A 10 cm x 10 cm CdTe Spectroscopic Imaging Detector based on the HEXITEC ASIC

被引:32
|
作者
Wilson, M. D. [1 ]
Dummott, L. [2 ]
Duarte, D. D. [1 ,2 ]
Green, F. H. [2 ]
Pani, S. [2 ]
Schneider, A. [1 ]
Scuffham, J. W. [2 ,3 ]
Seller, P. [1 ]
Veale, M. C. [1 ]
机构
[1] Sci & Technol Facil Council, Rutherford Appleton Lab, Didcot OX11 0QX, Oxon, England
[2] Univ Surrey, Dept Phys, Guildford GU2 7XH, Surrey, England
[3] Royal Surrey Cty Hosp, NHS Fdn Trust, Dept Nucl Med, Guildford GU2 7XX, Surrey, England
来源
JOURNAL OF INSTRUMENTATION | 2015年 / 10卷
基金
英国工程与自然科学研究理事会;
关键词
X-ray detectors; Imaging spectroscopy; Pixelated detectors and associated VLSI electronics;
D O I
10.1088/1748-0221/10/10/P10011
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The 250 mu m pitch 80x80 pixel HEXITEC detector systems have shown that spectroscopic imaging with an energy resolution of < 1 keV FWHM per pixel can be readily achieved in the range of 5-200 keV with Al-pixel CdTe biased to -500V. This level of spectroscopic imaging has a variety of applications but the ability to produce large area detectors remains a barrier to the adoption of this technology. The limited size of ASICs and defect free CdTe wafers dictates that building large area monolithic detectors is not presently a viable option. A 3-side buttable detector module has been developed to cover large areas with arrays of smaller detectors. The detector modules are 20.35 x 20.45 mm with CdTe bump bonded to the HEXITEC ASIC with coverage up to the edge of the module on three sides. The fourth side has a space of 3 mm to allow I/O wire bonds to be made between the ASIC and the edge of a PCB that routes the signals to a connector underneath the active area of the module. The detector modules have been assembled in rows of five modules with a dead space of 170 mu m between each module. Five rows of modules have been assembled in a staggered height array where the wire bonds of one row of modules are covered by the active detector area of a neighboring row. A data acquisition system has been developed to digitise, store and output the 24 Gbit/s data that is generated by the array. The maximum bias magnitude that could be applied to the CdTe detectors from the common voltage source was limited by the worst performing detector module. In this array of detectors a bias of -400V was used and the detector modules had 93 % of pixels with better than 1.2 keV FWHM at 59.5 keV. An example of K-edge enhanced imaging for mammography was demonstrated. Subtracting images from the events directly above and below the K-edge of the Iodine contrast agent was able to extract the Iodine information from the image of a breast phantom and improve the contrast of the images. This is just one example where the energy spectrum per pixel can be used to develop new and improve existing X-ray imaging techniques.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Research and Development of 10 cm×10 cm Bulk Micromegas Detector
    Li P.
    Zhou J.
    Chen L.
    Zhao M.
    Zhi Y.
    Liu W.
    Jia S.
    Zhang Y.
    Hu S.
    Yu W.
    Li X.
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2020, 54 (06): : 963 - 967
  • [2] 10 cm x 10 cm Single Gas Electron Multiplier (GEM) X-ray Fluorescence Detector for Dilute Elements
    Shaban, E. H.
    Siddons, D. P.
    Seifu, D.
    17TH PAN-AMERICAN SYNCHROTRON RADIATION INSTRUMENTATION CONFERENCE SRI2013, 2014, 493
  • [3] Gains, Uniformity and Signal Sharing in XY Readouts of the 10 cm x 10 cm Gas Electron Multiplier (GEM) Detector
    Rittirong, Anawat
    Saenboonruang, Kiadtisak
    JOURNAL OF PHYSICAL SCIENCE, 2018, 29 (01) : 121 - 132
  • [4] 128 Channels of Multi-gigasample-per-second Waveform Sampling and Digitization in a 10 cm x 10 cm x 8 cm Package
    Andrew, Matthew
    Lim, Chester
    Nishimura, Kurtis
    Ridley, Louis
    Varner, Gary
    2012 18TH IEEE-NPSS REAL TIME CONFERENCE (RT), 2012,
  • [5] Force and moment calibration for NCKU 10 cm x 10 cm magnetic suspension wind tunnel
    Lin, CE
    Yang, CK
    IMTC/97 - IEEE INSTRUMENTATION & MEASUREMENT TECHNOLOGY CONFERENCE: SENSING, PROCESSING, NETWORKING, PROCEEDINGS VOLS 1 AND 2, 1997, : 1060 - 1065
  • [6] Effects of High Gas Flow Rates on the Standard 10 cm x 10 cm GEM Prototype
    Saenboonruang, Kiadtisak
    Kumphiranon, Piyakul
    Kulasri, Kittipong
    Rittirong, Anawat
    CHIANG MAI JOURNAL OF SCIENCE, 2016, 43 (04): : 876 - 883
  • [7] Experimental determination of kQ factors for cylindrical ionization chambers in 10 cm x 10 cm and 3 cm x 3 cm photon beams from 4 MV to 25 MV
    Krauss, A.
    Kapsch, R. P.
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (15): : 4227 - 4246
  • [8] HEXITEC 2 x 2 tiled hard X-ray spectroscopic imaging detector system
    Jowitt, L.
    Wilson, M.
    Seller, P.
    Angelsen, C.
    Wheater, R. M.
    Cline, B.
    Schoene, D.
    Lauba, F.
    Goede, M.
    Ball, R.
    Verhoeven, M.
    Gottseleben, G.
    Boone, M. N.
    Van Assche, F.
    Veale, M. C.
    JOURNAL OF INSTRUMENTATION, 2022, 17 (01)
  • [9] 16.0% efficiency of large area (10 cm x 10 cm) thin film polycrystalline silicon solar cell
    Morikawa, H
    Nishimoto, Y
    Naomoto, H
    Kawama, Y
    Takami, A
    Arimoto, S
    Ishihara, T
    Namba, K
    SOLAR ENERGY MATERIALS AND SOLAR CELLS, 1998, 53 (1-2) : 23 - 28
  • [10] Fabrication and performance evaluation of 3-cell SOFC stack based on planar 10 cm x 10 cm anode-supported cells
    Jung, H. Y.
    Choi, S. -H.
    Kim, H.
    Son, J. -W.
    Kim, J.
    Lee, H. -W.
    Lee, J. -H.
    JOURNAL OF POWER SOURCES, 2006, 159 (01) : 478 - 483