Thickness of the Macula, Retinal Nerve Fiber Layer, and Ganglion Cell Layer in the Epiretinal Membrane: The Repeatability Study of Optical Coherence Tomography

被引:35
作者
Lee, Haeng-Jin [1 ]
Kim, Min-Su [1 ]
Jo, Young-Joon [1 ,2 ]
Kim, Jung-Yeul [1 ,2 ]
机构
[1] Chungnam Natl Univ, Dept Ophthalmol, Coll Med, Daejeon, South Korea
[2] Chungnam Natl Univ, Res Inst Med Sci, Coll Med, Daejeon, South Korea
关键词
epiretinal membrane; ganglion cell inner plexiform layer; repeatability; GLAUCOMATOUS EYES; REPRODUCIBILITY; SURGERY; CIRRUS; OCT;
D O I
10.1167/iovs.15-16949
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To analyze the repeatability of measurements of the thicknesses of the macula, retinal nerve fiber layer (RNFL), and ganglion cell inner plexiform layer (GCIPL) using spectral-domain optical coherence tomography (SD-OCT) in the epiretinal membrane (ERM). METHODS. The prospective study analyzed patients who visited our retinal clinic from June 2013 to January 2014. An experienced examiner measured the thicknesses twice using macular cube 512 x 128 and optic disc cube 200 3 200 scans. The repeatability of the thicknesses of the macula, RNFL, and GCIPL were compared using the intraclass correlation coefficient (ICC) of two groups based on the central macular thickness (group A, <= 450 mu m; group B, > 450 mu m). RESULTS. A total of 88 patients were analyzed. The average thicknesses of the central macula, RNFL, and GCIPL were 256.5, 96.6, and 84.4 mu m, respectively, in the normal fellow eye and 412.3, 94.6, and 56.7 mu m in the affected eye. The ICCs of the central macula, RNFL, and GCIPL were 0.995, 0.994, and 0.996, respectively, for the normal fellow eye and 0.991, 0.973, and 0.881 for the affected eye. The average thicknesses of the central macula, RNFL, and GCIPL in group A were 360.9, 93.5, and 63.4 mu m, respectively, and the ICCs were 0.997, 0.987, and 0.995. The thicknesses in group B were 489.5, 96.2, and 46.6 mu m, respectively, and the ICCs were 0.910, 0.942, and 0.603, significantly lower repeatability compared with group A (P < 0.05). CONCLUSIONS. The macular contour change with the ERM results in low repeatability and tendency to be measured thinner in GCIPL thickness using SD-OCT. This can be explained by the unstable gaze of the patient due to decreased visual acuity and autosegmentation error following changes in the macula.
引用
收藏
页码:4554 / 4559
页数:6
相关论文
共 50 条
[11]   Ganglion Cell-Inner Plexiform Layer Thickness in Retinal Diseases: Repeatability Study of Spectral-Domain Optical Coherence Tomography [J].
Lee, Haeng-Jin ;
Kim, Min-Su ;
Jo, Young-Joon ;
Kim, Jung-Yeul .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2015, 160 (02) :283-289
[12]   Estimating retinal nerve fiber layer thickness in normal schoolchildren with spectral-domain optical coherence tomography [J].
Tsai, Der-Chong ;
Huang, Nicole ;
Hwu, Jinn-Jong ;
Jueng, Ruo-Nan ;
Chou, Pesus .
JAPANESE JOURNAL OF OPHTHALMOLOGY, 2012, 56 (04) :362-370
[13]   Evaluation of the reproducibility of retinal nerve fiber layer thickness and optic nerve head parameters with optical coherence tomography [J].
Rau, Ricardo ;
de Freitas Silva, Andre Luiz ;
Alves Salame, Andre Luiz ;
Soares Melo Junior, Luiz Alberto .
REVISTA BRASILEIRA DE OFTALMOLOGIA, 2012, 71 (05) :280-284
[14]   Inter-Session Repeatability of Retinal Layer Thickness in Optical Coherence Tomography [J].
DeBuc, D. Cabrera ;
Tian, Jing ;
Bates, Nathan ;
Somfai, Gabor Mark .
MEDICAL IMAGING 2017: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2017, 10137
[15]   Spectral Domain Optical Coherence Tomography of Ganglion Cell Complex and Retinal Nerve Fiber Layer in Multiple Sclerosis [J].
Calafiore, Silvia ;
Scuderi, Gianluca ;
Martino, Francesco ;
Ducoli, Giorgio ;
Delle Fave, Martina Maria ;
Perdicchi, Andrea .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
[16]   Negative correlation between retinal displacement and ganglion cell layer thickness changes in eyes with epiretinal membrane [J].
Loiudice, Pasquale ;
Pellegrini, Marco ;
Montesel, Andrea ;
Nardi, Marco ;
Peiretti, Enrico ;
De Cilla, Stefano ;
Posarelli, Chiara ;
Figus, Michele ;
Casini, Giamberto .
EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2020, 30 (06) :1424-1431
[17]   Retinal Nerve Fiber Layer Configuration in Eyes with Epiretinal Membrane [J].
Oh, Jaeryung ;
Oh, Jong-Hyun ;
Do, Jae Rock ;
Chang, Minwook ;
Park, Choul Yong .
OPTOMETRY AND VISION SCIENCE, 2014, 91 (11) :1328-1334
[18]   Evaluation of retinal nerve fibre layer, ganglion cell layer and choroidal thickness with optical coherence tomography in migraine patients: a case-control study [J].
Gunes, Aygul ;
Karadag, Ayse Sevgi ;
Yazgan, Serpil ;
Celik, Haci Ugur ;
Simsek, Ali .
CLINICAL AND EXPERIMENTAL OPTOMETRY, 2018, 101 (01) :109-115
[19]   Peripapillary retinal nerve fiber layer thickness in normal Japanese eyes measured with optical coherence tomography [J].
Kanno, Makoto ;
Nagasawa, Michi ;
Suzuki, Michiro ;
Yamashita, Hidetoshi .
JAPANESE JOURNAL OF OPHTHALMOLOGY, 2010, 54 (01) :36-42
[20]   Comparison of Retinal Nerve Fiber Layer Thickness Measured by Cirrus HD and Stratus Optical Coherence Tomography [J].
Sung, Kyung Rim ;
Kim, Dong Yoon ;
Park, Sung Bae ;
Kook, Michael S. .
OPHTHALMOLOGY, 2009, 116 (07) :1264-1270