T-Configuration Horizontal Low-Profile Visualized Intraluminal Support (LVIS Jr) Device-Assisted Coiling for Treatment of Basilar Tip Aneurysms: A Technical Note

被引:2
作者
Dornbos, David, III [1 ]
Khandpur, Umang [1 ]
Youssef, Patrick P. [1 ]
机构
[1] Ohio State Univ, Wexner Med Ctr, Dept Neurol Surg, Columbus, OH 43210 USA
关键词
Coil embolization; Endovascular; Intracranial aneurysm; Intracranial stent; MULTICENTER;
D O I
10.1016/j.wneu.2019.05.220
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND: Provided certain anatomic considerations, stent-assisted coiling of basilar apex aneurysms can be performed using a single stent placed in a horizontal T-configuration across the aneurysm neck prior to coil deployment, although this has predominantly been described using older-generation stents. The development of the Low-profile Visualized Intraluminal Support (LVIS Jr) device has provided greater versatility than previous stents, including use in smaller vessels, the ability to be resheathed, and improved flow diversion properties. METHODS: To our knowledge, we report the first use of the LVIS Jr device in a horizontal T-configuration across the aneurysm neck prior to coil deployment in 2 patients for the treatment of basilar apex aneurysms. This technique requires robust posterior communicating artery aneurysms and a second site of vascular access. RESULTS: Both patients demonstrated good outcomes following the procedure with no adverse sequelae. One patient required retreatment for recurrence after 1 year and was able to undergo further coiling without difficulty. CONCLUSIONS: Use of the LVIS Jr device allows greater versatility in horizontal T-configuration stent-assisted coil embolization, when compared with other available intracranial stents. This provides another tool to treat basilar apex aneurysms with improved coil occlusion and a theoretically decreased risk of thromboembolic events.
引用
收藏
页码:428 / 431
页数:4
相关论文
共 7 条
[1]   Treatment of basilar tip aneurysms with horizontal PCA to PCA stent-assisted coiling: case series [J].
Blackburn, Spiros L. ;
Kadkhodayan, Yasha ;
Shekhtman, Eugenia ;
Derdeyn, Colin P. ;
Cross, DeWitte T., III ;
Moran, Christopher J. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2013, 5 (03) :212-216
[2]  
Chow MM, 2004, AM J NEURORADIOL, V25, P509
[3]   Y stenting assisted coiling using a new low profile visible intraluminal support device for wide necked basilar tip aneurysms: a technical report [J].
Conrad, Marcelo D. ;
Brasiliense, Leonardo B. C. ;
Richie, Alexa N. ;
Hanel, Ricardo A. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2014, 6 (04) :296-300
[4]  
Cross DT, 2005, AM J NEURORADIOL, V26, P2578
[5]   A Multicenter Study of Stent-Assisted Coiling of Cerebral Aneurysms With a Y Configuration [J].
Fargen, Kyle M. ;
Mocco, J. ;
Neal, Dan ;
Dewan, Michael C. ;
Reavey-Cantwell, John ;
Woo, Henry H. ;
Fiorella, David J. ;
Mokin, Maxim ;
Siddiqui, Adnan H. ;
Turk, Aquilla S. ;
Turner, Raymond D. ;
Chaudry, Imran ;
Kalani, M. Yashar S. ;
Albuquerque, Felipe ;
Hoh, Brian L. .
NEUROSURGERY, 2013, 73 (03) :466-472
[6]   Treatment of wide-necked aneurysms with the Low-profile Visualized Intraluminal Support (LVIS Jr) device: a multicenter experience [J].
Grossberg, J. A. ;
Hanel, R. A. ;
Dabus, G. ;
Keigher, K. ;
Haussen, D. C. ;
Sauvageau, E. ;
Linfante, I. ;
Gonsales, D. ;
Salinas, P. Aguilar ;
Bouslama, M. ;
Mayich, M. ;
Nogueira, R. G. ;
Lopes, D. K. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2017, 9 (11) :1098-1102
[7]   Horizontal stent placement plus coiling in a broad-based basilar-tip aneurysm: an alternative to the Y-stent technique [J].
Wanke, Isabel ;
Gizewski, Elke ;
Forsting, Michael .
NEURORADIOLOGY, 2006, 48 (11) :817-820