A compact SPECT/CT system for small animal imaging

被引:11
作者
Hong, Key Jo [1 ]
Choi, Yong
Lee, Sang Chul
Lee, Soo Yeol
Song, Tae Yong
Min, Byung Jun
Jung, Jin Ho
Choe, Yearn Seong
Lee, Kyung-Han
Kim, Byung-Tae
机构
[1] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Nucl Med, Seoul 135710, South Korea
[2] Kyung Hee Univ, Dept Biomed Engn, Yongin 449701, Kyungki, South Korea
关键词
multimodality imaging; small animal imaging; single photon emission computed tomography; X-ray imaging; PINHOLE COLLIMATOR; CT SYSTEM;
D O I
10.1109/TNS.2006.880972
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A dual-modality compact SPECT/CT system for small animal imaging was developed. The SPECT system cons sisted of a pinhole collimator and continuous NaI(Tl) scintillation crystal coupled to a PSPMT. The CT system consisted of a microfocus X-ray tube and a CMOS flat-panel detector. The SPECT system was mounted perpendicular to the X-ray system. Individual projections of the SPECT and the CT were acquired by rotating the animal on a vertical axis in front of the detectors. The SPECT and CT images were reconstructed using OSEM and Feldkamp's cone-beam algorithms, respectively. Mouse and rat SPECT images demonstrated detailed activity distribution at the expected structures. A CT image obtained with 40 kVp and 0.5 mA presented high-resolution anatomic details. Fused SPECT/CT images demonstrated good agreement between the CT images and the corresponding uptake of the radiotracer. The SPECT/CT system developed in this study provides high-quality dual-modality images and could be useful to obtain functional images with high resolution morphology information.
引用
收藏
页码:2601 / 2604
页数:4
相关论文
共 15 条
[1]   Characterization of pinhole SPECT acquisition geometry [J].
Bequé, D ;
Nuyts, J ;
Bormans, G ;
Suetens, P ;
Dupont, P .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2003, 22 (05) :599-612
[2]   X-ray micro-tomography system for small-animal imaging with zoom-in imaging capability [J].
Chun, IK ;
Cho, MH ;
Lee, SC ;
Cho, MH ;
Lee, SY .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (17) :3889-3902
[3]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[4]  
Iwata K, 2002, J NUCL MED, V43, p10P
[5]   Performance improvement of small gamma camera using NaI(Tl) plate and position sensitive photo-multiplier tubes [J].
Jeong, MH ;
Choi, Y ;
Chung, YH ;
Song, TY ;
Jung, JH ;
Hong, KJ ;
Min, BJ ;
Choe, YS ;
Lee, KH ;
Kim, BT .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (21) :4961-4970
[6]   Cardiovascular risk - through the ages [J].
Kastelein, J .
ATHEROSCLEROSIS SUPPLEMENTS, 2004, 5 (02) :1-2
[7]   Development of a miniature scintillation camera using an NaI(Tl) scintillator and PSPMT for scintimammography [J].
Kim, JH ;
Choi, Y ;
Joo, KS ;
Sihn, BS ;
Chong, JW ;
Kim, SE ;
Lee, KH ;
Choe, YS ;
Kim, BT .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :3481-3488
[8]  
LACKAS C, 2004, P 2004 NUCL SCI S ME, V6, P3893
[9]   A flat-panel detector based micro-CT system: performance evaluation for small-animal imaging [J].
Lee, SC ;
Kim, HK ;
Chun, IK ;
Cho, MH ;
Lee, SY ;
Cho, MH .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (24) :4173-4185
[10]   AMIDE: A Free Software Tool for Multimodality Medical Image Analysis [J].
Loening, Andreas Markus ;
Gambhir, Sanjiv Sam .
Molecular Imaging, 2003, 2 (03) :131-137