Determination of Cd in food using an electrothermal vaporization capacitively coupled plasma microtorch optical emission microspectrometer: Compliance with European legislation and comparison with graphite furnace atomic absorption spectrometry

被引:17
作者
Butaciu, Sinziana [1 ]
Frentiu, Tiberiu [1 ]
Senila, Marin [2 ]
Darvasi, Eugen [1 ]
Cadar, Sergiu [2 ]
Ponta, Michaela [1 ]
Petreus, Dorin [3 ]
Etz, Radu [3 ]
Frentiu, Maria [2 ]
机构
[1] Univ Babes Bolyai, Fac Chem & Chem Engn, Cluj Napoca 400028, Romania
[2] INCDO INOE 2000 Natl Inst Res & Dev Optoelect Buc, Res Inst Analyt Instrumentat, Cluj Napoca 400293, Romania
[3] Tech Univ Cluj Napoca, Fac Elect Telecommun & Informat Technol, Cluj Napoca 400027, Romania
关键词
Cadmium in food; Capacitively coupled plasma microtorch; Optical emission spectrometry; Graphite furnace atomic absorption spectrometry; European legislation; BIODEGRADABLE MATERIALS; MERCURY DETERMINATION; SAMPLE INTRODUCTION; WATER SAMPLES; CADMIUM; LEAD; PRECONCENTRATION; MICROEXTRACTION; NANOPARTICLES; MICROPLASMA;
D O I
10.1016/j.foodcont.2015.09.040
中图分类号
TS2 [食品工业];
学科分类号
0832 ;
摘要
A precise, accurate and sensitive method based on electrothermal vaporization capacitively coupled plasma microtorch optical emission spectrometry for Cd determination in various dietary products was developed. The instrumentation is inexpensive as it includes a low power (15 W) and low argon consumption (150 ml min(-1)) microplasma connected to a small-sized Rh electrothermal vaporization device and a low resolution microspectrometer. The sample was subjected to microwave assisted digestion in HNO3-H2O2 mixture and Cd was determined in 10 mu l sample without preconcentration. The standard addition method was used to compensate for the non-spectral matrix effects. The new method was in-house validated by analyzing certified reference materials and test dietary samples for which the Cd maximum level was established in Commission Regulation 488/2014/EU. The figures of merit of the proposed method and the experimental results were compared to those obtained in graphite furnace atomic absorption spectrometry as standardized method and discussed in relation with the requirements in the Commission Decision 2002/657/EC and Commission Regulations 2011/836/EU and 2007/333/EC. The limit of detection in fresh sample (0.001-0.009 mg kg(-1)), limit of quantification (0.002-0.018 mg kg(-1)), recovery (103 +/- 15%), trueness (-3.0)-(+7.3)% and precision (0.8-14.0%) were similar to those found in the reference method and comply with the minimal criteria imposed to a method to be used in official laboratories for Cd determination in food. Precision for Cd determination close to maximum admitted levels or higher was of 0.8-6.0%, while for lower concentration in the range 4.0-14.0%. The Bland and Altman statistical test allowed successful evaluation of the new method as an alternative to the well established graphite-furnace atomic absorption spectrometry. The proposed method has analytical potential and is easy to run, while the full miniaturized instrumentation could be considered as a useful tool for the future as an alternative to the traditional instrumentation. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:227 / 234
页数:8
相关论文
共 37 条
[31]   Development and validation of an analytical method for the determination of arsenic, cadmium and lead content in powdered infant formula by means of quadrupole Inductively Coupled Plasma Mass Spectrometry [J].
Sorbo, A. ;
Turco, A. C. ;
Di Gregorio, M. ;
Ciaralli, L. .
FOOD CONTROL, 2014, 44 :159-165
[32]   Determination of Cadmium in Fruit and Vegetables by Ionic Liquid Magnetic Microextraction and Flame Atomic Absorption Spectrometry [J].
Soylak, Mustafa ;
Yilmaz, Erkan .
ANALYTICAL LETTERS, 2015, 48 (03) :464-476
[33]  
Taylor A, 2015, J ANAL ATOM SPECTROM, V30, P542, DOI [10.1039/c5ja90001h, 10.1039/C5JA90001H]
[34]  
USEPA, 2000, RISK BAS CONC TABL, P2000
[35]   Microplasma Technology and Its Applications in Analytical Chemistry [J].
Yuan, Xin ;
Tang, Jie ;
Duan, Yixiang .
APPLIED SPECTROSCOPY REVIEWS, 2011, 46 (07) :581-605
[36]   Simple and robust method for lithium traces determination in drinking water by atomic emission using low-power capacitively coupled plasma microtorch and microspectrometer [J].
Zsigmond, Andreea R. ;
Frentiu, Tiberiu ;
Ponta, Michaela ;
Frentiu, Maria ;
Petreus, Dorin .
FOOD CHEMISTRY, 2013, 141 (04) :3621-3626
[37]   Methodological evaluation of method for dietary heavy metal intake [J].
Zukowska, J. ;
Biziuk, M. .
JOURNAL OF FOOD SCIENCE, 2008, 73 (02) :R21-R29