Effect of Structural Remodeling (Retraction and Recoil) of the Pipeline Embolization Device on Aneurysm Occlusion Rate

被引:16
作者
Jou, L. -D. [1 ]
Mitchell, B. D. [2 ]
Shaltoni, H. M. [1 ]
Mawad, M. E. [1 ]
机构
[1] Baylor Coll Med, Dept Radiol, Houston, TX 77030 USA
[2] Baylor Coll Med, Dept Neurosurg, Houston, TX 77030 USA
关键词
ACUTE STENT RECOIL; FLAT-PANEL-CT; INTRACRANIAL ANEURYSMS; FLOW-DIVERSION; SINGLE-CENTER; IN-VIVO; IMMEDIATE;
D O I
10.3174/ajnr.A3920
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: During endovascular treatment of unruptured aneurysms with the Pipeline Embolization Device, an oversized device is often selected to achieve better wall apposition; however, this device oversizing could be related to overelongation and possible delayed enlargement of the stented region. The purpose of this study is to investigate the relationship between oversize and treatment outcome. MATERIALS AND METHODS: The DynaCT images of 14 aneurysms treated by a single Pipeline Embolization Device were retrospectively analyzed. 3D images of the deployed device were compared with those acquired at the 6-month follow-up for qualitative and quantitative evaluation. The diameter and length of the Pipeline Embolization Device were measured at both time points and compared for determination of the device changes. RESULTS: Structural changes of the device have been observed, and it was found that the Pipeline Ernbolization Device influences the vessel curvature in some cases. On average, it increases its diameter by 0.23 mm and decreases its length by 2.88 mm within 6 months of initial deployment. Excessive elongation beyond its nominal length is correlated with a lower aneurysm occlusion rate at the 6-month follow-up. CONCLUSIONS: Not only does a Pipeline Embolization Device reconstruct the aneurysm and parent artery, but its entire structure goes through a gradual remodeling process. The relative deformation between the device and the artery indicates suboptimal wall apposition. Device oversizing does not have a direct effect on shortening or recoil. The aneurysm occlusion rate, however, is lowered by overelongation of the Pipeline Embolization Device.
引用
收藏
页码:1772 / 1778
页数:7
相关论文
共 28 条
[1]   Stent expansion: a combination of delivery balloon underexpansion and acute stent recoil reduces predicted stent diameter irrespective of reference vessel size [J].
Aziz, Shahid ;
Morris, John L. ;
Perry, Raphael A. ;
Stables, Rodney H. .
HEART, 2007, 93 (12) :1562-1566
[2]   Effects of overexpansion on stents' recoil, symmetry/asymmetry, and neointimal hyperplasia in aortas of hypercholesterolemic rabbits [J].
Berrocal, Daniel H. ;
Gonzalez, German E. ;
Fernandez, Alejandro ;
Perez, Susana ;
Wilensky, Luciana ;
Morales, Celina ;
Grinfeld, Liliana ;
Gelpi, Ricardo J. .
CARDIOVASCULAR PATHOLOGY, 2008, 17 (05) :289-296
[3]   Stents and flow diverters in the treatment of aneurysms: device deformation in vivo may alter porosity and impact efficacy [J].
Bing, Fabrice ;
Darsaut, Tim E. ;
Salazkin, Igor ;
Makoyeva, Alina ;
Gevry, Guylaine ;
Raymond, Jean .
NEURORADIOLOGY, 2013, 55 (01) :85-92
[4]   Spontaneous Delayed Migration/Shortening of the Pipeline Embolization Device: Report of 5 Cases [J].
Chalouhi, N. ;
Tjoumakaris, S. I. ;
Gonzalez, L. F. ;
Hasan, D. ;
Pema, P. J. ;
Gould, G. ;
Rosenwasser, R. H. ;
Jabbour, P. M. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2013, 34 (12) :2326-2330
[5]   Evaluation of stent visibility by flat panel detector CT in patients treated for intracranial aneurysms [J].
Clarencon, Frederic ;
Piotin, Michel ;
Pistocchi, Silvia ;
Babic, Drazenko ;
Blanc, Raphael .
NEURORADIOLOGY, 2012, 54 (10) :1121-1125
[6]   Immediate procedural outcomes in 35 consecutive pipeline embolization cases: a single-center, single-user experience [J].
Colby, Geoffrey P. ;
Lin, Li-Mel ;
Gomez, Juan F. ;
Pau, Alexandra R. ;
Huang, Judy ;
Tamargo, Rafael J. ;
Coon, Alexander L. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2013, 5 (03) :237-246
[7]   Marked decrease in coil and stent utilization following introduction of flow diversion technology [J].
Crobeddu, Emanuela ;
Lanzino, Giuseppe ;
Kallmes, David F. ;
Cloft, Harry J. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2013, 5 (04) :351-353
[8]   Delayed Ipsilateral Parenchymal Hemorrhage Following Flow Diversion for the Treatment of Anterior Circulation Aneurysms [J].
Cruz, J. P. ;
Chow, M. ;
O'Kelly, C. ;
Marotta, B. ;
Spears, J. ;
Montanera, W. ;
Fiorella, D. ;
Marotta, T. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2012, 33 (04) :603-608
[9]  
Danzi GB, 2001, CATHETER CARDIO INTE, V52, P147, DOI 10.1002/1522-726X(200102)52:2<147::AID-CCD1038>3.0.CO
[10]  
2-B