution may result in different agreement, especially because of a higher fraction of multiple scatter events. The algorithm was applied to a brain phantom as well as to clinical whole-body studies. It proved robust in the case of large patients, where the scatter fraction increases. The execution time, inclusive of interpolation, is typically under 5 min for a whole-body study (axial FoV: 81 cm) of a 100 kg patient.
引用
收藏
页码:2577 / 2598
页数:22
相关论文
共 50 条
  • [1] A new implementation of the single scatter simulation algorithm for 3D PET.
    Adam, LE
    Accorsi, R
    Charron, M
    Karp, JS
    JOURNAL OF NUCLEAR MEDICINE, 2002, 43 (05) : 57P - 57P
  • [2] A single scatter simulation technique for scatter correction in 3D pet
    Watson, CC
    Newport, D
    Casey, ME
    THREE-DIMENSIONAL IMAGE RECONSTRUCTION IN RADIOLOGY AND NUCLEAR MEDICINE, 1996, 4 : 255 - 268
  • [3] 3D implementation of Scatter Estimation in 3D PET
    Iatrou, M.
    Manjeshwar, R. M.
    Ross, S. G.
    Thielemans, K.
    Stearns, C. W.
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2142 - 2145
  • [4] DETECTOR EFFICIENCY AND COMPTON SCATTER IN FULLY 3D PET
    OLLINGER, JM
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) : 1168 - 1173
  • [5] Implementation of a single scatter simulation algorithm for 3D PET: Application to emission and transmission scanning
    Accorsi, R
    Adam, LE
    Werner, ME
    Karp, JS
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 816 - 820
  • [6] Scatter Correction with Combined Single-Scatter Simulation and Monte Carlo Simulation for 3D PET
    Ye, Jinghan
    Song, Xiyun
    Hu, Zhiqiang
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [7] Implementation and Evaluation of a 3D PET Single Scatter Simulation with TOF Modeling
    Werner, Matthew E.
    Surti, Suleman
    Karp, Joel S.
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 1768 - 1773
  • [8] Analytic correction for scatter in fully 3D PET: Statistical issues
    Ollinger, JM
    1997 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1 & 2, 1998, : 1386 - 1389
  • [9] Attenuation and scatter correction in fully 3D limited angle PET
    Pönisch, F
    Enghardt, W
    Lauckner, K
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 2115 - 2118
  • [10] Model-based scatter correction for fully 3D PET
    Ollinger, JM
    PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01): : 153 - 176