Biomechanical comparison of the femoral neck system versus InterTan nail and three cannulated screws for unstable Pauwels type III femoral neck fracture

被引:25
|
作者
Wang, Zheng [1 ]
Yang, Yong [1 ]
Feng, Gangning [1 ]
Guo, Haohui [1 ]
Chen, Zhirong [1 ]
Chen, Yaogeng [2 ]
Jin, Qunhua [1 ]
机构
[1] Ningxia Med Univ, Dept Orthopaed, Gen Hosp, Yinchuan 750004, Ningxia, Peoples R China
[2] Ningxia Med Univ, Yinchuan 750004, Ningxia, Peoples R China
基金
中国国家自然科学基金;
关键词
Pauwels type III femoral neck fracture; Biomechanical properties; Internal fixation; Femoral neck system; Artificial composite femur bone; INTRACAPSULAR HIP FRACTURE; INTERNAL-FIXATION; LOCKING; FORCES; REDUCTION; STABILITY; FAILURE; DEVICES; FEMURS;
D O I
10.1186/s12938-022-01006-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background There are a variety of internal fixation methods for unstable femoral neck fractures (FNFs), but the best method is still unclear. Femoral neck system (FNS) is a dynamic angular stabilization system with cross screws, and is a new internal fixation implant designed for minimally invasive fixation of FNFs. In this study, we conducted a biomechanical comparison of FNS, InterTan nail and three cannulated screws for the treatment of Pauwels III FNFs and investigate the biomechanical properties of FNS. Methods A total of 18 left artificial femurs were selected and randomly divide into Group A (fixation with FNS), Group B (fixation with InterTan nail) and Group C (fixation with three cannulated screws), with 6 specimens in each group. After creating Pauwels type III FNF models, the specimens in each were tested with non-destructive quasi-static tests, including torsion, A-P bending and axial compression tests. The average slope of the linear load-deformation curve obtained from quasi-static tests defines the initial torsional stiffness, A-P bending stiffness, and axial compression stiffness. After cyclic loading test was applied, the overall deformation of models and local deformation of implant holes in each group were assessed. The overall deformation was estimated as the displacement recorded by the software of the mechanical testing apparatus. Local deformation was defined as interfragmental displacement. Data were analyzed by one-way analysis of variance (ANOVA) followed by Bonferroni post hoc test using the SPSS software (version 24.0, IBM, New York, NY, USA). Correlation analysis was performed using Pearson's correlation analysis. Results Group B exhibited significantly higher axial stiffness and A-P bending stiffness than the other two groups (P < 0.01), while Group A had significantly higher axial stiffness and A-P bending stiffness than Group C (P < 0.01). Groups A and B exhibited significantly higher torsional stiffness than Group C (P < 0.01), no statistical significance was observed between Groups A and B (P > 0.05). Group B exhibited significantly lower overall and local deformations than the other two groups (P < 0.01), while Group A had significantly lower overall and local deformations than Group C (P < 0.01). Correlation analysis revealed positive correlation between axial stiffness and A-P bending stiffness (r = 0.925, P < 0.01), torsional stiffness (r = 0.727, P < 0.01), between torsional stiffness and A-P bending stiffness; negative correlation between overall, local deformations and axial stiffness (r = - 0.889, - 0.901, respectively, both P < 0.01), and positive correlation between the two deformations (r = - 0.978, P < 0.01). Conclusion For fixation of unstable FNFs, InterTan nail showed the highest axial stiffness and A-P bending stiffness, followed by FNS, and then three cannulated screws. Torsional stiffness of FNS was comparable to that of the InterTan nail. FNS, as a novel minimally invasive implant, can create good mechanical environment for the healing of unstable FNFs. Clinical studies are needed to confirm the potential advantages of FNS observed in this biomechanical study.
引用
收藏
页数:14
相关论文
共 50 条
  • [41] Mechanical effects of surgical variations in the femoral neck system on Pauwels type III femoral neck fracture A FINITE ELEMENT ANALYSIS
    Jung, C-H
    Cha, Y.
    Yoon, H. S.
    Park, C. H.
    Yoo, J-, I
    Kim, J-T
    Jeon, Y.
    BONE & JOINT RESEARCH, 2022, 11 (02): : 102 - 111
  • [42] Comparison of femoral neck system versus cannulated compression screws in treating femoral neck fractures: A systematic review and meta-analysis
    Zhang, Jian
    Chang, Xiaohu
    Sun, Zhonghui
    Tang, Xin
    ASIAN JOURNAL OF SURGERY, 2023, 46 (08) : 3259 - 3260
  • [43] Femoral neck system vs. cannulated screws on treating femoral neck fracture: a meta-analysis and system review
    Zhou, Yimin
    Li, Zongyang
    Lao, Kecheng
    Wang, Zixiu
    Zhang, Li
    Dai, Shiyou
    Fan, Xiao
    FRONTIERS IN SURGERY, 2023, 10
  • [44] Biomechanical Research of Three Parallel Cannulated Compression Screws in Oblique Triangle Configuration for Fixation of Femoral Neck Unstable Fractures
    Zhang, Ru-Yi
    Zhang, Wu-Peng
    Yang, Guang-Min
    Wang, Dao-Feng
    Su, Peng
    Zhang, Yi
    Nie, Shao-Bo
    Li, Jia
    Zhao, Zhe
    Li, Jian-Tao
    Zhang, Li-Cheng
    Tang, Pei-Fu
    ORTHOPAEDIC SURGERY, 2024, 16 (04) : 953 - 964
  • [45] Letter to the editor regarding "Comparison of femoral neck system versus cannulated cancellous screws for the fixation of femoral neck fracture: a single-center retrospective cohort study"
    Pan, Zhicheng
    Liu, Wangxin
    Zhang, Qi
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2024, 34 (06): : 3215 - 3216
  • [46] Fracture morphology and biomechanical characteristics of Pauwels III femoral neck fractures in young adults
    Wang, Kaiyang
    Ni, Ming
    Liao, Peng
    Dou, Bang
    Yan, Xu
    Lv, Lin
    Zhang, Fangfang
    Mei, Jiong
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2021, 52 (11): : 3227 - 3238
  • [47] Comparison of femoral neck system and three cannulated cancellous screws in the treatment of vertical femoral neck fractures: clinical observation and finite element analysis
    Huang, Shaolong
    Zhang, Yazhong
    Zhang, Xu
    Zhou, Chengqiang
    Li, Wenbo
    Wang, Yunqing
    Wang, Bin
    Zhu, Ziqiang
    BIOMEDICAL ENGINEERING ONLINE, 2023, 22 (01)
  • [48] Comparison of femoral neck system and three cannulated cancellous screws in the treatment of vertical femoral neck fractures: clinical observation and finite element analysis
    Shaolong Huang
    Yazhong Zhang
    Xu Zhang
    Chengqiang Zhou
    Wenbo Li
    Yunqing Wang
    Bin Wang
    Ziqiang Zhu
    BioMedical Engineering OnLine, 22
  • [49] Is there an increased risk for subtrochanteric stress fracture with the Femoral Neck System versus multiple cannulated screws fixation?
    Hsu, Megan R.
    Shu, Henry T.
    Luksameearunothai, Kitchai
    Margalit, Adam
    Yu, Andrew T.
    Hasenboehler, Erik A.
    Shafiq, Babar
    JOURNAL OF ORTHOPAEDICS, 2022, 30 : 127 - 133
  • [50] Internal fixation of femoral neck fractures with posterior comminutionA biomechanical comparison of DHS® and Intertan nail®
    Martin Rupprecht
    Lars Grossterlinden
    Kai Sellenschloh
    Michael Hoffmann
    Klaus Püschel
    Michael Morlock
    Johannes M. Rueger
    Wolfgang Lehmann
    International Orthopaedics, 2011, 35 : 1695 - 1701