Benchtop Assessment of a New Single-Use Flexible Ureteroscope

被引:2
作者
Whelan, Patrick [1 ]
Terry, Russell S. [1 ]
Qi, Robert [1 ]
Ketterman, Brian [2 ]
Preminger, Glenn M. [1 ]
Lipkin, Michael E. [1 ]
机构
[1] Duke Univ, Med Ctr, Div Urol, 40 Duke Med Circle,Room 1573, Durham, NC 27710 USA
[2] Duke Univ, Sch Med, Durham, NC USA
关键词
ureteroscopy; single-use ureteroscope; nephrolithiasis; laser lithotripsy; ASSOCIATION/ENDOUROLOGICAL SOCIETY GUIDELINE; SURGICAL-MANAGEMENT; STONES; LITHOVUE; COSTS;
D O I
10.1089/end.2020.0836
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Single-use flexible ureteroscopes are an increasingly popular alternative to reusable ureteroscopes. In this study, we performed a benchtop examination of the physical and optical properties of the new Dornier Axis (TM) (Webling, Germany) single-use ureteroscope. Methods: Ten new, never-used Dornier Axis ureteroscopes were assessed for optical performance, maximal tip deflection, and irrigation flow rate with an empty working channel and with insertion of 200 and 365 mu m laser fibers, and a 1.9F nitinol basket. All ureteroscopes were then fully deflected 100 times in each direction, and maximal deflection angles were re-measured with and without instruments in the working channel. All measurements were performed in duplicate. In vitro optical testing for resolution, image distortion, and depth of field was performed and compared vs the LithoVue (TM) (Boston Scientific, Marlborough, MA) single-use ureteroscope. Statistical analyses using paired Wilcoxon rank-sum tests and Kruskal-Wallis multiple-group comparison tests were performed in R. Results: Median maximal deflection angles exceeded 300 degrees in both directions before and after 100 full deflection cycles for all groups except the 365 mu m laser fiber group. After 100 deflection cycles, there was no change in the majority of working instruments, except a decrease in upward flexion with an empty channel and 200 mu m Moses (TM) laser fiber, and downward flexion with 200 mu m Flexiva (TM) laser fiber (all <10 degrees). After excluding the 365 mu m fiber, there was no difference in multi-group comparison for upward and downward flexion pre- and post-cycling. Median flow rate through an empty channel was 48.0 mL/min, and it decreased significantly with all used instruments (p < 0.001). Compared with the LithoVue, the Axis demonstrated superior resolution at all tested distances and less distortion. Conclusions: The new Dornier Axis single-use ureteroscope demonstrates excellent tip deflection, which remains unchanged after 100 manual flexions in each direction. The Axis also demonstrates superior optical performance compared with the LithoVue in benchtop testing.
引用
收藏
页码:755 / 760
页数:6
相关论文
共 24 条
  • [1] Surgical Management of Stones: American Urological Association/Endourological Society Guideline, PART II
    Assimos, Dean
    Krambeck, Amy
    Miller, Nicole L.
    Monga, Manoj
    Murad, M. Hassan
    Nelson, Caleb P.
    Pace, Kenneth T.
    Pais, Vernon M., Jr.
    Pearle, Margaret S.
    Preminger, Glenn M.
    Razvi, Hassan
    Shah, Ojas
    Matlaga, Brian R.
    [J]. JOURNAL OF UROLOGY, 2016, 196 (04) : 1161 - 1169
  • [2] Assimos D, 2016, J UROLOGY, V196, P1153, DOI 10.1016/j.juro.2016.05.090
  • [3] Boston Scientific, 2016, LITHOVUE SINGLE US D
  • [4] Outbreak of ertapenem-resistant Enterobacter cloacae urinary tract infections due to a contaminated ureteroscope
    Chang, C. -L.
    Su, L. -H.
    Lu, C. -M.
    Tai, F. -T.
    Huang, Y. -C.
    Chang, K. -K.
    [J]. JOURNAL OF HOSPITAL INFECTION, 2013, 85 (02) : 118 - 124
  • [5] Evaluation of Performance Parameters of the Disposable Flexible Ureterorenoscope (LITHOVUE) in Patients with Renal Stones: A Prospective, Observational, Single-arm, Multicenter Study
    Cho, Sung Yong
    Lee, Joo Yong
    Shin, Dong Gil
    Seo, Ill Young
    Yoo, Sangjun
    Park, Hyung Keun
    [J]. SCIENTIFIC REPORTS, 2018, 8
  • [6] Evaluation of a Novel Single-Use Flexible Ureteroscope
    Dale, Joanne
    Kaplan, Adam G.
    Radvak, Daniela
    Shin, Richard
    Ackerman, Anika
    Chen, Tony
    Scales, Charles D., Jr.
    Ferrandino, Michael N.
    Simmons, Walter Neal
    Preminger, Glenn M.
    Lipkin, Michael E.
    [J]. JOURNAL OF ENDOUROLOGY, 2021, 35 (06) : 903 - 907
  • [7] Dornier MedTech, 2019, DORN AX SINGL US DIG
  • [8] Characteristics of current digital single-use flexible ureteroscopes versus their reusable counterparts: an in-vitro comparative analysis
    Dragos, Laurian B.
    Somani, Bhaskar K.
    Keller, Etienne X.
    De Coninck, Vincent M. J.
    Herrero, Maria Rodriguez-Monsalve
    Kamphuis, Guido M.
    Bres-Niewada, Ewa
    Sener, Emre T.
    Doizi, Steeve
    Wiseman, Oliver J.
    Traxer, Olivier
    [J]. TRANSLATIONAL ANDROLOGY AND UROLOGY, 2019, 8 : S359 - S370
  • [9] FDA, 2019, 00810116020690 DORN
  • [10] Gridley CM, 2017, RES REP UROL, V9, P19, DOI 10.2147/RRU.S104229