Objective: Assess endocochlear trauma by adjusting: 1) location of cochleostomy or round window insertion, and 2) size of precontoured electrode array. Study Design: Cadaveric temporal bone study. Methods: Locations of electrode placement into the cochlea were as follows: 1) round window, 2) anterior inferior to the round window, 3) anterior inferior to the round window niche, 4) superior to the round window niche. Two types of electrode arrays were used: a larger precontoured electrode and thinner pre-curved research electrode. Histologic sections were made by a blinded third party. Results: Fourteen bones were included in the study. Six (42.8%) of the bones were right ears. Seven bones had no endocochlear trauma. Seven bones have intracochlear trauma. Round window insertions had a high incidence of intracochlear trauma with precontoured electrodes (3/4 bones). Superior cochleostomies with electrode placement had significant intracochlear trauma (2/2 bones). Insertions made anterior inferior to the round window annulus had a 50% incidence of intracochlear trauma (2/4 bones). No endocochlear trauma observed for insertions through traditional cochleostomies (4/4 bones). Less endocochlear trauma was observed with the thinner electrode: 57% versus 42% in the larger electrode array. A higher incidence of tip fold-over was observed with the thinner electrode array (2/7 electrodes). No tip fold-over was noted in the larger electrode. Conclusion: The traditional cochleostomy had the least incidence of endocochlear trauma. The smaller electrode array did not significantly affect the incidence of endocochlear trauma, but the thinner array had a higher incidence of tip fold-over, which caused trauma distally.
引用
收藏
页码:52 / 57
页数:6
相关论文
共 21 条
[1]   Preservation of basal inner ear structures in cochlear implantation [J].
Adunka, O ;
Gstoettner, W ;
Hambek, M ;
Unkelbach, MH ;
Radeloff, A ;
Kiefer, J .
ORL-JOURNAL FOR OTO-RHINO-LARYNGOLOGY AND ITS RELATED SPECIALTIES, 2004, 66 (06) :306-312
[2]   Combining perimodiolar electrode placement and atraumatic insertion properties in cochlear implantation - fact or fantasy? [J].
Adunka, OF ;
Pillsbury, HC ;
Kiefer, J .
ACTA OTO-LARYNGOLOGICA, 2006, 126 (05) :475-482
[3]   Impact of electrode insertion depth on intracochlear trauma [J].
Adunka, Oliver ;
Kiefer, Jan .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2006, 135 (03) :374-382
[4]   Scala tympani cochleostomy I: Results of a survey [J].
Adunka, Oliver F. ;
Buchman, Craig A. .
LARYNGOSCOPE, 2007, 117 (12) :2187-2194
[5]   Conservation of residual acoustic hearing after cochlear implantation [J].
Balkany, Thomas J. ;
Connell, Sarah S. ;
Hodges, Annelle V. ;
Payne, Stacy L. ;
Telischi, Fred F. ;
Eshraghi, Adrien A. ;
Angeli, Simon I. ;
Germani, Ross ;
Messiah, Sarah ;
Arheart, Kristopher L. .
OTOLOGY & NEUROTOLOGY, 2006, 27 (08) :1083-1088
[6]  
Briggs R J, 2001, Cochlear Implants Int, V2, P135, DOI 10.1179/cim.2001.2.2.135
[7]  
Briggs RJS, 2011, 2011 COCHLEAR IMPLAN, V12, P130
[8]   Residual hearing conservation and electroacoustic stimulation with the nucleus 24 contour advance cochlear implant [J].
Fraysse, Bernard ;
Macias, Angel Ramos ;
Sterkers, Olivier ;
Burdo, Sandro ;
Ramsden, Richard ;
Deguine, Olivier ;
Klenzner, Thomas ;
Lenarz, Thomas ;
Rodriguez, Manuel Manrique ;
Von Wallenberg, Ernst ;
James, Chris .
OTOLOGY & NEUROTOLOGY, 2006, 27 (05) :624-633
[9]  
Hodges A, 1997, AM J OTOL, V18, P179
[10]  
LEHNHARDT E, 1993, HNO, V41, P356