Removal of the gadolinium interference from the measurement of selenium in human serum by use of collision cell quadrupole inductively coupled plasma mass spectrometry (Q-ICP-MS)

被引:15
作者
Harrington, Chris F. [1 ]
Walter, Alan [1 ]
Nelms, Simon [2 ]
Taylor, Andrew [1 ]
机构
[1] Univ Surrey, Fac Hlth & Med Sci, Trace Element Lab, Supra Reg Assay Serv, Guildford GU2 7XH, Surrey, England
[2] Thermo Fisher Sci, Hemel Hempstead, Herts, England
关键词
mass spectrometry; instrumentation; laboratory methods; Analytical systems; GADOBENATE DIMEGLUMINE; SAFETY; LIVER;
D O I
10.1177/0004563213504386
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background Measurement of selenium in serum is an important clinical biomarker of nutritional status. The presence of gadolinium (Gd) in samples following administration of the contrast agents used for magnetic resonance imaging (MRI) results in a significant positive bias when using quadrupole inductively coupled plasma mass spectrometry (Q-ICP-MS). Methods Three instrumental set-ups were assessed: standard mode with no collision gas and collision cell mode with either a hydrogen:helium mixture or hydrogen. The effect of Gd on the selenium (Se) signal was assessed using external quality assurance (EQA) specimens and internal quality control (IQC) materials, both unspiked and spiked with Gd. Serum previously shown to contain high concentrations of Gd-containing contrast agents were also analysed. Results Recoveries of Se in the spiked compared to the unspiked samples were: between 500% and 1300% using standard mode; 100% and 29,000% using collision cell mode with hydrogen:helium mixture; and between 99% and 103% using hydrogen. The use of H-2 in the collision cell provided accurate results, indicating that the charge exchange reaction (CER) of Gd2+ with H-2 removes this interference. Analysis of patient serum known to contain the Gd contrast agent using the method gave results within the selenium reference range (adults 0.89-1.65 mu mol/L). The presence of Gd, as low as 0.2 mg/L, in serum samples causes a positive interference on the measurement of Se by ICP-MS. Conclusions Using a CER mode with pure H-2 in the collision cell it was possible to fully remove the interference due to Gd2+ from the signal for Se.
引用
收藏
页码:386 / 391
页数:6
相关论文
共 7 条
[1]   Effects of selenium supplementation for cancer prevention in patients with carcinoma of the skin a randomized controlled trial - A randomized controlled trial [J].
Clark, LC ;
Combs, GF ;
Turnbull, BW ;
Slate, EH ;
Chalker, DK ;
Chow, J ;
Davis, LS ;
Glover, RA ;
Graham, GF ;
Gross, EG ;
Krongrad, A ;
Lesher, JL ;
Park, HK ;
Sanders, BB ;
Smith, CL ;
Taylor, JR .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (24) :1957-1963
[2]  
Davies BE, INVEST RADIOL, V200, P299
[3]   Clinical and biological consequences of transmetallation induced by contrast agents for magnetic resonance imaging:: a review [J].
Idee, Jean-Marc ;
Port, Marc ;
Raynal, Isabelle ;
Schaefer, Michel ;
Le Greneur, Soizic ;
Corot, Claire .
FUNDAMENTAL & CLINICAL PHARMACOLOGY, 2006, 20 (06) :563-576
[4]   Safety, tolerance, biodistribution, and MR imaging enhancement of the liver with gadobenate dimeglumine: Results of clinical pharmacologic and pilot imaging studies in nonpatient and patient volunteers [J].
Spinazzi, A ;
Lorusso, V ;
Pirovano, G ;
Kirchin, M .
ACADEMIC RADIOLOGY, 1999, 6 (05) :282-291
[5]   Safety and pharmacokinetic profile of gadobenate dimeglumine in subjects with renal impairment [J].
Swan, SK ;
Lambrecht, LJ ;
Townsend, R ;
Davies, BE ;
McCloud, S ;
Parker, JR ;
Bensel, K ;
LaFrance, ND .
INVESTIGATIVE RADIOLOGY, 1999, 34 (07) :443-448
[6]   Detection and monitoring of disorders of essential trace elements [J].
Taylor, A .
ANNALS OF CLINICAL BIOCHEMISTRY, 1996, 33 :486-510
[7]   Interference of gadolinium on the measurement of selenium in human serum by inductively coupled plasma-quadrupole mass spectrometry [J].
Walter, Alan ;
Nelms, Simon ;
Harrington, Chris F. ;
Taylor, Andrew .
ANNALS OF CLINICAL BIOCHEMISTRY, 2011, 48 :176-177