摘要
Alumina ceramic-on-ceramic bearings perform exceptionally well under standard hip simulator conditions, but in vivo some retrieved bearings have shown an unusual stripe pattern of wear. We studied 16 bearings retrieved from a series of 1,588 cementless hip arthroplasties with third generation alumina ceramic-on-ceramic bearings to characterize the mechanism of stripe wear formation. None of these bearings were retrieved for bearing failure. The average wear volume was 0.4 mm 3 per year in the heads and 0.3 mm(3) per year in the liners. Mapping of wear stripes on the heads and liners showed that the majority do not occur with normal walking; instead they probably occur with edge loading when the hip is flexed, such as with rising from a chair or with climbing a high step. (C) 2004 Elsevier Inc. All rights reserved.
引用
收藏
页码:402 / 413
页数:12
相关论文
共 22 条
[1]   The effect of femoral component head size on posterior dislocation of the artificial hip joint [J].
Bartz, RL ;
Noble, PC ;
Kadakia, NR ;
Tullos, HS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (09) :1300-1307
[2]   Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[3]   The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios [J].
D'Lima, DD ;
Urquhart, AG ;
Buehler, KO ;
Walker, RH ;
Colwell, CW .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (03) :315-321
[4]  
DIGAS GK, 2003, 49 ANN M ORTHR RES S
[5]   2 STRATEGIES OF TRANSFERRING FROM SIT-TO-STAND - THE ACTIVATION OF MONOARTICULAR AND BIARTICULAR MUSCLES [J].
DOORENBOSCH, CAM ;
HARLAAR, J ;
ROEBROECK, ME ;
LANKHORST, GJ .
JOURNAL OF BIOMECHANICS, 1994, 27 (11) :1299-1307
[6]   A literature review of the association between wear rate and osteolysis in total hip arthroplasty [J].
Dumbleton, JH ;
Manley, MT ;
Edidin, AA .
JOURNAL OF ARTHROPLASTY, 2002, 17 (05) :649-661
[7]  
FISHER J, 2002, IMECHE JUN 13 15 LON
[8]   Effect of muscle strength and movement speed on the biomechanics of rising from a chair in healthy elderly and young women [J].
Gross, MM ;
Stevenson, PJ ;
Charette, SL ;
Pyka, G ;
Marcus, R .
GAIT & POSTURE, 1998, 8 (03) :175-185
[9]   Effects of clinically relevant alumina ceramic wear particles on TNF-α production by human peripheral blood mononuclear phagocytes [J].
Hatton, A ;
Nevelos, JE ;
Matthews, JB ;
Fisher, J ;
Ingham, E .
BIOMATERIALS, 2003, 24 (07) :1193-1204
[10]  
HOLLEY KFB, 2003, HOSP SPEC SURG ALUMN