Diffusion and Dispersion in Tubes in Supercritical Fluid Chromatography Using Sub-2 μm Packings

被引:8
作者
Berger, Terry A. [1 ]
机构
[1] SFC Solut Inc, Englewood, FL 34224 USA
关键词
Supercritical fluid chromatography (SFC); Sub-2 mu m packing; Diffusion coefficients; Dm; D-m versus pressure and % methanol; Connector tubing variance; Deviation from poiseuille flow; KINETIC PLOT METHOD; M PARTICLES; LIQUID-CHROMATOGRAPHY; COLUMN; PERFORMANCE; PRESSURE; SPEED; COEFFICIENTS; LIMITATIONS; EFFICIENCY;
D O I
10.1007/s10337-020-03996-8
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Detailed plots of diffusion coefficients, D-m, versus % methanol in CO2 and pressure, P, are presented. Using sub-2 mu m particle columns with small ID connector tubing make for large pressure drops, Delta P, across both the system and the column. This creates low average D-m, and large gradients in D-m, slowing the chromatography. The Delta P with 75 mu m ID tubing varies non-linearly with flow creating U shaped plots of reduced plate height versus flow rate. Plumbing does but should never change h versus F plots. Using larger ID tubes to estimate the extra-column variance sigma(2)(e-c) of smaller tubes, by keeping the aspect ratio (L/r) constant, assumes the same constant D-m in both tubes, which is not true in SFC. Contrary to prediction, calculated tube plate count, N, indicates connector tubing is more dispersive in SFC compared to high performance liquid chromatography (HPLC). Current supercritical fluid chromatographs (SFC) have up to 20 x higher variance than appropriate for use with 2.1 and 3 mm columns <= 100 mm long, packed with sub-2 mu m particles. Lower volume UV detector flow cells are needed for use with sub-2 mu m particles. The modular nature of instruments makes it difficult to further minimize tube lengths while retaining column temperature control. Very small ID tubes, like 75 mu m should probably be avoided. It remains difficult to characterize and minimize sigma(2)(e-c).
引用
收藏
页码:167 / 177
页数:11
相关论文
共 28 条
[1]   ON THE DISPERSION OF A SOLUTE IN A FLUID FLOWING THROUGH A TUBE [J].
ARIS, R .
PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1956, 235 (1200) :67-77
[2]   DISPERSION OF PEAKS BY SHORT STRAIGHT OPEN TUBES IN LIQUID-CHROMATOGRAPHY SYSTEMS [J].
ATWOOD, JG ;
GOLAY, MJE .
JOURNAL OF CHROMATOGRAPHY, 1981, 218 (1-3) :97-122
[3]   Reduced Plate Height of 1.65 on a 20 x 3 mm Column Packed with 1.8 μm Particles in Supercritical Fluid Chromatography (SFC) [J].
Berger, Terry A. .
CHROMATOGRAPHIA, 2019, 82 (06) :971-974
[4]   High-Speed, High-Efficiency Achiral SFC on a 3x20-mm Column Packed with 1.8-μm Particles Facilitated by a Low-Dispersion Chromatograph [J].
Berger, Terry A. .
CHROMATOGRAPHIA, 2019, 82 (02) :537-542
[5]   Kinetic performance of a 50 mm long 1.8 μm chiral column in supercritical fluid chromatography [J].
Berger, Terry A. .
JOURNAL OF CHROMATOGRAPHY A, 2016, 1459 :136-144
[6]   Instrument modifications that produced reduced plate heights &lt;2 with sub-2 μm particles and 95% of theoretical efficiency at k=2 in supercritical fluid chromatography [J].
Berger, Terry A. .
JOURNAL OF CHROMATOGRAPHY A, 2016, 1444 :129-144
[7]   Demonstration of High Speeds with Low Pressure Drops Using 1.8 μm Particles in SFC [J].
Berger, Terry A. .
CHROMATOGRAPHIA, 2010, 72 (7-8) :597-602
[8]   VARIANCE OF A ZONE MIGRATING IN A NONUNIFORM TIME-INVARIANT LINEAR MEDIUM [J].
BLUMBERG, LM ;
BERGER, TA .
JOURNAL OF CHROMATOGRAPHY, 1992, 596 (01) :1-13
[9]   Understanding and diminishing the extra-column band broadening effects in supercritical fluid chromatography [J].
De Pauw, Ruben ;
Shoykhet , Konstantin ;
Desmet, Gert ;
Broeckhoven, Ken .
JOURNAL OF CHROMATOGRAPHY A, 2015, 1403 :132-137
[10]   Exploring the speed-resolution limits of supercritical fluid chromatography at ultra-high pressures [J].
De Pauw, Ruben ;
Shoykhet , Konstantin ;
Desmet, Gert ;
Broeckhoven, Ken .
JOURNAL OF CHROMATOGRAPHY A, 2014, 1374 :247-253