Long focal length, asymmetric fan beam collimation for transmission acquisition with a triple camera SPECT system

被引:15
作者
Gilland, DR
Wang, HL
Coleman, RE
Jaszczak, RJ
机构
[1] Department of Radiology, Duke University Medical Center, Durham, NC
关键词
D O I
10.1109/23.596986
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A system design for fast sequential SPECT/transmission CT on a three-headed SPECT camera is evaluated using Monte Carlo simulation. A long focal length (114 cm) asymmetric fan beam collimator is used with a stationary line source positioned at the focal line for transmission acquisition. The advantages of this system include: large field-of-view with 360 degree acquisition, high sensitivity to the transmission line source, high spatial resolution, easy transition between SPECT and transmission acquisition, and no moving line source. For transmission reconstruction an ML-EM algorithm was implemented for the asymmetric fan beam geometry. The results showed that with a 40 cm detector width, transmission reconstructions of a 38 x 26 cm elliptical body were effectively artifact-free. Reconstructions of a 46 x 31 cm body contained only minor truncation artifacts that did not substantially affect the attenuation compensated SPECT image. Contamination of the transmission data from radionuclide within the patient can be substantially reduced with this system by using a high resolution fan beam collimator. We conclude that long focal length, asymmetric fan beam collimation, combined with iterative reconstruction, offers a promising approach for fast sequential SPECT/TCT on a three-headed SPECT camera.
引用
收藏
页码:1191 / 1196
页数:6
相关论文
共 12 条
[1]  
ANGER HO, 1964, J NUCL MED, V5, P515
[2]   ASYMMETRIC FAN TRANSMISSION CT ON SPECT SYSTEMS [J].
CHANG, W ;
LONCARIC, S ;
HUANG, G ;
SANPITAK, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (05) :913-928
[3]  
FICARO EP, 1995, J NUCL MED, V36, P921
[4]  
Gilland D. R., 1996, Journal of Nuclear Medicine, V37, p154P
[5]  
GILLAND DR, 1993, J NUCL MED, V34, P30
[6]  
HAWMAN EG, 1994, J NUCL MED, V35, pP92
[7]  
JASZCZAK RJ, 1993, J NUCL MED, V34, P1577
[8]  
LANGE K, 1984, J COMPUT ASSIST TOMO, V8, P306
[9]   TRUNCATION ARTIFACT SUPPRESSION IN CONE-BEAM RADIONUCLIDE TRANSMISSION CT USING MAXIMUM-LIKELIHOOD TECHNIQUES - EVALUATION WITH HUMAN-SUBJECTS [J].
MANGLOS, SH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (03) :549-562
[10]   FAST CALCULATION OF THE EXACT RADIOLOGICAL PATH FOR A 3-DIMENSIONAL CT ARRAY [J].
SIDDON, RL .
MEDICAL PHYSICS, 1985, 12 (02) :252-255