Development of Needle Trap Technology for On-Site Determinations: Active and Passive Sampling

被引:36
作者
Asl-Hariri, Saba [1 ]
Gomez-Rios, German A. [1 ]
Gionfriddo, Emanuela [1 ]
Dawes, Peter [2 ]
Pawliszyn, Janusz [1 ]
机构
[1] Univ Waterloo, Dept Chem, Waterloo, ON N2L 3G1, Canada
[2] SGE Analyt Sci Pty Ltd, Ringwood, Vic 3134, Australia
关键词
SOLID-PHASE MICROEXTRACTION; IN-VIVO; SPME; DEVICE; OPTIMIZATION; MS;
D O I
10.1021/ac500801v
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This study presents a thorough evaluation of new prototypes of extended tip needle trap devices (NT), as well as their application to in situ sampling of biological emissions and active/passive on-site sampling of indoor air. A new NT prototype was constructed with a side hole above the sorbent and an extended tip that fits inside the restriction of the narrow neck liner to increase desorption efficiency. New prototype needles were initially packed with divinylbenzene particles at SGE Analytical Science for the purpose of studying biogenic emissions of pine trees. Prior to their final application, they were evaluated in terms of robustness after multiple use (n > 10), as well as amount extracted of volatile organic compounds (VOCs). An ANOVA test for all the probes showed that at a 95% level of confidence, there were not statistical differences observed among the 9 NTs tested. In addition, the needles were also packed in laboratory with synthesized highly cross-linked PDMS as a frit to immobilize carboxen (Car) particles for spot sampling. For passive sampling, the needles were packed with Car particles embedded in PDMS to simplify calculations in passive mode. The use of NTs as spot samplers, as well as a passive sampler under controlled conditions in the laboratoryyielded a relative standard deviation of less than 15%. Finally, a new, reusable and readily deployable penlike diffusive sampler for needle traps (PDS-NT) was built and tested. Application of the PDS-NT in combination with NT-spot sampling toward the analysis of indoor air in a polymer synthesis laboratory showed good agreement between both techniques for the analyte studied, yielding averages of 0.03 and 0.025 ng/mL of toluene, respectively.
引用
收藏
页码:5889 / 5897
页数:9
相关论文
共 34 条
[1]   Diffusive sampling of methylene chloride with solid phase microextraction [J].
Chen, Cheng-Yao ;
Hsiech, Chunming ;
Lin, Jia-Ming .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1137 (02) :138-144
[2]   Kinetics and the on-site application of standards in a solid-phase microextration fiber [J].
Chen, Y ;
Pawliszyn, J .
ANALYTICAL CHEMISTRY, 2004, 76 (19) :5807-5815
[3]   Calibration for on-site analysis of hydrocarbons in aqueous and gaseous samples using solid-phase microextraction [J].
Chen, Y ;
Koziel, JA ;
Pawliszyn, J .
ANALYTICAL CHEMISTRY, 2003, 75 (23) :6485-6493
[4]   A NEW METHOD FOR PREDICTION OF BINARY GAS-PHASE DIFFUSION COEFFECIENTS [J].
FULLER, EN ;
SCHETTLE, PD ;
GIDDINGS, JC .
INDUSTRIAL AND ENGINEERING CHEMISTRY, 1966, 58 (05) :19-+
[5]   Development of a new in-vial standard gas system for calibrating solid-phase microextraction in high-throughput and on-site applications [J].
Gomez-Rios, German A. ;
Reyes-Garces, Nathaly ;
Pawliszyn, Janusz .
JOURNAL OF SEPARATION SCIENCE, 2013, 36 (17) :2939-2945
[6]   Development and application of a needle trap device for time-weighted average diffusive sampling [J].
Gong, Ying ;
Eom, In-Yong ;
Lou, Da-Wei ;
Hein, Dietmar ;
Pawliszyn, Janusz .
ANALYTICAL CHEMISTRY, 2008, 80 (19) :7275-7282
[7]   Passive sampling [J].
Górecki, T ;
Namiesnik, J .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2002, 21 (04) :276-291
[8]   Porous crosslinked PDMS-microchannels coatings [J].
Juchniewicz, Marcin ;
Stadnik, Dorota ;
Biesiada, Krzysztof ;
Olszyna, Andrzej ;
Chudy, Michal ;
Brzozka, Zbigniew ;
Dybko, Artur .
SENSORS AND ACTUATORS B-CHEMICAL, 2007, 126 (01) :68-72
[9]   Sampling and analysis of airborne particulate matter and aerosols using in-needle trap and SPME fiber devices [J].
Koziel, JA ;
Odziemkowski, M ;
Pawliszyn, J .
ANALYTICAL CHEMISTRY, 2001, 73 (01) :47-54
[10]   On-site monitoring of biogenic emissions from Eucalyptus dunnii leaves using membrane extraction with sorbent interface combined with a portable gas chromatograph system [J].
Liu, XY ;
Pawliszyn, R ;
Wang, LM ;
Pawliszyn, J .
ANALYST, 2004, 129 (01) :55-62