Capillary gas chromatography coupled with negative ionization microplasma mass spectrometry for halogen-selective detection

被引:18
|
作者
Brede, C
Pedersen-Bjergaard, S
Lundanes, E
Greibrokk, T
机构
[1] Univ Oslo, Dept Chem, N-0315 Oslo, Norway
[2] Univ Oslo, Sch Pharm, N-0316 Oslo, Norway
关键词
D O I
10.1039/a904211c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Gas chromatography was coupled with plasma mass spectrometry with a microplasma ion source for negative ion detection. The ion source, which was kept inside the high vacuum chamber of the mass spectrometer, was a rigid fused silica capillary tube containing a capacitively coupled radiofrequency helium plasma. This made the setup quite simple: eliminating the sampler-skimmer pressure-reducing interface traditionally used in plasma mass spectrometry. The present study describes the utilization of halogen-selective negative ion detection. A high selectivity of fluorine to hydrocarbon compounds (3 x 10(3)), and a highly sensitive detection for F, Cl, Br, and I (0.13-12 pg s(-1)), were obtained. The mechanisms of negative ion formation and breakdown were discussed in conjunction with the results that were achieved.