Image Reconstruction Using OS-EM Method in Cone-beam CT

被引:2
作者
Dong, Baoyu [1 ]
机构
[1] Dalian Jiaotong Univ, Sch Elect & Informat, Dalian, Peoples R China
来源
2008 IEEE INTERNATIONAL SYMPOSIUM ON KNOWLEDGE ACQUISITION AND MODELING WORKSHOP PROCEEDINGS, VOLS 1 AND 2 | 2008年
关键词
OS-EM algorithm; Cone-beam CT; MLE; Image reconstruction;
D O I
10.1109/KAMW.2008.4810436
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Traditional CT reconstructions are limited by many kinds of artifacts, and they give dissatisfactory image. To reduce image noise and artifacts, we propose a statistical iterative method for image reconstruction in cone-beam CT. First the theory of maximum likelihood estimation (MLE) is extended from emission CT to X-ray scan, then an expectation-maximization (EM) formula is deduced for direct reconstruction of cone-beam CT. EM algorithm is an iterative method that can produce good quality reconstruction, but compared with fast and robust FDK algorithm, EM algorithm is computer intensive and convergence slow. In order to accelerate the convergence speed of EM algorithm, ordered subset (OS) is applied in Cone-beam CT. Experimental results with computer simulated data and real CT data show that OS-EM algorithm can provide good quality reconstructions after only a few iterations. In addition, the point spread function of the OS-EM algorithm is analyzed for evaluating the imaging system performance.
引用
收藏
页码:108 / 111
页数:4
相关论文
共 8 条
[1]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[2]  
HERBERT T, 1989, IEEE T MED IMAGING, V8, P194
[3]   ACCELERATED IMAGE-RECONSTRUCTION USING ORDERED SUBSETS OF PROJECTION DATA [J].
HUDSON, HM ;
LARKIN, RS .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) :601-609
[4]   MAXIMUM LIKELIHOOD APPROACH TO EMISSION IMAGE-RECONSTRUCTION FROM PROJECTIONS [J].
ROCKMORE, AJ ;
MACOVSKI, A .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1976, 23 (04) :1428-1432
[5]  
Shepp L A, 1982, IEEE Trans Med Imaging, V1, P113, DOI 10.1109/TMI.1982.4307558
[6]   Performance evaluation of local ROI algorithms for exact ROI reconstruction in spiral cone-beam computed tomography [J].
Sourbelle, K ;
Lauritsch, G ;
Tam, KC ;
Noo, F ;
Kalender, WA .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (03) :697-702
[7]   Iterative deblurring for CT metal artifact reduction [J].
Wang, G ;
Snyder, DL ;
OSullivan, JA ;
Vannier, MW .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1996, 15 (05) :657-664
[8]   Feldkamp-type cone-beam tomography in the wavelet framework [J].
Zhao, SY ;
Wang, G .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (09) :922-929