Biomechanical qualities of threaded acetabular cups

被引:14
作者
Klanke, J [1 ]
Partenheimer, A [1 ]
Westermann, K [1 ]
机构
[1] Nordstadt Hosp, Dept Traumatol, D-30167 Hannover, Germany
关键词
D O I
10.1007/s00264-002-0368-y
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We examined 25 different threaded cups using a specific biomechanical procedure. Great differences between the tested cups concerning their biomechanical parameters could be found. There is some evidence that the design of several cups might be inappropriate for stable acetabular fixation, and this may be the reason for early loosening. The new procedure is the first to allow biomechanical assessment of threaded cups by using quantified data, which correspond to long-term stability of the cup.
引用
收藏
页码:278 / 282
页数:5
相关论文
共 14 条
[1]   Survivorship analysis of 111 noncemented acetabular screw cups (Mecron) - A follow-up study [J].
Bertram, R ;
Muller, RT .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1998, 136 (04) :317-320
[2]   Survivorship of rough-surfaced threaded acetabular cups - 382 consecutive primary Zweymuller cups followed for 0.2-12 years [J].
Delaunay, CP ;
Kapandji, AI .
ACTA ORTHOPAEDICA SCANDINAVICA, 1998, 69 (04) :379-383
[3]   The cementless PM hip arthroplasty - Four-to-seven-year results [J].
Dickob, M ;
Martini, T .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1996, 78B (02) :195-199
[4]  
GEBAUER D, 1989, BIOMED TECHNIK, V35, P15
[5]  
HARRIS WH, 1983, CLIN ORTHOP RELAT R, P7
[6]  
KLANKE J, 1999, SICOT 21 TRIENN WORL, P74
[7]   Fixation of acetabular cups without cement in total hip arthroplasty - A comparison of three different implant surfaces at a minimum duration of follow-up of five years [J].
Manley, MT ;
Capello, WN ;
D'Antonio, JA ;
Edidin, AA ;
Geesink, RGT .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (08) :1175-1185
[8]  
PERREN SM, 1983, ZEMENTLOSE FIXATION
[9]  
PILLIAR RM, 1986, CLIN ORTHOP RELAT R, P108
[10]   Factors influencing initial stability of uncemented acetabular components. An in-vitro study [J].
Pitto, RP ;
Bohner, J ;
Hofmeister, V .
BIOMEDIZINISCHE TECHNIK, 1997, 42 (12) :363-368