Bi-enzyme sensor based on thick-film carbon electrode modified with electropolymerized tyramine

被引:19
作者
Suprun, EV
Budnikov, HC
Evtugyn, GA
Brainina, KZ
机构
[1] Kazan VI Lenin State Univ, Chem Fac, Kazan 420008, Russia
[2] Kazan VI Lenin State Univ, Ecol Fac, Kazan 420008, Russia
[3] Ural Econ Univ, Ekaterinburg 620219, Russia
关键词
biosensor; enzyme sensor; electropolymerization; polytyramine;
D O I
10.1016/j.bioelechem.2003.10.025
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Bi-enzyme sensor based on thick-film epoxy-carbon electrode modified with polytyramine has been developed and examined for the determination of peroxidase substrates and cholinesterase inhibitors. Polytyramine was obtained on the electrode surface by repeated scanning of the potential from + 600 to + 1800 mV vs. Ag/AgCl in tyramine solution. The enzymes were immobilized in the polytyramine matrix by cross-linking with glutaraldehyde. The biosensor developed provides a reliable and inexpensive way for preliminary testing of common environmental pollutants with a single sensor in accordance with assumed toxic effect by the choice of appropriate substrate and measurement conditions. The bi-enzyme sensor makes it possible to determine substituted phenols and aromatic amines in the micromolar range of their concentrations and anticholinesterase pesticides with detection limits of 0.1 (Coumaphos) and 0.03 mumol l(-1) (Chloropyrifos-methyl). (C) 2004 Published by Elsevier B.V.
引用
收藏
页码:281 / 284
页数:4