Patient Size-Specific Analysis of Dose Indexes From CT Lung Cancer Screening

被引:17
作者
Fujii, Keisuke [1 ]
McMillan, Kyle [2 ]
Bostani, Maryam [2 ,3 ]
Cagnon, Christopher [2 ,3 ]
McNitt-Gray, Michael [2 ,3 ]
机构
[1] Nagoya Univ, Grad Sch Med, Dept Radiol Sci, Higashi Ku, 1-1-20 Daikominami, Nagoya, Aichi 4618673, Japan
[2] Univ Calif Los Angeles, David Geffen Sch Med, Phys & Biol Med Grad Program, Los Angeles, CA USA
[3] Univ Calif Los Angeles, David Geffen Sch Med, Dept Radiol Sci, Los Angeles, CA USA
关键词
low-dose CT; lung cancer screening; volumetric CT dose index; TUBE CURRENT MODULATION; CHEST CT; TRIAL; PROGRAM;
D O I
10.2214/AJR.16.16082
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The U.S. Centers for Medicare & Medicaid Services (CMS) recently approved the use of low-dose CT for lung cancer screening and described volumetric CT dose index (CTDIvol) requirements. These were based on the National Lung Screening Trial, which used only fixed-tube-current techniques. The aim of this study was to evaluate dose index data from a lung cancer screening program using automatic exposure control (AEC) techniques to ensure compliance with requirements and to correlate dose index values with patient size. MATERIALS AND METHODS. CTDIvol, dose-length product (DLP), and body mass index (BMI) data were collected for 563 lung cancer screening examinations performed with AEC between January 1, 2014, through August 31, 2015. CTDIvol and DLP were analyzed according to the patient's BMI classification. Results were compared with the CMS requirement that the CTDIvol for a standard-sized patient (height, 170 cm; weight, 70 kg) be 3.0 mGy or less, with adjustments for patients of different sizes. For a subset of patients, the average water-equivalent diameter and size-specific dose estimate were estimated. RESULTS. The average CTDIvol for a standard-sized patient was 1.8 mGy, which meets CMS requirements. CTDIvol values were lower for smaller patients and higher for larger patients. Overall, the mean CTDIvol and DLP were 2.1 mGy and 74 mGy.cm, respectively. The size-specific dose estimate for the average water-equivalent diameter (27.5 cm) of the patient subset was 2.6 mGy. CONCLUSION. The screening protocols using AEC resulted in CTDIvol values that were compliant with CMS requirements. CTDIvol values greater than 3.0 mGy were only observed for overweight or obese patients.
引用
收藏
页码:144 / 149
页数:6
相关论文
共 19 条
[1]  
AAPM, 2011, SIZ SPEC DOS EST SSD
[2]  
aapm, AAPM lung cancer screening CT protocols version 5.1-13 th September 2019
[3]  
American College of Radiology (ACR), DES LUNG CANC SCREEN
[4]  
American College of Radiology (ACR), AM COLL RAD NEMA XR
[5]  
American College of Radiology (ACR), NAT RAD DAT REG
[6]  
American College of Radiology (ACR), LUNG CANC SCREEN RES
[7]  
American College of Radiology (ACR), REQ FIELDS LCSR
[8]   Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation [J].
Angel, Erin ;
Yaghmai, Nazanin ;
Jude, Cecilia Matilda ;
DeMarco, John J. ;
Cagnon, Christopher H. ;
Goldin, Jonathan G. ;
McCollough, Cynthia H. ;
Primak, Andrew N. ;
Cody, Dianna D. ;
Stevens, Donna M. ;
McNitt-Gray, Michael F. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 193 (05) :1340-1345
[9]   Monte Carlo simulations to assess the effects of tube current modulation on breast dose for multidetector CT [J].
Angel, Erin ;
Yaghmai, Nazanin ;
Jude, Cecilia Matilda ;
DeMarco, John J. ;
Cagnon, Christopher H. ;
Goldin, Jonathan G. ;
Primak, Andrew N. ;
Stevens, Donna M. ;
Cody, Dianna D. ;
McCollough, Cynthia H. ;
McNitt-Gray, Michael F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (03) :497-511
[10]  
[Anonymous], 2014, USE WATER EQUIVALENT, DOI [10.37206/146, DOI 10.37206/146]