Evaluation of auto-planning in IMRT and VMAT for head and neck cancer

被引:29
|
作者
Ouyang, Zi [1 ]
Shen, Zhilei Liu [1 ]
Murray, Eric [1 ]
Kolar, Matt [1 ]
LaHurd, Danielle [1 ]
Yu, Naichang [1 ]
Joshi, Nikhil [1 ]
Koyfman, Shiomo [1 ]
Bzdusek, Karl [2 ]
Xia, Ping [1 ]
机构
[1] Cleveland Clin, Dept Radiat Oncol, Cleveland, OH 44106 USA
[2] Philips Healthcare Inc, Fitchburg, WI USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2019年 / 20卷 / 07期
关键词
auto-planning; head and neck; IMRT; step and shoot; VMAT; QUALITY; DVH;
D O I
10.1002/acm2.12652
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The purposes of this work are to (a) investigate whether the use of auto-planning and multiple iterations improves quality of head and neck (HN) radiotherapy plans; (b) determine whether delivery methods such as step-and-shoot (SS) and volumetric modulated arc therapy (VMAT) impact plan quality; (c) report on the observations of plan quality predictions of a commercial feasibility tool. Materials and methods Twenty HN cases were retrospectively selected from our clinical database for this study. The first ten plans were used to test setting up planning goals and other optimization parameters in the auto-planning module. Subsequently, the other ten plans were replanned with auto-planning using step-and-shoot (AP-SS) and VMAT (AP-VMAT) delivery methods. Dosimetric endpoints were compared between the clinical plans and the corresponding AP-SS and AP-VMAT plans. Finally, predicted dosimetric endpoints from a commercial program were assessed. Results All AP-SS and AP-VMAT plans met the clinical dose constraints. With auto-planning, the dose coverage of the low dose planning target volume (PTV) was improved while the dose coverage of the high dose PTV was maintained. Compared to the clinical plans, the doses to critical organs, such as the brainstem, parotid, larynx, esophagus, and oral cavity were significantly reduced in the AP-VMAT (P < 0.05); the AP-SS plans had similar homogeneity indices (HI) and conformality indices (CI) and the AP-VMAT plans had comparable HI and improved CI. Good agreement in dosimetric endpoints between predictions and AP-VMAT plans were observed in five of seven critical organs. Conclusion With improved planning quality and efficiency, auto-planning module is an effective tool to enable planners to generate HN IMRT plans that are meeting institution specific planning protocols. DVH prediction is feasible in improving workflow and plan quality.
引用
收藏
页码:39 / 47
页数:9
相关论文
共 50 条
  • [31] EMPHASIZING CONFORMAL AVOIDANCE VERSUS TARGET DEFINITION FOR IMRT PLANNING IN HEAD-AND-NECK CANCER
    Harari, Paul M.
    Song, Shiyu
    Tome, Wolfgang A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 77 (03): : 950 - 958
  • [32] Novel VMAT planning technique improves dosimetry for head and neck cancer patients undergoing definitive chemoradiotherapy
    DiBartolo, David
    Carpenter, Todd
    Santoro, Joseph P.
    Lischalk, Jonathan W.
    Ebling, David
    Haas, Jonathan A.
    Witten, Matthew
    Rybstein, Marissa
    Vaezi, Alec
    Repka, Michael C.
    ACTA ONCOLOGICA, 2023, 62 (02) : 189 - 193
  • [33] Multi-institutional evaluation of knowledge-based planning performance of volumetric modulated arc therapy (VMAT) for head and neck cancer
    Kamima, Tatsuya
    Ueda, Yoshihiro
    Fukunaga, Jun-ichi
    Shimizu, Yumiko
    Tamura, Mikoto
    Ishikawa, Kazuki
    Monzen, Hajime
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 64 : 174 - 181
  • [34] Evaluation of larynx-sparing techniques with IMRT when treating the head and neck
    Webster, Gareth J.
    Rowburrom, Carl G.
    Ho, Kean F.
    Slevin, Nick J.
    Mackay, Ranald I.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (02): : 617 - 622
  • [35] An evaluation of MR based deep learning auto-contouring for planning head and neck radiotherapy
    Hague, C.
    McPartlin, A.
    Lee, L. W.
    Hughes, C.
    Mullan, D.
    Beasley, W.
    Green, A.
    Price, G.
    Whitehurst, P.
    Slevin, N.
    van Herk, M.
    West, C.
    Chuter, R.
    RADIOTHERAPY AND ONCOLOGY, 2021, 158 : 112 - 117
  • [36] Evaluation of volumetric modulated arc therapy (VMAT) with Oncentra MasterPlan® for the treatment of head and neck cancer
    Alvarez-Moret, Judith
    Pohl, Fabian
    Koelbl, Oliver
    Dobler, Barbara
    RADIATION ONCOLOGY, 2010, 5
  • [37] Evaluation of set-up errors using CBCT in Head and Neck cancer patients treated with IMRT in UMMC
    Ramachandran, Rangasamy
    Satar, Nur Fadhlina Abdul
    Jamalludin, Zulaikha
    Jong, Wei Loong
    Yuin, Jasmin Loh Pei
    RADIOTHERAPY AND ONCOLOGY, 2024, 197 : S324 - S325
  • [38] A comprehensive evaluation of the quality and complexity of prostate IMRT and VMAT plans generated by an automated inverse planning tool
    Wilkinson, Dean
    Mackie, Kelly
    Novy, Dean
    Beaven, Frances
    McNamara, Joanne
    Bailey, Renee
    Currie, Michael
    Nasser, Elias
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2022, 21 (04) : 506 - 512
  • [39] Intensity modulated proton therapy (IMPT) - The future of IMRT for head and neck cancer
    Moreno, Amy C.
    Frank, Steven J.
    Garden, Adam S.
    Rosenthal, David I.
    Fuller, Clifton D.
    Gunn, Gary B.
    Reddy, Jay P.
    Morrison, William H.
    Williamson, Tyler D.
    Holliday, Emma B.
    Phan, Jack
    Blanchard, Pierre
    ORAL ONCOLOGY, 2019, 88 : 66 - 74
  • [40] Dosimetric impact of Acuros XB dose-to-water and dose-to-medium reporting modes on VMAT planning for head and neck cancer
    Munoz-Montplet, Caries
    Marruecos, Jordi
    Buxo, Maria
    Jurado-Bruggeman, Diego
    Romera-Martinez, Ingrid
    Bueno, Marta
    Vilanova, Joan C.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 55 : 107 - 115