A bimodal optoelectronic flow-through detector for phosphate determination

被引:17
作者
Fiedoruk, Marta [1 ]
Mieczkowska, Elzbieta [1 ]
Koncki, Robert [1 ]
Tymecki, Lukasz [1 ]
机构
[1] Univ Warsaw, Dept Chem, PL-02093 Warsaw, Poland
关键词
Phosphate; Photometry; Fluorimetry; Flow analysis; Optoelectronic detector; LIGHT-EMITTING-DIODES; ULTRATRACE ANALYSIS; INJECTION-ANALYSIS; ANALYSIS SYSTEM; PHOSPHORUS; DEVICE; SENSOR; FLUORESCENCE; ABSORBENCY; PHOTOMETRY;
D O I
10.1016/j.talanta.2014.04.086
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A miniature flow-through detector useful for bimodal, photometric and fluorimetric, determination of phosphates has been developed. This optoelectronic device made of four light emitting diodes (LEDs) integrated in the form of 85 mu L optical cell is easily applied in flow analysis manifolds. These LEDs play the roles of light source for photometric measurements, fluorescence inductors and detector of absorbance and fluorescence. For photometric mode of determinations a phosphomolybdenum blue method has been applied. The fluorimetric method of phosphate determination is based on quenching of rhodamine fluorescence by the heteropolyacid. The developed detector used in a simple three-channel flow injection analysis (FIA) system allows photometric or fluorimetric determination of phosphate in the wide range of concentration. The detection limits found for photometric and fluorimetric modes of FIA measurements are 5.5 mg L-1 and 10.4 mu g L-1, respectively. The potential utility of the flow-through detector for the needs of food and clinical analysis has been demonstrated. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:211 / 214
页数:4
相关论文
共 39 条
  • [1] Towards the development of a miniaturized fiberless optofluidic biosensor for glucose
    Cocovi-Solberg, David J.
    Miro, Manuel
    Cerda, Victor
    Pokrzywnicka, Marta
    Tymecki, Lukasz
    Koncki, Robert
    [J]. TALANTA, 2012, 96 : 113 - 120
  • [2] Integrated flow analysis platform for the direct detection of nitrate in water using a simplified chromotropic acid method
    Cogan, Deirdre
    Cleary, John
    Phelan, Thomas
    McNamara, Eoghan
    Bowkett, Mark
    Diamond, Dermot
    [J]. ANALYTICAL METHODS, 2013, 5 (18) : 4798 - 4804
  • [3] Portable integrated microfluidic analytical platform for the monitoring and detection of nitrite
    Czugala, Monika
    Fay, Cormac
    O'Connor, Noel E.
    Corcoran, Brian
    Benito-Lopez, Fernando
    Diamond, Dermot
    [J]. TALANTA, 2013, 116 : 997 - 1004
  • [4] Optical sensing system based on wireless paired emitter detector diode device and ionogels for lab-on-a-disc water quality analysis
    Czugala, Monika
    Gorkin, Robert, III
    Phelan, Thomas
    Gaughran, Jennifer
    Curto, Vincenzo Fabio
    Ducree, Jens
    Diamond, Dermot
    Benito-Lopez, Fernando
    [J]. LAB ON A CHIP, 2012, 12 (23) : 5069 - 5078
  • [5] Light emitting diode-based detectors absorbance, fluorescence and spectroelectrochemical measurements in a planar flow-through cell
    Dasgupta, PK
    Eom, IY
    Morris, KJ
    Li, JZ
    [J]. ANALYTICA CHIMICA ACTA, 2003, 500 (1-2) : 337 - 364
  • [6] Absorbance measurements with light-emitting diodes as sources: Silicon photodiodes or light-emitting diodes as detectors?
    Duy Anh Bui
    Hauser, Peter C.
    [J]. TALANTA, 2013, 116 : 1073 - 1078
  • [7] Flow analysis techniques for phosphorus:: an overview
    Estela, JM
    Cerdà, V
    [J]. TALANTA, 2005, 66 (02) : 307 - 331
  • [8] Fiedoruk M., 2014, TALANTA UNPUB
  • [9] Using sequential injection analysis for fast determination of phosphate in coastal waters
    Frank, Carsten
    Schroeder, Friedhelm
    Ebinghaus, Ralf
    Ruck, Wolfgang
    [J]. TALANTA, 2006, 70 (03) : 513 - 517
  • [10] FLUOROPHOTOMETRIC DETERMINATION OF PHOSPHATE AFTER COLLECTION ON A MEMBRANE-FILTER AS ION-PAIR OF MOLYBDOPHOSPHATE WITH RHODAMINE-6G
    KAN, M
    NASU, T
    TAGA, M
    [J]. ANALYTICAL SCIENCES, 1991, 7 (01) : 87 - 91