Ablation lesion size correlates with pacing threshold: A physiological basis for use of pacing to assess ablation lesions

被引:20
作者
Sapp, JL
Soejima, K
Cooper, JM
Epstein, LM
Stevenson, WG
机构
[1] Harvard Univ, Brigham & Womens Hosp, Sch Med, Dept Internal Med,Cardiovasc Div, Boston, MA 02115 USA
[2] Univ Penn, Med Ctr, Philadelphia, PA 19104 USA
[3] Dalhousie Univ, QEII Hlth Sci Ctr, Dept Med, Div Cardiol, Halifax, NS, Canada
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2004年 / 27卷 / 07期
关键词
catheter ablation; radiofrequency lesions; pacing threshold; electrogram amplitude; virtual electrode;
D O I
10.1111/j.1540-8159.2004.00561.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The virtual electrode model predicts that pacing stimulus strength should reflect proximity of the pacing electrode to excitable myocardium, allowing pacing threshold to assess radiofrequency (RF) ablation lesions and unexcitable scar. The purpose of this study is to correlate RF lesion size with pacing threshold and electrogram (EG) amplitude change at the ablation site. In four swine (32-58 kg, 20 ventricular RF lesions were created using a 4-mm tip electrode catheters underfluoroscopic and electroanatomic guidance. Unipolar pacing threshold and bipolar and unipolar EG amplitude were measured before and after ablation and compared with lesion size measured in the flxed, serially sectioned tissue. Lesion diameter ranged from 6.4 to 19 mm and volume ranged from 29 to 1920 mm.(3) Ablation increased the pacing threshold by 320%, from 0.9 +/- 0.3 to 3.6 +/- 2.6 mA, P < 0.001. The change in pacing threshold correlated with lesion volume R = 0.88, P < 0.001). Linear regression predicts that lesion volume (mm 3) = 160 X rise in pacing threshold + 13. Ablation reduced peak to peak bipolar EG amplitude by 56%, from 2.5 +/- 2.0 mV to 1.1 +/- 0.6 mV (P = 0.005). Unipolar EG amplitude diminished by only 22% from 4.0 +/- 1.6 to 3.2 +/- 0.9 mV postablation (P = 0.005). The correlations of lesion volume with change in either bipolar R = 0.14, P = 0.6) or unipolar R = 0.18, P = 0.6) EG amplitude were poor. Pacing threshold correlates with RF ablation lesion size, consistent with the virtual electrode model. In normal myocardium, change in pacing threshold is likely to be a better marker of lesion size than electrogram amplitude.
引用
收藏
页码:933 / 937
页数:5
相关论文
共 13 条
[1]   PHYSICS AND ENGINEERING OF TRANSCATHETER CARDIAC TISSUE ABLATION [J].
AVITALL, B ;
KHAN, M ;
KRUM, D ;
HARE, J ;
LESSILA, C ;
DHALA, A ;
DESHPANDE, S ;
JAZAYERI, M ;
SRA, J ;
AKHTAR, M .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (03) :921-932
[2]   Assessment of radiofrequency ablation effect from unipolar pacing threshold [J].
Delacretaz, E ;
Soejima, K ;
Brunckhorst, CB ;
Maisel, WH ;
Friedman, PL ;
Stevenson, WG .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2003, 26 (10) :1993-1996
[3]   THE BIOPHYSICS OF RADIOFREQUENCY CATHETER ABLATION IN THE HEART - THE IMPORTANCE OF TEMPERATURE MONITORING [J].
HAINES, DE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03) :586-591
[4]   CLOSED CHEST CATHETER DESICCATION OF THE ATRIOVENTRICULAR JUNCTION USING RADIOFREQUENCY ENERGY - A NEW METHOD OF CATHETER ABLATION [J].
HUANG, SK ;
BHARATI, S ;
GRAHAM, AR ;
LEV, M ;
MARCUS, FI ;
ODELL, RC .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 9 (02) :349-358
[5]   Prospective comparison of lesions created using a multipolar microcatheter ablation system with those created using a pullback approach with standard radiofrequency ablation in the canine atrium [J].
Jumrussirikul, P ;
Atiga, WL ;
Lardo, AC ;
Berger, RD ;
Halperin, H ;
Hutchins, GM ;
Calkins, H .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (02) :203-213
[6]   Electrophysiological correlates of transmural linear ablation [J].
Liem, LB ;
Pomeranz, M ;
Riseling, K ;
Anderson, S ;
Berry, GJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (01) :40-46
[7]   EFFECT OF PACING CURRENT STRENGTH ON INDEXES OF MYOCARDIAL ACTIVATION IN HUMANS - INFLUENCE OF CHRONIC INFARCTION AND POLARITY OF PACING [J].
MEESMANN, M ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (02) :393-401
[8]  
ODONNELL D, 2002, PACE, V25, P544
[9]   Electrically unexcitable scar mapping based on pacing threshold for identification of the reentry circuit isthmus - Feasibility for guiding ventricular tachycardia ablation [J].
Soejima, K ;
Stevenson, WG ;
Maisel, WH ;
Sapp, JL ;
Epstein, LM .
CIRCULATION, 2002, 106 (13) :1678-1683
[10]   Ventricular tachycardia associated with myocardial infarct scar - A spectrum of therapies for a single patient [J].
Soejima, K ;
Stevenson, WG .
CIRCULATION, 2002, 106 (02) :176-179