Disposable real-time microPCR device: lab-on-a-chip at a low cost

被引:99
作者
Neuzil, Pavel [1 ]
Pipper, Juergen [1 ]
Hsieh, Tseng Ming [1 ]
机构
[1] Inst Bioengn & Nanotechnol, Singapore 138669, Singapore
关键词
D O I
10.1039/b605957k
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
We have designed, fabricated and tested a real-time micro polymerase chain reaction (microPCR) system. It consists of a microscope glass cover slip placed on top of a micromachined silicon chip integrated with a heater and a temperature sensor. A single mu L of a sample containing DNA was placed on the glass and encapsulated with mineral oil to prevent the evaporation of water, thus forming a virtual reaction chamber (VRC). The PCR chip required half a second to heat up from 72 to 94 degrees C and two seconds to cool from 94 to 55 degrees C, corresponding to a cooling rate of -20 K s(-1). The real-time PCR yield was determined by a fluorescence method. The melting curve analysis method as well as capillary electrophoresis was performed to determine the purity of the PCR product. As the glass slip is disposable, cross-contamination from sample to sample is eliminated. The total cost of running the PCR is given by the value of the cover slip and its treatment.
引用
收藏
页码:292 / 298
页数:7
相关论文
共 31 条
[1]   Real-time PCR detection of Listeria monocytogenes using an integrated microfluidics platform [J].
Cady, NC ;
Stelick, S ;
Kunnavakkam, MV ;
Batt, CA .
SENSORS AND ACTUATORS B-CHEMICAL, 2005, 107 (01) :332-341
[2]   Analysis of ligase chain reaction products amplified in a silicon-glass chip using capillary electrophoresis [J].
Cheng, J ;
Shoffner, MA ;
Mitchelson, KR ;
Kricka, LJ ;
Wilding, P .
JOURNAL OF CHROMATOGRAPHY A, 1996, 732 (01) :151-158
[3]   Light emitting diode-based detectors absorbance, fluorescence and spectroelectrochemical measurements in a planar flow-through cell [J].
Dasgupta, PK ;
Eom, IY ;
Morris, KJ ;
Li, JZ .
ANALYTICA CHIMICA ACTA, 2003, 500 (1-2) :337-364
[4]  
FINDLAY JB, 1993, CLIN CHEM, V39, P1927
[5]   THEORY OF DNA MELTING CURVES [J].
FIXMAN, M ;
FREIRE, JJ .
BIOPOLYMERS, 1977, 16 (12) :2693-2704
[6]   A BioMEMS review: MEMS technology for physiologically integrated devices [J].
Grayson, ACR ;
Shawgo, RS ;
Johnson, AM ;
Flynn, NT ;
Li, YW ;
Cima, MJ ;
Langer, R .
PROCEEDINGS OF THE IEEE, 2004, 92 (01) :6-21
[7]   Planar chip device for PCR and hybridization with surface acoustic wave pump [J].
Guttenberg, Z ;
Müller, H ;
Habermüller, H ;
Geisbauer, A ;
Pipper, J ;
Felbel, J ;
Kielpinski, M ;
Scriba, J ;
Wixforth, A .
LAB ON A CHIP, 2005, 5 (03) :308-317
[8]   Rapid PCR in a continuous flow device [J].
Hashimoto, M ;
Chen, PC ;
Mitchell, MW ;
Nikitopoulos, DE ;
Soper, SA ;
Murphy, MC .
LAB ON A CHIP, 2004, 4 (06) :638-645
[9]   Noncontact infrared-mediated thermocycling for effective polymerase chain reaction amplification of DNA in nanoliter volumes [J].
Hühmer, AFR ;
Landers, JP .
ANALYTICAL CHEMISTRY, 2000, 72 (21) :5507-5512
[10]   Chemical amplification: Continuous-flow PCR on a chip [J].
Kopp, MU ;
de Mello, AJ ;
Manz, A .
SCIENCE, 1998, 280 (5366) :1046-1048