Evaluation of nucleated red blood cell count by Sysmex XE-2100 in patients with thalassaemia or sickle cell anaemia and in neonates

被引:8
|
作者
Buoro, Sabrina [1 ]
Vavassori, Mauro [1 ]
Pipitone, Silvia [2 ]
Benegiamo, Anna [2 ]
Lochis, Eleonora [1 ]
Fumagalli, Sabina [1 ]
Falanga, Anna [3 ]
Marchetti, Marina [3 ]
Crippa, Alberto [1 ]
Ottomano, Cosimo [2 ]
Lippi, Giuseppe [2 ]
机构
[1] Papa Giovanni XXIII Hosp, Clin Chem Lab, Bergamo, Italy
[2] Univ Hosp Parma, Diagnost Blood Chem Unit, Parma, Italy
[3] Papa Giovanni XXIII Hosp, Immunohaematol & Transfus Med Unit, Bergamo, Italy
关键词
nucleated red blood cells; XE-2100; thalassaemia; sickle cell disease; neonates; ANALYZER;
D O I
10.2450/2015.0283-14
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Current haematology analysers have variable sensitivity and accuracy for counting nucleated red blood cells in samples with low values and in all those conditions characterised by altered sensitivity of red blood cells to the lysing process, such as in beta-thalassaemia or sickle-cell diseases and in neonates. The aim of our study was to evaluate the performance of the automated analyser XE-2100 at counting nucleated red blood cells in the above-mentioned three categories of subjects with potentially altered red blood cell lysis sensitivity and yet a need for accurate nucleated red blood cell counts. Materials and methods. We measured nucleated red blood cell count by XE-2100 in peripheral blood samples of 187 subjects comprising 55 patients with beta-thalassaemia (40 major and 15 traits), 26 sickle-cell patients, 56 neonates and 50 normal subject. Results were compared with those obtained by optical microscopy. Agreement between average values of the two methods was estimated by means of Pearson's correlation and bias analysis, whereas diagnostic accuracy was estimated by analysis of receiver operating characteristic curves. Results. The comparison between the two methods showed a Pearson's correlation of 0.99 (95% CI; 0.98-0.99; p < 0.001) and bias of -0.61 (95% CI, -1.5-0.3). The area under the curve of the nucleated red blood cell count in all samples was 0.98 (95% CI, 0.96-1.00; p < 0.001). Sub-analysis revealed an area under curve of 0.99 (95% CI, 0.98-1.00; p < 0.001) for patients with thalassaemia, 0.94 (95% CI, 0.85-1.00; p < 0.001) for patients with sickle cell anaemia, and 1.00 (95% CI, 1.0-1.0) for neonates. Discussion. XE-2100 has excellent performance for nucleated red blood cell counting, especially in critical populations such as patients with haemoglobinopathies and neonates.
引用
收藏
页码:588 / 594
页数:7
相关论文
共 50 条
  • [1] INTERFERENCE FROM LIPEMIA IN CELL BLOOD COUNT BY THE SYSMEX XE-2100
    Drieghe, S.
    Stove, V.
    Delanghe, J.
    ACTA CLINICA BELGICA, 2010, 65 (03): : 217 - 217
  • [2] An evaluation of the performance of Sysmex XE-2100 in enumerating nucleated red cells in peripheral blood
    Gulati, Gene
    Behling, Eric
    Kocher, William
    Schwarting, Roland
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2007, 131 (07) : 1077 - 1083
  • [3] Comparative performance efficiencies of the Abbott Cell-Dyn Sapphire and Sysmex XE-2100 hematology analyzers for nucleated red blood cell detection
    Brouwer, Rick
    Liesting, Eline
    Pranjic-Tumenko, Natasa
    Van Dongen-Lases, Edmee
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2007, 29 : 47 - 47
  • [4] Use of hematopoietic progenitor cell count on the Sysmex XE-2100 for peripheral blood stem cell harvest monitoring
    Letestu, Remi
    Marzac, Christophe
    Audat, Francoise
    Belhocine, Ramdane
    Tondeur, Sylvie
    Baccini, Veronique
    Garcon, Loic
    Dal Cortivo, Liliane
    Perrot, Jean-Yves
    Lefrere, Francois
    Valensi, Francoise
    Ajchenbaum-Cymbalista, Florence
    LEUKEMIA & LYMPHOMA, 2007, 48 (01) : 89 - 96
  • [5] Hematopoietic progenitor cell count on SYSMEX® XE-2100:: a new tool for peripheral blood stem cell harvest monitoring
    Letestu, R
    Audat, F
    Marzac, C
    Belhocine, R
    Baccini, V
    Garcon, L
    Louisy, D
    Persiaux, C
    Dal Cortivo, L
    Perrot, JY
    Lefrère, F
    Valensi, F
    Cymbalista, F
    BONE MARROW TRANSPLANTATION, 2003, 31 : S251 - S251
  • [6] CANDIDAEMIA INTERFERENCE IN THE WHITE BLOOD CELL AND NUCLEATED RED BLOOD CELL COUNT ON THE SYSMEX XE5000™ ANALYSER
    Seaux, Liesbeth
    Coucke, Line
    Mulliez, Sylvie
    Stove, Veronique
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2013, 35 : 60 - 61
  • [7] Automated analysis of bone marrow fluid with the Sysmex XE-2100™ blood cell counter
    Zini, G
    Mistretta, G
    Giordano, G
    Laurenti, L
    d'Onofrio, G
    INFUSION THERAPY AND TRANSFUSION MEDICINE-INFUSIONSTHERAPIE UND TRANSFUSIONSMEDIZIN, 2001, 28 (05): : 277 - 279
  • [8] New red cell parameters on the Sysmex XE-2100™ as potential markers of functional iron deficiency
    Briggs, C
    Rogers, R
    Thompson, B
    Machin, SJ
    INFUSION THERAPY AND TRANSFUSION MEDICINE-INFUSIONSTHERAPIE UND TRANSFUSIONSMEDIZIN, 2001, 28 (05): : 256 - 262
  • [9] The new hematology analyzer Sysmex XE-2100 - Performance evaluation of a novel white blood cell differential technology
    Ruzicka, K
    Veitl, M
    Thalhammer-Scherrer, R
    Schwarzinger, I
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2001, 125 (03) : 391 - 396
  • [10] EVALUATION OF SYSMEX XN-2000 AND XE-2100 IN UMBILICAL CORD BLOOD
    Kim, Hanah
    Hur, Mina
    Choi, Sang-Gyeu
    Moon, Hee-Won
    Yun, Yeo-Min
    Hwang, Han Sung
    Kwon, Han Sung
    Sohn, In Sook
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 : 33 - 34