Improved myocardial tagging contrast in cine balanced SSFP images

被引:18
作者
Ibrahim, El-Sayed H.
Stuber, Matthias
Schar, Michael
Osman, Nael F.
机构
[1] Johns Hopkins Univ, Dept Elect & Comp Engn, Baltimore, MD 21218 USA
[2] Johns Hopkins Univ, Russell H Morgan Dept Radiol & Radiol Sci, Baltimore, MD USA
[3] Philips Med Syst Inc, Cleveland, OH USA
关键词
cardiac MRI; SSFP; CSPAMM tagging; SNR; SPGR;
D O I
10.1002/jmri.20730
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To improve the tag persistence throughout the whole cardiac cycle by providing a constant tag-contrast throughout all the cardiac phases when using balanced steady-state free precession (bSSFP) imaging. Materials and Methods: The flip angles of the imaging radiofrequency pulses were optimized to compensate for the tagging contrast-to-noise ratio (Tag-CNR) fading at later cardiac phases in bSSFP imaging. Complementary spatial modulation of magnetization (CSPAMM) tagging was implemented to improve the Tag-CNR. Numerical simulations were performed to examine the behavior of the Tag-CNR with the proposed method, and to compare the resulting Tag-CNR with that obtained from the more commonly used spoiled gradient echo (SPGR) imaging. A gel phantom, as well as five healthy human volunteers, were scanned on a 1.5T scanner using bSSFP imaging with and without the proposed technique. The phantom was also scanned with SPGR imaging. Results: With the proposed technique, the Tag-CNR remained almost constant during the whole cardiac cycle. Using bSSFP imaging, the Tag-CNR was about double that of SPGR. Conclusion: The tag persistence was significantly improved when the proposed method was applied, with better Tag-CNR during the diastolic cardiac phase. The improved Tag-CNR will support automated tagging analysis and quantification methods.
引用
收藏
页码:1159 / 1167
页数:9
相关论文
共 34 条
[1]   MR IMAGING OF MOTION WITH SPATIAL MODULATION OF MAGNETIZATION [J].
AXEL, L ;
DOUGHERTY, L .
RADIOLOGY, 1989, 171 (03) :841-845
[2]  
Deimling M., 1994, PROC 2 ANN M SMR SAN, P495
[3]   Reduction of transient signal oscillations in True-FISP using a linear flip angle series magnetization preparation [J].
Deshpande, VS ;
Chung, YC ;
Zhang, Q ;
Shea, SM ;
Li, DB .
MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (01) :151-157
[4]  
FAHMY AS, 2005, P 8 ANN M SCMR SAN F
[5]   TRUE MYOCARDIAL MOTION TRACKING [J].
FISCHER, SE ;
MCKINNON, GC ;
SCHEIDEGGER, MB ;
PRINS, W ;
MEIER, D ;
BOESIGER, P .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (04) :401-413
[6]   IMPROVED MYOCARDIAL TAGGING CONTRAST [J].
FISCHER, SE ;
MCKINNON, GC ;
MAIER, SE ;
BOESIGER, P .
MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (02) :191-200
[7]   Starter sequence for steady-state free precession imaging [J].
Foxall, DL .
MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (04) :919-929
[8]   Off-resonance effects in the transient response of SSFP sequences [J].
Ganter, C .
MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (02) :368-375
[9]   Fat-suppressed steady-state free precession imaging using phase detection [J].
Hargreaves, BA ;
Vasanawala, SS ;
Nayak, KS ;
Hu, BS ;
Nishimura, DG .
MAGNETIC RESONANCE IN MEDICINE, 2003, 50 (01) :210-213
[10]   Characterization and reduction of the transient response in steady-state MR imaging [J].
Hargreaves, BA ;
Vasanawala, SS ;
Pauly, JM ;
Nishimura, DG .
MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (01) :149-158