Stress analysis of cemented glenoid prostheses in Total Shoulder Arthroplasty

被引:42
作者
Gupta, S
van der Helm, FCT
van Keulen, F
机构
[1] Delft Univ Technol, Fac Design Engn & Prod, Dept Mech Engn & Marine Technol, Man Machine Syst Grp, NL-2628 CD Delft, Netherlands
[2] Deemd Univ, Bengal Engn Coll, Dept Appl Mech, Howrah 711103, W Bengal, India
[3] Delft Univ Technol, Fac Design Engn & Prod, Dept Mech Engn & Marine Technol, Struct Optimisat & Computat Mech Grp, NL-2628 CD Delft, Netherlands
关键词
shoulder; glenoid prosthesis; finite element analysis; computed tomography;
D O I
10.1016/j.jbiomech.2004.02.001
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Glenoid component loosening is the most-frequently encountered problem in the total shoulder arthroplasty. The purpose of the study was to investigate whether failure of the glenoid component is caused by stresses generated within the cement mantle, implant materials and at the various interfaces during humeral abduction, using 3-D FE analyses of implanted glenoid structures. FE models, one total polyethylene and the other, metal backed polyethylene, were developed using CT-scan data and submodelling technique, which was based on an overall solution of a natural scapula model acted upon by all the muscles, ligaments and joint reaction forces. Material interfaces were assumed to be fully bonded. Based on the FE stress analysis, the following observations were made. (1) The submodelling technique, which required a large-size submodel and the use of prescribed displacements at cut-boundaries located far away from the glenoid, was crucial for evaluations on glenoid component. (2) Total polyethylene results in lower-peak stresses (tensile: 10 MPa, Von-Mises: 8.31 MPa) in the cement as compared to a metal-backed design (tensile: 11.5 MPa, Von-Mises: 9.81 MPa). The maximum principal (tensile) stresses generated in the cement mantle for both the designs were below its failure strength, but might evoke crack initiation. (3) The cement-bone interface adjacent to the tip of the keel seemed very likely to fail for both the designs. In case of metal-backed design, this interface adjacent to the tip of the keel appears even more likely to fail. (4) High metal-cement interface stresses for a moderate load might indicate failure at higher load. (5) It appears that both the designs were vulnerable to failure in some ways or the other. A part of the subchondral bone along the longitudinal axis of the glenoid cavity should be preserved to strengthen the glenoid structure and to reduce the use of cement. (C) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1777 / 1786
页数:10
相关论文
共 40 条
  • [1] TOTAL SHOULDER ARTHROPLASTY
    BARRETT, WP
    FRANKLIN, JL
    JACKINS, SE
    WYSS, CR
    MATSEN, FA
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1987, 69A (06) : 865 - 872
  • [2] BATTE SWP, 1996, T ORTHOP RES SOC, V21, P707
  • [3] Boyd A D Jr, 1990, J Arthroplasty, V5, P329
  • [4] TOTAL SHOULDER ARTHROPLASTY WITH THE NEER PROSTHESIS
    COFIELD, RH
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1984, 66A (06) : 899 - 906
  • [5] Finite element analysis of the mechanical behavior of a scapula implanted with a glenoid prosthesis
    Couteau, B
    Mansat, P
    Estivalèzes, É
    Darmana, R
    Mansat, M
    Egan, J
    [J]. CLINICAL BIOMECHANICS, 2001, 16 (07) : 566 - 575
  • [6] DEVELOPMENT AND VALIDATION OF A 3-DIMENSIONAL FINITE-ELEMENT MODEL OF THE PELVIC BONE
    DALSTRA, M
    HUISKES, R
    VANERNING, L
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1995, 117 (03): : 272 - 278
  • [7] DALSTRA M, 1993, THESIS U NIJMEGAN NE
  • [8] COMPARISON OF THE FATIGUE CHARACTERISTICS OF CENTRIFUGED AND UNCENTRIFUGED SIMPLEX P-BONE CEMENT
    DAVIES, JP
    BURKE, DW
    OCONNOR, DO
    HARRIS, WH
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 1987, 5 (03) : 366 - 371
  • [9] FRICH LH, 1994, THESIS U AARHUS AARH
  • [10] Friedman R J, 1992, J Shoulder Elbow Surg, V1, P261, DOI 10.1016/S1058-2746(09)80068-2