Accuracy of the EASI 12-lead electrocardiogram compared to the standard 12-lead electrocardiogram for diagnosing multiple cardiac abnormalities

被引:66
作者
Drew, BJ
Pelter, MM
Wung, SF
Adams, MG
Taylor, C
Evans, GT
Foster, E
机构
[1] Univ Calif San Francisco, Sch Nursing, Dept Physiol Nursing, San Francisco, CA 94143 USA
[2] Univ Calif San Francisco, Dept Med, Div Cardiol, San Francisco, CA 94143 USA
关键词
electrocardiogram; myocardial ischemia; ST segments; physiologic monitoring;
D O I
10.1016/S0022-0736(99)90033-X
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study was performed to compare a derived 12-lead electrocardiogram (ECG) using a simple 5-electrode lead configuration (EASI 12-lead) with the standard ECG for multiple cardiac diagnoses. Accurate diagnosis of arrhythmias and ischernia often require analysis of multiple (ideally, 12) ECG leads; however, continuous 12-lead monitoring is impractical in hospital settings. EASI and standard ECGs were compared in 540 patients, 426 of whom also had continuous 12-lead ST segment monitoring with both lead methods. Independent standards relative to a correct diagnosis were used whenever possible, for example, echocardiographic data for chamber enlargement-hypertrophy, and troponin levels for acute infarction. Percent agreement between the 2 methods were: cardiac rhythm, 100%; chamber enlargement-hypertrophy, 84%-99%; right and left bundle branch block 95% and 97%, respectively; left anterior and posterior fascicular block, 97% and 99%, respectively; prior anterior and inferior infarction, 95% and 92%, respectively. There was very little variation between the 2 lead methods in cardiac interval measurements; however, there was more variation in P, QRS, and T-wave axes. Of the 426 patients with ST monitoring, 138 patients had a total of 238 ST events (26, acute infarction; 62, angioplasty-induced ischemia; 150, spontaneous transient ischemia). There was 100% agreement between the 2 methods for acute infarction, 95% agreement for angioplasty-induced ischemia, and 89% agreement for transient ischemia. EASI and standard 12-lead ECGs are comparable for multiple cardiac diagnoses; however, serial ECG changes (eg, T-wave changes) should be assessed using one consistent 12-lead method.
引用
收藏
页码:38 / 47
页数:10
相关论文
共 22 条
[1]   A NEW APPROACH TO THE DIFFERENTIAL-DIAGNOSIS OF A REGULAR TACHYCARDIA WITH A WIDE QRS COMPLEX [J].
BRUGADA, P ;
BRUGADA, J ;
MONT, L ;
SMEETS, J ;
ANDRIES, EW .
CIRCULATION, 1991, 83 (05) :1649-1659
[2]  
Dower G.E., 1996, EASI 12 LEAD ELECTRO
[3]   DERIVING THE 12-LEAD ELECTROCARDIOGRAM FROM 4 (EASI) ELECTRODES [J].
DOWER, GE ;
YAKUSH, A ;
NAZZAL, SB ;
JUTZY, RV ;
RUIZ, CE .
JOURNAL OF ELECTROCARDIOLOGY, 1988, 21 :S182-S187
[4]  
Drew B J, 1998, Am J Crit Care, V7, P355
[5]  
Drew B J, 1996, Am J Crit Care, V5, P198
[6]   ECG criteria to distinguish between aberrantly conducted supraventricular tachycardia and ventricular tachycardia: Practical aspects for the immediate care setting [J].
Drew, BJ ;
Scheinman, MM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (12) :2194-2208
[7]   Comparison of standard and derived 12-lead electrocardiograms for diagnosis of coronary angioplasty-induced myocardial ischemia [J].
Drew, BJ ;
Adams, MG ;
Pelter, MM ;
Wung, SF ;
Caldwell, MA .
AMERICAN JOURNAL OF CARDIOLOGY, 1997, 79 (05) :639-644
[8]   COMPARISON OF A VECTORCARDIOGRAPHICALLY DERIVED 12-LEAD ELECTROCARDIOGRAM WITH THE CONVENTIONAL ELECTROCARDIOGRAM DURING WIDE QRS COMPLEX TACHYCARDIA, AND ITS POTENTIAL APPLICATION FOR CONTINUOUS BEDSIDE MONITORING [J].
DREW, BJ ;
SCHEINMAN, MM ;
EVANS, GT .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 69 (06) :612-618
[9]   RECONSTRUCTION OF THE ELECTROCARDIOGRAM DURING HEART-SURGERY [J].
EDENBRANDT, L ;
ZENG, ZR ;
EDENBRANDT, A ;
MANSSON, C ;
SORNMO, L ;
OLSSON, SB .
COMPUTERS AND BIOMEDICAL RESEARCH, 1992, 25 (06) :538-546
[10]   AN ACCURATE, CLINICALLY PRACTICAL SYSTEM FOR SPATIAL VECTORCARDIOGRAPHY [J].
FRANK, E .
CIRCULATION, 1956, 13 (05) :737-749