3D-imaging using mobile low-end C-arm systems

被引:0
作者
Lütjens, J [1 ]
Koppe, R [1 ]
Klotz, E [1 ]
Grass, M [1 ]
Rasche, V [1 ]
机构
[1] Philips Res Labs, Hamburg, Germany
来源
MEDICAL IMAGING 2002: IMAGE PROCESSING, VOL 1-3 | 2002年 / 4684卷
关键词
low-end C-arm systems; cone-beam reconstruction;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
During the last years, three-dimensional X-ray imaging has become a well-established imaging modality, setting the golden standard for spatial resolution in three-dimensional X-ray imaging. Firstly introduced on a motorized C-arm system, it gained benefit from the high spatial resolution of the image intensifier. Using cone-beam reconstruction, it provided fast access to truly three-dimensional imaging with isotropic voxel dimensions. However, the non-rigid mechanics and the image distortion in the image intensifier required dedicated calibration processes and obligated the developers to use the most stable and reliable system in the C-arm device family. The need for system calibration also required the system to be able to reproducibly adjust the C-arm to the pre-calibrated positions, which seemed only possible with the motorized movement of a high-end system. On mobile, non-motorized C-arm systems, which are often used for guiding surgical procedures, however, 3D application has not been feasible due to the non-reproducibility of the mechanical movement. In this paper, first results regarding the feasibility of this approach are presented. The data were acquired on a Philips BV 26 surgical C-arm. This device is fully movable. The "C" arc is adjusted manually.
引用
收藏
页码:901 / 912
页数:12
相关论文
共 14 条
[1]  
Bronstein I.N., 1989, TASCHENBUCH MATH
[2]   A CONE-BEAM RECONSTRUCTION ALGORITHM USING SHIFT-VARIANT FILTERING AND CONE-BEAM BACKPROJECTION [J].
DEFRISE, M ;
CLACK, R .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (01) :186-195
[3]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[4]  
GLASER F, 1999, THESIS U LUBECK
[5]   Three-dimensional reconstruction of high contrast objects using C-arm image intensifier projection data [J].
Grass, M ;
Koppe, R ;
Klotz, E ;
Proksa, R ;
Kuhn, MH ;
Aerts, H ;
Op de Beek, J ;
Kemkers, R .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1999, 23 (06) :311-321
[6]   CUBIC CONVOLUTION INTERPOLATION FOR DIGITAL IMAGE-PROCESSING [J].
KEYS, RG .
IEEE TRANSACTIONS ON ACOUSTICS SPEECH AND SIGNAL PROCESSING, 1981, 29 (06) :1153-1160
[7]  
KOPPE R, 1995, P CAR 95
[8]  
KOPPE R, 1994, 81794 PHIL GMBH FORS
[9]   ON THE COMPARISON OF INTERPOLATION METHODS [J].
MAELAND, E .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1988, 7 (03) :213-217
[10]  
Parker J, 1983, IEEE Trans Med Imaging, V2, P31, DOI 10.1109/TMI.1983.4307610