Kinetic Passive Sampling: In Situ Calibration Using the Contaminant Mass Measured in Parallel Samplers with Different Thicknesses

被引:8
作者
Fuchte, Hanna E. [1 ]
Schaeffer, Andreas [1 ,2 ]
Booij, Kees [3 ]
Smith, Kilian E. C. [1 ]
机构
[1] Rhein Westfal TH Aachen, Inst Environm Res, D-52074 Aachen, Germany
[2] Nanjing Univ, Sch Environm, State Key Lab Pollut Control & Resource Reuse, Nanjing 210093, Peoples R China
[3] Pass Sampling Organ Cpds PaSOC, NL-8821 LV Kimswerd, Netherlands
关键词
POLYCYCLIC AROMATIC-HYDROCARBONS; SEMIPERMEABLE-MEMBRANE DEVICES; ORGANIC CONTAMINANTS; CHEMICALS; WATER; EQUILIBRIUM; PERFORMANCE; POLLUTANTS; EXTRACTION; SEDIMENTS;
D O I
10.1021/acs.est.0c04437
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The use of single-phase passive samplers is a common method for sampling bioavailable concentrations of hydrophobic aquatic pollutants. Often such samplers are used in the kinetic stage, and in situ calibration is necessary. Most commonly, exchange kinetics are derived from the release rates of performance reference compounds (PRCs). In this study, a complementary calibration approach was developed, in which measuring the contaminant mass ratio (CMR) from two samplers with different thicknesses allows the dissolved concentrations to be determined. This new CMR calibration was tested (1) in a laboratory experiment with defined and constant concentrations and (2) in the field, at a storm water retention site. Silicone passive samplers with different thicknesses were used to sample a range of dissolved polycyclic aromatic hydrocarbons. In the laboratory study, the concentrations derived from the CMR calibration were compared with those from water extraction and passive dosing and differences below a factor 2 were found. In the field study, CMR-derived concentrations were compared to those from PRC calibration. Here, differences ranged by only a factor 1 to 3 between both methods. These findings indicate that the CMR calibration can be applied as a stand-alone or complementary calibration method for kinetic passive sampling.
引用
收藏
页码:15759 / 15767
页数:9
相关论文
共 3 条
  • [1] Using performance reference compounds to compare mass transfer calibration methodologies in passive samplers deployed in the water column
    Joyce, Abigail S.
    Burgess, Robert M.
    ENVIRONMENTAL TOXICOLOGY AND CHEMISTRY, 2018, 37 (08) : 2089 - 2097
  • [2] Calibration of Silicone Rubber Rods as Passive Samplers for Pesticides at Two Different Flow Velocities: Modeling of Sampling Rates Under Water Boundary Layer and Polymer Control
    Martin, Alexis
    Margoum, Christelle
    Jolivet, Antoine
    Assoumani, Azziz
    El Moujahid, Bachir
    Randon, Jerome
    Coquery, Marina
    ENVIRONMENTAL TOXICOLOGY AND CHEMISTRY, 2018, 37 (04) : 1208 - 1218
  • [3] Rapid in situ quantitation of photoinitiators in packaging by two-points kinetic calibration of liquid microjunction surface sampling-mass spectrometry
    Mei, Yu
    Wu, Qian
    Zhou, Shiwen
    Wang, Zhiguo
    Liang, Qiuju
    Li, Youmei
    Lu, Hongmei
    TALANTA, 2020, 216