Patient Radiation Exposure in a Modern, Large-Volume, Pediatric Cardiac Catheterization Laboratory

被引:56
作者
Glatz, Andrew C. [1 ,2 ,3 ]
Patel, Akash [1 ,2 ]
Zhu, Xiaowei [4 ]
Dori, Yoav [1 ,2 ]
Hanna, Brian D. [1 ,2 ]
Gillespie, Matthew J. [1 ,2 ]
Rome, Jonathan J. [1 ,2 ]
机构
[1] Childrens Hosp Philadelphia, Div Cardiol, Philadelphia, PA 19104 USA
[2] Univ Penn, Dept Pediat, Perelman Sch Med, Philadelphia, PA 19104 USA
[3] Univ Penn, Ctr Clin Epidemiol & Biostat, Perelman Sch Med, Philadelphia, PA 19104 USA
[4] Childrens Hosp Philadelphia, Dept Radiol, Philadelphia, PA 19104 USA
关键词
Catheterization; Radiation; Pediatric; Risk factors; CANCER-RISKS; CHILDREN; REDUCTION; MORTALITY; SAFETY; DAMAGE;
D O I
10.1007/s00246-014-0869-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Radiation exposure from pediatric cardiac catheterization may be substantial, although published estimates vary. We sought to report patient radiation dose across a range of diagnostic and interventional cases in a modern, high-volume pediatric catheterization laboratory. We retrospectively reviewed diagnostic and interventional cases performed in our pediatric catheterization laboratory from 1 April 2009 to 30 September 2011 for which radiation usage data were available as reported by the Artis Zee(A (R)) (Siemens Medical Solutions) system. Electrophysiology cases were excluded. Radiation dose was quantified as air kerma dose (mGy) and dose-area product (DAP; mu Gy m(2)). The DAP was converted to an effective dose millisievert (mSv) using the Monte Carlo method. Radiation usage data were available from 2,265 diagnostic and interventional cases with an overall median air kerma dose of 135 mGy [interquartile range (IQR) 59-433], median DAP of 760 mu Gy m(2) (IQR 281-2,810), of which 75 % (IQR 59-90 %) was derived from fluoroscopy, and median effective dose of 6.2 mSv (IQR 2.7-14.1). Air kerma dose from a single camera > 2,000 mGy occurred in 1.8 % of cases. Significant differences in all measures of radiation exposure existed based on procedural and interventional types (p = 0.0001), with interventional cases associated with the highest effective dose after adjusting for patient weight category (p < 0.001). Patient weight, age, fluoroscopy time, and proportional use of digital acquisition were independent predictors of exposure (p a parts per thousand currency sign 0.001; R (2) = 0.59-0.64). In a modern, large-volume pediatric catheterization laboratory, the median effective dose is 6.2 mSv with a wide range of exposure based on patient- and procedure-specific factors. Radiation monitoring is an important component of a pediatric laboratory and further dose reduction strategies are warranted.
引用
收藏
页码:870 / 878
页数:9
相关论文
共 50 条
[31]   A Virtual Cardiac Catheterization Laboratory for Patient Education: The Angiogram Procedure [J].
Wilcocks, K. ;
Halabi, N. ;
Kartick, P. ;
Uribe-Quevedo, A. ;
Chow, C. ;
Kapralos, B. .
2017 8TH INTERNATIONAL CONFERENCE ON INFORMATION, INTELLIGENCE, SYSTEMS & APPLICATIONS (IISA), 2017, :576-579
[32]   Radiation reduction in a modern catheterization laboratory: A single-center experience [J].
Shekhar, Shashank ;
Ajay, Abhishek ;
Agrawal, Ankit ;
Kumar, Anirudh ;
Kaur, Manpreet ;
Isogai, Toshiaki ;
Saad, Anas ;
Abushouk, Abdelrahman, I ;
Abdelfattah, Omar ;
Lak, Hassan ;
Farwati, Medhat ;
Ahuja, Keerat ;
Verma, Beni R. ;
Wunderle, Kevin ;
Ellis, Stephen ;
Khatri, Jaikirshan ;
Ziada, Khaled ;
Krishnaswamy, Amar ;
Kapadia, Samir .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2022, 100 (04) :575-584
[33]   Radiation exposure during cardiac catheterization: Implications for the transradial approach [J].
Komocsi, A. ;
Konyi, A. ;
Kovacs, E. ;
Ungi, I. .
INTERVENTIONAL MEDICINE AND APPLIED SCIENCE, 2010, 2 (03) :105-109
[34]   Reduction in Radiation Dose in a Pediatric Cardiac Catheterization Lab Using the Philips AlluraClarity X-ray System [J].
Sullivan, Patrick M. ;
Harrison, David ;
Badran, Sarah ;
Takao, Cheryl M. ;
Ing, Frank F. .
PEDIATRIC CARDIOLOGY, 2017, 38 (08) :1583-1591
[35]   Use of Smart Technology for Remote Consultation in the Pediatric Cardiac Catheterization Laboratory [J].
Seckeler, Michael D. ;
Gordon, Brent M. ;
Williams, Derek A. ;
Goldstein, Bryan H. .
CONGENITAL HEART DISEASE, 2015, 10 (06) :E288-E294
[36]   Radiation Reduction in the Pediatric Catheterization Laboratory Using a Novel Imaging System [J].
Manu, Stephen ;
Suntharos, Patcharapong ;
Boyle, Gerard J. ;
Wang, Lu ;
Prieto, Lourdes R. .
JOURNAL OF INVASIVE CARDIOLOGY, 2018, 30 (01) :28-33
[37]   Effect of a Real-Time Radiation Monitoring Device on Operator Radiation Exposure During Cardiac Catheterization The Radiation Reduction During Cardiac Catheterization Using Real-Time Monitoring Study [J].
Christopoulos, Georgios ;
Papayannis, Aristotelis C. ;
Alomar, Mohammed ;
Kotsia, Anna ;
Michael, Tesfaldet T. ;
Rangan, Bavana V. ;
Roesle, Michele ;
Shorrock, Deborah ;
Makke, Lorenza ;
Layne, Ronald ;
Grabarkewitz, Rebecca ;
Haagen, Donald ;
Maragkoudakis, Spyros ;
Mohammad, Atif ;
Sarode, Karan ;
Cipher, Daisha J. ;
Chambers, Charles E. ;
Banerjee, Subhash ;
Brilakis, Emmanouil S. .
CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2014, 7 (06) :744-750
[38]   Patient effective doses and radiation risks in cardiac catheterization procedures [J].
Sulieman, A. ;
Abdelrazig, A. ;
Alkhorayef, M. ;
Babikir, E. ;
Al-Mohammed, H. I. ;
Al-Mugren, K. S. ;
Bradley, D. .
RADIATION PHYSICS AND CHEMISTRY, 2019, 157 :54-59
[39]   Incidence of and Risk Factors for Catheterization-Induced Complete Heart Block in the Pediatric Cardiac Catheterization Laboratory [J].
Mah, Douglas Y. ;
Porras, Diego ;
Bergersen, Lisa ;
Marshall, Audrey C. ;
Walsh, Edward P. ;
Triedman, John K. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2014, 7 (01) :127-133
[40]   Assessing the Efficacy of RADPAD Protection Drape in Reducing Radiation Exposure to Operators in the Cardiac Catheterization Laboratory: A Systematic Review and MetaAnalysis [J].
Bahar, Abdul Rasheed ;
Khanal, Resha ;
Hamza, Mohammad ;
Goru, Rohit K. ;
Shafiq, Aimen ;
Haider, Mobeen Z. ;
Basit, Salman Abdul ;
Bahar, Yasemin ;
Umer, Ahmed Muaaz ;
Sattar, Yasar ;
Alraies, M. Chadi .
CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (04)