The effect of plate rotation on the stiffness of femoral LISS: A mechanical study

被引:35
作者
Khalafi, Afshin [1 ]
Curtiss, Shane [1 ]
Hazelwood, Scott [1 ]
Wolinsky, Philip [1 ]
机构
[1] Univ Calif Davis, Dept Orthopaed Surg, Sacramento, CA 95817 USA
关键词
Less Invasive Stabilization System; mechanical testing; distal femur fractures; stiffness; modes of failure;
D O I
10.1097/01.bot.0000244996.45127.ad
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: Malposition of the femoral Less Invasive Stabilization System (LISS) plate may alter its biomechanical behavior. This study compares the mechanical stability of "correctly" affixed LISS plates matching the slope of the lateral femoral condyle to "incorrectly" placed LISS plates fixed in external rotation relative to the distal femur. Methods: A fracture gap model was created to simulate a comminuted supracondylar femur fracture (AO/OTA33-A3). Fixation was achieved using two different plate positions: the LISS plate was either placed "correctly" by internally rotating the plate to match the slope of the lateral femoral condyle, or "incorrectly" by externally rotating the plate relative to the distal femur. Following fixation, the constructs were loaded in axial, torsional, and cyclical axial modes in a material testing machine. Main Outcome Measurement: Stiffness in axial and torsional loading; total deformation and irreversible (plastic) deformation in cyclical axial loading. Results: The mean axial stiffness for the correctly placed LISS constructs was 21.5% greater than the externally rotated LISS constructs (62.7 N/mm vs. 49.3 N/mm; P = 0.0007). No significant difference was found in torsional stiffness between the two groups. Cyclical axial loading caused significantly less (P < 0.0001) plastic deformation in the correct group (0.6 mm) compared with externally rotated group (1.3 mm). All the constructs in the incorrect group failed, where failure was defined as a complete closure of the medial fracture gap, prior to completion of the test cycles. Conclusion: Correct positioning of the LISS plate for fixation of distal femur fractures results in improved mechanical stability as reflected by an increased stiffness in axial loading and decreased plastic deformation at the bone-screw interface.
引用
收藏
页码:542 / 546
页数:5
相关论文
共 21 条
[1]   Failure of less invasive stabilization system plates in the distal femur - A report of four cases [J].
Button, G ;
Wolinsky, P ;
Hak, D .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (08) :565-570
[2]  
Cordey J, 1999, INJURY, V30, pS26
[3]   Mechanical validation of whole bone composite femur models [J].
Cristofolini, L ;
Viceconti, M ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 1996, 29 (04) :525-535
[4]   The development of the distal femur Less Invasive Stabilization System (LISS) [J].
Frigg, R ;
Appenzeller, A ;
Christensen, R ;
Frenk, A ;
Gilbert, S ;
Schavan, R .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2001, 32 :24-31
[5]   Structural properties of a new design of composite replicate femurs and tibias [J].
Heiner, AD ;
Brown, TD .
JOURNAL OF BIOMECHANICS, 2001, 34 (06) :773-781
[6]   Biomechanical analysis of supracondylar femoral fractures fixed with modern retrograde intramedullary nails [J].
Hora, N ;
Markel, DC ;
Haynes, A ;
Grimm, MJ .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1999, 13 (08) :539-544
[7]   Supracondylar femur fracture fixation -: Mechanical comparison of the 95° condylar side plate and screw versus 95° angled blade plate [J].
Jaakkola, JI ;
Lundy, DW ;
Moore, T ;
Jones, B ;
Ganey, TM ;
Hutton, WC .
ACTA ORTHOPAEDICA SCANDINAVICA, 2002, 73 (01) :72-76
[8]   Distal femoral fixation: A biomechanical comparison of the standard condylar buttress plate, a locked buttress plate, and the 95-degree blade plate [J].
Koval, KJ ;
Hoehl, JJ ;
Kummer, FJ ;
Simon, JA .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1997, 11 (07) :521-524
[9]   Distal femoral fixation: A laboratory comparison of the 95 degrees plate, antegrade and retrograde inserted reamed intramedullary nails [J].
Koval, KJ ;
Kummer, FJ ;
Bharam, S ;
Chen, DY ;
Halder, S .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1996, 10 (06) :378-382
[10]   Distal femoral fracture fixation utilizing the Less Invasive Stabilization System (LISS): The technique and early results [J].
Kregor, PJ ;
Stannard, J ;
Zlowodzki, M ;
Cole, PA ;
Alonso, J .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2001, 32 :32-47