共 32 条
Dispersive Liquid-Liquid Microextraction of Silver Prior to Determination by Microsample Introduction-Flame Atomic Absorption Spectrometry
被引:36
作者:
Jafarvand, Sanaz
[1
,2
]
Bidari, Araz
[1
,2
]
Hemmatkhah, Payam
[1
,2
]
Hosseini, Mohammad Reza Milani
[1
,2
]
Assadi, Yaghoub
[1
,2
]
机构:
[1] Iran Univ Sci & Technol, Fac Chem, Dept Analyt Chem, Tehran, Iran
[2] Iran Univ Sci & Technol, Electroanalyt Chem Res Ctr, Tehran, Iran
关键词:
Dispersive liquid-liquid microextraction;
flame atomic absorption spectrometry;
microsample introduction;
silver;
water analysis;
GAS-CHROMATOGRAPHY;
TRACE AMOUNTS;
SELECTIVE DETERMINATION;
SENSITIVE METHOD;
ARRAY DETECTION;
WASTE-WATER;
SILICA-GEL;
PRECONCENTRATION;
EXTRACTION;
SAMPLES;
D O I:
10.1080/00032710903137400
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
A new simple and sensitive method has been proposed for rapid determination of trace levels of silver in environmental water samples, using dispersive liquid-liquid microextraction (DLLME) prior to its microsample introduction-flame atomic absorption spectrometry. Under the optimum conditions, the linear range was 0.1-7 mu g L-1 and limit of detection was 0.018 mu g L-1. The relative standard deviation for 0.50 and 5.00 mu g L-1 of silver in water sample was 4.0 and 1.7%, respectively. The relative recoveries of silver from tap, well, river, and seawater samples at spiking levels of 1.00 and 5.00 mu g L-1 were in the range of 86.4-98.6%.
引用
收藏
页码:2214 / 2231
页数:18
相关论文