Functional properties and performance of new and reprocessed coronary angioplasty balloon catheters

被引:14
作者
Fedel, Mariangela [1 ]
Tessarolo, Francesco
Ferrari, Paolo
Loesche, Christian
Ghassemieh, Nikou
Guarrera, Giovanni Maria
Nollo, Giandomenico
机构
[1] Univ Trent, Dept Mat Engn & Ind Technol, Trento, Italy
[2] Univ Trent, Dept Phys, Trento, Italy
[3] Vanguard AG, Dept Res & Dev, Berlin, Germany
[4] Azienda Prov Servzi Sanit, Dept Care & Rehabil, Trento, Italy
[5] ITC irst, Trento, Italy
关键词
single use medical devices; PTCA catheters; reuse; reprocessing; functionality tests;
D O I
10.1002/jbm.b.30496
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The need of health costs control has prompted the reuse of devices originally manufactured for single use only. To assess reprocessing feasibility of disposable medical devices, hygienic, technical, and functional aspects must be taken into account in addition to economical, ethical, and legal implications. This study aims to characterize coronary angioplasty catheters and to evaluate performance changes induced by reprocessing. Multiple analysis including crossing profile, slipperiness, compliance, mechanical, and burst pressure tests were performed at different steps of the protocol on 25 catheters reprocessed up to two times. The results highlighted that both use and reprocessing can affect the features of angioplasty catheters. Mechanical stress caused by clinical inflation and thermal- chemical stress undergone by polymers during cleaning and sterilization procedures caused partial modifications of material properties, inducing an overall shrinking effect on balloons. Compliance tests reported a maximum variation of 6.2% from nominal values, showing the conformity of reprocessed devices with manufacturers' original specifications (+/- 10%). The burst pressure of reprocessed devices was 80% higher than the rated burst pressure certified by manufacturers, thus reducing concerns of breakage during reuse. A strict dependence on device model in the behavior of catheters was found, especially for balloons crossing profile and slipperiness. Main changes occurred after the first reprocessing cycle, while a second cleaning and sterilization did not introduce further significant alterations. On the whole, the magnitude of modifications introduced up to two reprocessing cycles did not compromise catheters performance. (c) 2006 Wiley Periodicals. Inc.
引用
收藏
页码:364 / 372
页数:9
相关论文
共 32 条
[1]   EFFECTS OF REUSE ON THE PHYSICAL CHARACTERISTICS OF ANGIOGRAPHIC CATHETERS [J].
BENTOLILA, P ;
JACOB, R ;
ROBERGE, F .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1990, 14 (06) :254-259
[2]   The reuse of single-use balloon angioplasty catheters: Is it now legitimate? [J].
Bourassa, MG .
AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 (06) :673-674
[3]  
Brown Stanley A., 2001, Biomedical Instrumentation and Technology, V35, P312
[4]   Initial experience with reuse of coronary angioplasty catheters in the United States [J].
Browne, KF ;
Maldonado, R ;
Telatnik, M ;
Vlietstra, RE ;
Brenner, AS .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1997, 30 (07) :1735-1740
[5]  
Burtt BL, 1995, Gesneriana, V1, P1
[6]  
FECHT G, 1998, ZENTR STERIL, V6, P12
[7]  
*FOOD DRUG ADM CTR, 1999, FDA PROP STRAT REUS
[8]   Assessment of parameters associated to the risk of PVC catheter reuse [J].
Granados, DL ;
Jiménez, A ;
Cuadrado, TR .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2001, 58 (05) :505-510
[9]  
Grimandi G, 1996, CATHETER CARDIO DIAG, V38, P123, DOI 10.1002/(SICI)1097-0304(199606)38:2<123::AID-CCD1>3.0.CO
[10]  
2-B