Performance characteristics of five new anesthesia ventilators and four intensive care ventilators in pressure-support mode - A comparative bench study

被引:45
作者
Jaber, Samir
Tassaux, Didier
Sebbane, Mustapha
Pouzeratte, Yvan
Battisti, Anne
Capdevila, Xavier
Eledjam, Jean-Jacques
Jolliet, Philippe
机构
[1] Univ Montpellier 1, Hop St Eloi, Ctr Hosp, DAR B,Dept Anesthesia & Crit Care B, F-34295 Montpellier 5, France
[2] Univ Geneva, Cantonal Hosp, Div Med Intens Care, CH-1211 Geneva, Switzerland
[3] Univ Montpellier 1, CHU Montpellier, Hop Lapeyronie, Dept Anesthesia & Crit Care A,DAR A, F-34006 Montpellier, France
关键词
D O I
10.1097/00000542-200611000-00015
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: During the past few years, many manufacturers have introduced new modes of ventilation in anesthesia ventilators, especially partial-pressure modalities. The current bench test study was designed to compare triggering and pressurization of five new anesthesia ventilators with four intensive care unit ventilators. Methods: Ventilators were connected to a two-compartment lung model. One compartment was driven by an intensive care unit ventilator to mimic "patient" inspiratory effort, whereas the other was connected to the tested ventilator. The settings of ventilators were positive end-expiratory pressures of 0 and 5 cm H2O, and pressure-support ventilation levels of 10, 15, and 20 cm H2O with normal and high "patient" inspiratory effort. For the anesthesia ventilators, all the measurements were obtained for a low (1 l/min) and a high (10 l/min) fresh gas flow. Triggering delay, triggering workload, and pressurization at 300 and 500 ms were analyzed. Results: For the five tested anesthesia ventilators, the pressure-support ventilation modality functioned correctly. For inspiratory triggering, the three most recent anesthesia machines (Fabius, Dragerwerk AG, Lubeck, Germany; Primus, Dragerwerk AG; and Avance, GE-Datex-Ohemda, Munchen, Germany) had a triggering delay of less than 100 ms, which is considered clinically satisfactory and is comparable to intensive care unit machines. The use of positive end-expiratory pressure modified the quality of delivered pressure support for two anesthesia ventilators (Kion, Siemens AG, Munich, Germany; and Felix, Taema, Antony, France). Three of the five anesthesia ventilators exhibited pressure-support ventilation performance characteristics comparable to those of the intensive care unit machines. Increasing fresh gas flow (1 to 10 l/min) in the internal circuit did not influence the pressure-support ventilation performance of the anesthesia ventilators. Conclusion: Regarding trigger sensitivity and the system's ability to meet inspiratory flow during pressure-supported breaths, the most recent anesthesia ventilators have comparable performances of recent-generation intensive care unit ventilators.
引用
收藏
页码:944 / 952
页数:9
相关论文
共 28 条
[1]   Effects of flow triggering on breathing effort during partial ventilatory support [J].
Aslanian, P ;
El Atrous, S ;
Isabey, D ;
Valente, E ;
Corsi, D ;
Harf, A ;
Lemaire, F ;
Brochard, L .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1998, 157 (01) :135-143
[2]   Performance characteristics of 10 home mechanical ventilators in pressure-support mode - A comparative bench study [J].
Battisti, A ;
Tassaux, D ;
Janssens, JP ;
Michotte, JB ;
Jaber, S ;
Jolliet, P .
CHEST, 2005, 127 (05) :1784-1792
[3]   Variable ventilation improves perioperative lung function in patients undergoing abdominal aortic aneurysmectomy [J].
Boker, A ;
Haberman, CJ ;
Girling, L ;
Guzman, RP ;
Louridas, G ;
Tanner, JR ;
Cheang, M ;
Maycher, BW ;
Bell, DD ;
Doak, GJ .
ANESTHESIOLOGY, 2004, 100 (03) :608-616
[4]   Pressure support improves efficiency of spontaneous breathing during inhalation anesthesia [J].
Bosek, V ;
Roy, L ;
Smith, RA .
JOURNAL OF CLINICAL ANESTHESIA, 1996, 8 (01) :9-12
[5]  
Brégeon F, 2005, ANESTHESIOLOGY, V102, P331
[6]   Pressure support ventilation versus continuous positive airway pressure with the laryngeal mask airway -: A randomized crossover study of anesthetized adult patients [J].
Brimacombe, J ;
Keller, C ;
Hörmann, C .
ANESTHESIOLOGY, 2000, 92 (06) :1621-1623
[7]   Performance characteristics of bilevel pressure ventilators - A lung model study [J].
Bunburaphong, T ;
Imanaka, H ;
Nishimura, M ;
Hess, D ;
Kacmarek, RM .
CHEST, 1997, 111 (04) :1050-1060
[8]  
CHRISTIE JM, 1992, ANESTH ANALG, V75, P167
[9]   THE LARYNGEAL MASK AIRWAY AND POSITIVE-PRESSURE VENTILATION [J].
DEVITT, JH ;
WENSTONE, R ;
NOEL, AG ;
ODONNELL, MP .
ANESTHESIOLOGY, 1994, 80 (03) :550-555
[10]   PHRENIC-NERVE STIMULATION DURING HALOTHANE ANESTHESIA [J].
HEDENSTIERNA, G ;
TOKICS, L ;
LUNDQUIST, H ;
ANDERSSON, T ;
STRANDBERG, A ;
BRISMAR, B .
ANESTHESIOLOGY, 1994, 80 (04) :751-760