Novel cation selective exhaustive injection-sweeping procedure for 5-nitroimidazole determination in waters by micellar electrokinetic chromatography using dispersive liquid-liquid microextraction

被引:29
|
作者
Hernandez-Mesa, Maykel [1 ]
Airado-Rodriguez, Diego [1 ]
Cruces-Blanco, Carmen [1 ]
Garcia-Campana, Ana M. [1 ]
机构
[1] Univ Granada, Fac Sci, Dept Analyt Chem, E-18071 Granada, Spain
关键词
5-Nitroimidazoles; Dispersive liquid-liquid microextraction; Cation selective exhaustive injection; Sweeping; Micellar electrokinetic chromatography; SOLID-PHASE EXTRACTION; CAPILLARY-ELECTROPHORESIS; NITROIMIDAZOLE RESIDUES; MASS-SPECTROMETRY; DETECTION LIMITS; METABOLITES; SAMPLES; METRONIDAZOLE; SENSITIVITY; DIMETRIDAZOLE;
D O I
10.1016/j.chroma.2014.03.026
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel method consisting of cation-selective exhaustive injection and sweeping (CSEI-sweeping) as on-line preconcentration followed by a micellar electrokinetic chromatography (MEKC) separation has been developed for the determination of 5-nitroimidazoles (5-NDZ) in environmental waters. Moreover, dispersive liquid-liquid microextraction (DLLME) has been proposed for first time as sample treatment technique prior to CSEI-sweeping-MEKC. DLLME was applied to 5 mL of sample. Dibromomethane (1156 mu L) and 2-butanol (1363 mu L) were employed as extractant and dispersive solvents, respectively. Salting-out effect was achieved by the addition of 16% (w/v) NaCl to the samples. After DLLME and organic solvent evaporation, the residue was redissolved in a low conductivity solvent (5 mM phosphoric acid with 5% of methanol) and electrokinetically injected at 9.8 kV for 632 s in a bare fused-silica capillary (57.2 cm, 50 mu m I.D.). Prior to the injection, the capillary was rinsed with 50 mM phosphate buffer pH 2.5, followed by a plug of a higher conductivity buffer (100 mM phosphate pH 2.5, 50 mbar, 264 s) and a plug of water (50 mbar, 2 s). Separation was carried out applying -30 kV at 20 degrees C in 44 mM phosphate buffer pH 2.5, containing 8% tetrahydrofuran and 123 mM sodium dodecyl sulfate. Analytical signals were monitored at 276 nm. Validation was performed in river and well waters, obtaining satisfactory results in terms of linearity, precision (% RSD generally lower than 10%) and trueness (recoveries higher than 70% in almost all cases). LODs ranged from 0.61 to 2.44 ng/mL. The combination of this microextraction technique with the proposed capillary electrophoresis methodology supposes a simple, sensitive and cheap alternative for 5-NDZ analyses, in accordance with the aims of green chemistry. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:65 / 72
页数:8
相关论文
共 50 条
  • [1] Green methodology based on dispersive liquid-liquid microextraction and micellar electrokinetic chromatography for 5-nitroimidazole analysis in water samples
    Hernandez-Mesa, Maykel
    Cruces-Blanco, Carmen
    Garcia-Campana, Ana M.
    JOURNAL OF SEPARATION SCIENCE, 2013, 36 (18) : 3050 - 3058
  • [2] Dispersive liquid-liquid microextraction combined with sweeping micellar electrokinetic chromatography for the determination of some neonicotinoid insecticides in cucumber samples
    Zhang, Shuaihua
    Yang, Xiumin
    Yin, Xiaofang
    Wang, Chun
    Wang, Zhi
    FOOD CHEMISTRY, 2012, 133 (02) : 544 - 550
  • [3] Concentration and determination of cotinine in serum by cation-selective exhaustive injection and sweeping micellar electrokinetic chromatography
    Xu, Xin
    Fan, Z. Hugh
    ELECTROPHORESIS, 2012, 33 (16) : 2570 - 2576
  • [4] Cyclodextrin-assisted dispersive liquid-liquid microextraction for the preconcentration of carbamazepine and clobazam with subsequent sweeping micellar electrokinetic chromatography
    Chen, Szu-Ying
    Chen, Wen-Chi
    Chang, Sarah Y.
    JOURNAL OF SEPARATION SCIENCE, 2018, 41 (08) : 1871 - 1879
  • [5] Use of dispersive liquid-liquid microextraction for the determination of carbamates in juice samples by sweeping-micellar electrokinetic chromatography
    Moreno-Gonzalez, David
    Gamiz-Gracia, Laura
    Garcia-Campana, Ana M.
    Bosque-Sendra, Juan M.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2011, 400 (05) : 1329 - 1338
  • [6] Dispersive liquid-liquid microextraction for the determination of phenols by acetonitrile stacking coupled with sweeping-micellar electrokinetic chromatography with large-volume injection
    He, Hui
    Liu, Shuhui
    Meng, Zhaofu
    Hu, Shibing
    JOURNAL OF CHROMATOGRAPHY A, 2014, 1361 : 291 - 298
  • [7] Dispersive Liquid-Liquid Microextraction Combined With Micellar Electrokinetic Chromatography for the Determination of Some Photoinitiators in Fruit Juice
    Zhang, Yaohai
    Jiao, Bining
    FOOD ANALYTICAL METHODS, 2015, 8 (04) : 973 - 981
  • [8] Dispersive Liquid-Liquid Microextraction Combined with Micellar Electrokinetic Chromatography for the Determination of Pesticide in Apple Sample
    Wang, Yanjun
    Fan, Zhefeng
    JOURNAL OF DISPERSION SCIENCE AND TECHNOLOGY, 2014, 35 (09) : 1319 - 1324
  • [9] Determination of ephedrine and codeine in human urine by cation-selective exhaustive injection and sweeping micellar electrokinetic chromatography
    Li, Li Jun
    Li, Si Guang
    Li, Hai Yan
    Cai, Zhuo
    Cheng, Hao
    CHINESE CHEMICAL LETTERS, 2009, 20 (01) : 84 - 87
  • [10] Evaluation of the combination of micellar electrokinetic capillary chromatography with sweeping and cation selective exhaustive injection for the determination of 5-nitroimidazoles in egg samples
    Airado-Rodriguez, Diego
    Hernandez-Mesa, Maykel
    Garcia-Campana, Ana M.
    Cruces-Blanco, Carmen
    FOOD CHEMISTRY, 2016, 213 : 215 - 222