Head-mounted active noise control system to achieve speech communication

被引:0
作者
Miyazaki, Nobuhiro [1 ]
Yamakawa, Kohei [1 ]
Kajikawa, Yoshinobu [1 ]
机构
[1] Kansai Univ, Fac Engn Sci, Suita, Osaka 5648680, Japan
来源
2013 ASIA-PACIFIC SIGNAL AND INFORMATION PROCESSING ASSOCIATION ANNUAL SUMMIT AND CONFERENCE (APSIPA) | 2013年
关键词
AUDIO; IMPLEMENTATION; PERFORMANCE; PREDICTION;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this paper, we propose a head-mounted active noise control (ANC) system with speech communication. Magnetic resonance imaging (MRI) device is one of medical equipment. Recently, MRI device is utilized for microwave coagulation therapy. However, MRI device generates a serious noise (over 100 dB SPL). Hence, surgeons and other medical staff are exposed to the large MR noise for many hours and cannot verbally communicate with each other. We have therefore proposed a head-mounted ANC system for reducing MR noise and realizing verbal communication under such a loud noise environment. However, MRI device is generally controlled by the operator outside the MRI room. Hence, speech communication between inside and outside the room is needed. We therefore integrate the speech communication function with the head-mounted ANC system. Concretely, the error microphones and secondary loudspeakers are also used as an interface to realize the speech communication. In this case, the outside voice may be returned through the error microphone, so an audio-integrated ANC system based on the echo cancellation is utilized. Linear prediction filter is also utilized for separating the inside voice from residual noise. In this paper, we demonstrate the validity of the proposed ANC system through some noise reduction experiments and subjective assessment tests on phoneme articulation.
引用
收藏
页数:6
相关论文
共 29 条
[1]  
[Anonymous], 1985, Adaptive signal processing prentice-hall
[2]  
Bouchard M, 1998, J SOUND VIB, V210, P517, DOI 10.1006/jsvi.1997.1297
[3]  
Chakravarthy S., 2007, P IEEE INT C CONS EL
[4]  
Datta A., 1998, Adaptive Internal Model Control
[5]   Active noise control [J].
Elliott, S. J. ;
Nelson, P. A. .
IEEE SIGNAL PROCESSING MAGAZINE, 1993, 10 (04) :12-35
[6]   Performance of feedforward and feedback systems for active control [J].
Elliott, SJ ;
Sutton, TJ .
IEEE TRANSACTIONS ON SPEECH AND AUDIO PROCESSING, 1996, 4 (03) :214-223
[7]   A STABLE ADAPTIVE IMPLEMENTATION OF THE INTERNAL MODEL PRINCIPLE [J].
FENG, G ;
PALANISWAMI, M .
IEEE TRANSACTIONS ON AUTOMATIC CONTROL, 1992, 37 (08) :1220-1225
[8]  
Gan W. S., 2003, P 2003 INT S CIRC SY
[9]   Adaptive feedback active noise control headset: Implementation, evaluation and its extensions [J].
Gan, WS ;
Mitra, S ;
Kuo, SM .
IEEE TRANSACTIONS ON CONSUMER ELECTRONICS, 2005, 51 (03) :975-982
[10]   An integrated audio and active noise control headsets [J].
Gan, WS ;
Kuo, SM .
IEEE TRANSACTIONS ON CONSUMER ELECTRONICS, 2002, 48 (02) :242-247