An electrochemical immunosensor for digoxin using core-shell gold coated magnetic nanoparticles as labels

被引:71
作者
Ahmadi, Anita [1 ,2 ]
Shirazi, Hanieh [1 ]
Pourbagher, Narges [1 ]
Akbarzadeh, Abolfazl [3 ]
Omidfar, Kobra [1 ,2 ]
机构
[1] Univ Tehran Med Sci, Biosensor Res Ctr, Endocrinol & Metab Mol Cellular Sci Inst, Tehran, Iran
[2] Univ Tehran Med Sci, Endocrinol & Metab Res Inst, Endocrinol & Metab Res Ctr, Tehran, Iran
[3] Tabriz Univ Med Sci, Dept Med Chem, Tabriz, Iran
关键词
Core-shell gold coated magnetic nanoparticles; Immunosensor; Digoxin; COLLOIDAL GOLD; SIGNAL AMPLIFICATION; HEART-FAILURE; SERUM DIGOXIN; ANTIBODY; RADIOIMMUNOASSAY; ASSAY;
D O I
10.1007/s11033-013-3014-4
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
A simple, sensitive, and low-cost immunosensor was designed for the detection of digoxin through core-shell gold coated magnetic nanoparticles (Fe3O4-Au-NPs) as an electrochemical label. Having had such a large potential for a variety of applications, Fe3O4-Au-NPs have attracted a considerable attention and are actively investigated recently. Digoxin is a cardiac glycoside which, at high level, can indicate an increased risk of toxicity. This new competitive electrochemical immunosensor was developed based on antigen-antibody reaction employing antigen (Ag) labeled Fe3O4-Au-NPs and PVA modified screen-printed carbon electrode surface in order to detect the serum digoxin. The structures of Fe3O4-Au-NPs were studied by transmission electron microscopy, X-ray diffraction and Fourier transformed infrared spectroscopy. Cyclic voltammetry and differential pulse voltammetry (DPV) were employed to determine the physicochemical and electrochemical properties of immunosensor. DPV was employed for quantitative detection of digoxin in biological samples. The developed immunosensor was capable to detect digoxin in the range from 0.5 to 5 ng mL(-1), with a detection limit as low as 0.05 ng mL(-1). The proposed method represented acceptable reproducibility, stability, and reliability for the rapid detection of digoxin in serum samples.
引用
收藏
页码:1659 / 1668
页数:10
相关论文
共 42 条
  • [1] The underuse of digoxin in heart failure, and approaches to appropriate use
    Ahmed, Ali
    Young, James B.
    Gheorghiade, Mihai
    [J]. CANADIAN MEDICAL ASSOCIATION JOURNAL, 2007, 176 (05) : 641 - 643
  • [2] Magnetic nanoparticles for drug delivery
    Arruebo, Manuel
    Fernandez-Pacheco, Rodrigo
    Ibarra, M. Ricardo
    Santamaria, Jesus
    [J]. NANO TODAY, 2007, 2 (03) : 22 - 32
  • [3] Gold nanoparticle-based signal amplification for biosensing
    Cao, Xiaodong
    Ye, Yongkang
    Liu, Songqin
    [J]. ANALYTICAL BIOCHEMISTRY, 2011, 417 (01) : 1 - 16
  • [4] Gold nanoparticles-coated magnetic microspheres as affinity matrix for detection of hemoglobin Al c in blood by microfluidic immunoassay
    Chen, Shao-Peng
    Yu, Xiao-Dong
    Xu, Jing-Juan
    Chen, Hong-Yuan
    [J]. BIOSENSORS & BIOELECTRONICS, 2011, 26 (12) : 4779 - 4784
  • [5] A sensitive immunosensor using colloidal gold as electrochemical label
    Chen, Zhao-Peng
    Peng, Zhao-Feng
    Zhang, Peng
    Jin, Xue-Fang
    Jiang, Jian-Hui
    Zhang, Xiao-Bing
    Shen, Guo-Li
    Yu, Ru-Qin
    [J]. TALANTA, 2007, 72 (05) : 1800 - 1804
  • [6] Labeling digoxin antibody with colloidal gold and ferrocene for its use in a membrane immunostrip and immunosensor
    Choi, MJ
    Kim, SY
    Choi, J
    Paeng, IR
    [J]. MICROCHEMICAL JOURNAL, 1999, 63 (01) : 92 - 99
  • [7] Synthesis of GoldMag particles with assembled structure and their applications in immunoassay
    Cui Yali
    Zhang Lianying
    Su Jing
    Zhang Caifeng
    Li Qi
    Cui Ting
    Jin Boquan
    Chen Chao
    [J]. SCIENCE IN CHINA SERIES B-CHEMISTRY, 2006, 49 (06): : 534 - 540
  • [8] The synthesis of GoldMag nano-particles and their application for antibody immobilization
    Cui, YL
    Wang, YN
    Hui, WL
    Zhang, ZF
    Xin, XF
    Chen, C
    [J]. BIOMEDICAL MICRODEVICES, 2005, 7 (02) : 153 - 156
  • [9] A nanocatalyst-based assay for proteins:: DNA-free ultrasensitive electrochemical detection using catalytic reduction of p-nitrophenol by gold-nanoparticle labels
    Das, Jagotamoy
    Aziz, Md. Abdul
    Yang, Haesik
    [J]. JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2006, 128 (50) : 16022 - 16023
  • [10] A new enzyme-linked chemiluminescent immunosorbent digoxin assay is virtually free from interference of spironolactone, potassium canrenoate, and their common metabolite canrenone
    Dasgupta, Arnitava
    Kang, Edward
    Datta, Pradip
    [J]. JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2006, 20 (05) : 204 - 208