Development of Implant Stability Quotient values of implants placed with simultaneous sinus floor elevation - results of a prospective study with 109 implants

被引:33
作者
Kuchler, Ulrike [1 ,2 ]
Chappuis, Vivianne [1 ]
Bornstein, Michael M. [1 ]
Siewczyk, Marta [1 ,3 ]
Gruber, Reinhard [4 ,5 ]
Maestre, Laura [1 ]
Buser, Daniel [1 ]
机构
[1] Univ Bern, Dept Oral Surg & Stomatol, Bern, Switzerland
[2] Med Univ Vienna, Dept Oral Surg, Vienna, Austria
[3] Univ Bern, Sch Dent Med, Dept Reconstruct Dent Med & Gerodontol, Bern, Switzerland
[4] Univ Bern, Sch Dent Med, Lab Oral Cell Biol, Bern, Switzerland
[5] Med Univ Vienna, Dept Oral Biol, Vienna, Austria
关键词
autogenous bone grafts; Implant Stability Quotient; dental implant; deproteinized bovine bone mineral; early implant loading; resonance frequency analysis; SLActive; sinus floor elevation; ACID-ETCHED SURFACE; RESONANCE FREQUENCY-ANALYSIS; CHEMICALLY-MODIFIED SLACTIVE((R)); NONSUBMERGED TITANIUM IMPLANTS; DEHISCENCE-TYPE DEFECTS; LONG-TERM STABILITY; AUTOGENOUS BONE; CONSENSUS STATEMENTS; EDENTULOUS MAXILLA; DIFFERENT RATIOS;
D O I
10.1111/clr.12768
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
ObjectivesIn patients with implant placement and simultaneous sinus floor elevation (SFE), healing periods of 6months have been the standard of care for more than 25years. The primary objective of this prospective case series study was to determine what percentage of implants placed with SFE reach a threshold Implant Stability Quotient (ISQ) of 70 after 8weeks of healing using Resonance Frequency Analysis (RFA). Material and MethodsA total of 109 dental implants were placed in 97 patients. SFE was carried out with a lateral window approach and a mixture of autogenous bone chips and deproteinized bovine bone mineral (DBBM). Titanium screw-type, tissue-level implants with a chemically modified SLA surface were used. ISQ values were measured after implant insertion (ISQ(BL)) and after 8weeks of healing (ISQ(8wk)). Patients showing ISQ(8wk)70 subsequently underwent restoration. Implants with an ISQ value<70 were recalled at 2-week intervals. ResultsThe ISQ at baseline had a mean value of 68.3 (SD9.8). At 8weeks, the mean ISQ value was 73.6 (SD +/- 6.4). This increase was statistically significant (P<0.001). An ISQ(8wk) value 70 was observed for 91 implants (83%). One implant (0.9%) with a peri-implant infection and severe bone loss at 8weeks was considered an early failure. ConclusionsThis study showed that 83% of implants reached the threshold level of ISQ70 after 8weeks, allowing an early loading protocol. The early failure rate was considered low with 0.9%. The RFA technology is a suitable method to objectively monitor implant stability longitudinally.
引用
收藏
页码:109 / 115
页数:7
相关论文
共 54 条
  • [1] A 15-YEAR STUDY OF OSSEOINTEGRATED IMPLANTS IN THE TREATMENT OF THE EDENTULOUS JAW
    ADELL, R
    LEKHOLM, U
    ROCKLER, B
    BRANEMARK, PI
    [J]. INTERNATIONAL JOURNAL OF ORAL SURGERY, 1981, 10 (06): : 387 - 416
  • [2] [Anonymous], 2019, R PACKAGE VERSION 31
  • [3] Blomqvist J E, 1996, Int J Oral Maxillofac Implants, V11, P512
  • [4] Early Loading at 21 Days of Non-Submerged Titanium Implants With a Chemically Modified Sandblasted and Acid-Etched Surface: 3-Year Results of a Prospective Study in the Posterior Mandible
    Bornstein, Michael M.
    Wittneben, Julia-Gabriela
    Bragger, Urs
    Buser, Daniel
    [J]. JOURNAL OF PERIODONTOLOGY, 2010, 81 (06) : 809 - 818
  • [5] Early Loading of Nonsubmerged Titanium Implants with a Chemically Modified Sand-Blasted and Acid-Etched Surface: 6-Month Results of a Prospective Case Series Study in the Posterior Mandible Focusing on Peri-Implant Crestal Bone Changes and Implant Stability Quotient (ISQ) Values
    Bornstein, Michael M.
    Hart, Christopher N.
    Halbritter, Sandro A.
    Morton, Dean
    Buser, Daniel
    [J]. CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2009, 11 (04) : 338 - 347
  • [6] Brügger OE, 2015, INT J ORAL MAX IMPL, V30, P151, DOI 10.11607/jomi.3769
  • [7] Evaluation of filling materials in membrane-protected bone defects - A comparative histomorphometric study in the mandible of miniature pigs
    Buser, D
    Hoffmann, B
    Bernard, JP
    Lussi, A
    Mettler, D
    Schenk, RK
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 1998, 9 (03) : 137 - 150
  • [8] Long-term Stability of Early Implant Placement with Contour Augmentation
    Buser, D.
    Chappuis, V.
    Kuchler, U.
    Bornstein, M. M.
    Wittneben, J. G.
    Buser, R.
    Cavusoglu, Y.
    Belser, U. C.
    [J]. JOURNAL OF DENTAL RESEARCH, 2013, 92 (S12) : 176S - 182S
  • [9] Enhanced bone apposition to a chemically modified SLA titanium surface
    Buser, D
    Broggini, N
    Wieland, M
    Schenk, RK
    Denzer, AJ
    Cochran, DL
    Hoffmann, B
    Lussi, A
    Steinemann, SG
    [J]. JOURNAL OF DENTAL RESEARCH, 2004, 83 (07) : 529 - 533
  • [10] Buser D, 1999, PROCEEDINGS OF THE 3RD EUROPEAN WORKSHOP ON PERIODONTOLOGY: IMPLANT DENTISTRY, P88