Influence of lens aberrations, specimen thickness and tilt on differential phase contrast STEM images

被引:19
作者
Buerger, Julius [1 ,2 ,3 ]
Riedl, Thomas [1 ,2 ,3 ]
Lindner, Joerg K. N. [1 ,2 ,3 ]
机构
[1] Paderborn Univ, Dept Phys, Warburger Str 100, D-33098 Paderborn, Germany
[2] Paderborn Univ, Ctr Optoelect & Photon Paderborn CeOPP, D-33098 Paderborn, Germany
[3] Paderborn Univ, Inst Lightweight Design Hybrid Mat ILH, D-33098 Paderborn, Germany
关键词
Differential phase contrast; Scanning transmission electron microscopy; Atomic resolution; Lens aberration; C-s correction; Simulation; ELECTRIC-FIELDS; DETECTOR; MICROSCOPY; CRYSTALS;
D O I
10.1016/j.ultramic.2020.113118
中图分类号
TH742 [显微镜];
学科分类号
摘要
Differential phase contrast (DPC) imaging in scanning transmission electron microscopy (STEM) allows for measuring electric and magnetic fields in solids on scales ranging from picometres to micrometres. The DPC technique mainly uses the direct beam, which is deflected by the electric and magnetic fields of the specimen and measured with a beam position sensitive detector. The beam deflection and thus the DPC signal is strongly influenced by specimen thickness, specimen tilt and lens aberrations. Understanding these influences is critical for a solid interpretation and quantification of contrasts in DPC images. To this end, the present study employs DPC-STEM image simulations of SrTiO3 [001] at atomic resolution to analyse the influence of lens aberrations, specimen tilt and thickness and also to give a guideline for the detection of parameters affecting the contrast by performing an analysis of associated scattergrams. Simulations are obtained using the multislice algorithm implemented in the Dr. Probe software with conditions corresponding to a JEOL ARM200F microscope equipped with an octa-segmented annular detector, but results should be similar for other microscopes. Simulations show that due to a non-rigid shift of the detected intensity distribution correct values of projected potentials of specimens thicker than one unit-cell cannot be determined. Regarding the impact of residual lens aberrations, it is found that the shape of the lens aberration phase function determines the symmetry and features in the DPC image. Specimen tilt leads to an elongation of features perpendicular to the tilt axis. The results are confirmed by comparing simulated with experimental DPC images of Si [110] yielding good agreement. Overall, a high sensitivity of DPC-STEM imaging to lens aberrations, specimen tilt and diffraction effects is evidenced.
引用
收藏
页数:14
相关论文
共 39 条
[1]   Thickness and defocus dependence of inter-atomic electric fields measured by scanning diffraction [J].
Addiego, Christopher ;
Gao, Wenpei ;
Pan, Xiaoqing .
ULTRAMICROSCOPY, 2020, 208
[2]   Dr. Probe: A software for high-resolution STEM image simulation [J].
Barthel, J. .
ULTRAMICROSCOPY, 2018, 193 :1-11
[3]   Measuring nanometre-scale electric fields in scanning transmission electron microscopy using segmented detectors [J].
Brown, H. G. ;
Shibata, N. ;
Sasaki, H. ;
Petersen, T. C. ;
Paganin, D. M. ;
Morgan, M. J. ;
Findlay, S. D. .
ULTRAMICROSCOPY, 2017, 182 :169-178
[4]  
Brown H.G., 2017, MICROSC MICROANAL, V23, P442
[5]   Towards quantitative, atomic-resolution reconstruction of the electrostatic potential via differential phase contrast using electrons [J].
Close, R. ;
Chen, Z. ;
Shibata, N. ;
Findlay, S. D. .
ULTRAMICROSCOPY, 2015, 159 :124-137
[6]   THE SCATTERING OF ELECTRONS BY ATOMS AND CRYSTALS .1. A NEW THEORETICAL APPROACH [J].
COWLEY, JM ;
MOODIE, AF .
ACTA CRYSTALLOGRAPHICA, 1957, 10 (10) :609-619
[7]  
Egerton R.F., 2011, ELECT ENERGY LOSS SP
[8]   Atomic electrostatic maps of 1D channels in 2D semiconductors using 4D scanning transmission electron microscopy [J].
Fang, Shiang ;
Wen, Yi ;
Allen, Christopher S. ;
Ophus, Colin ;
Han, Grace G. D. ;
Kirkland, Angus I. ;
Kaxiras, Efthimios ;
Warner, Jamie H. .
NATURE COMMUNICATIONS, 2019, 10 (1)
[9]   Picometer-scale atom position analysis in annular bright-field STEM imaging [J].
Gao, Peng ;
Kumamoto, Akihito ;
Ishikawa, Ryo ;
Lugg, Nathan ;
Shibata, Naoya ;
Ikuhara, Yuichi .
ULTRAMICROSCOPY, 2018, 184 :177-187
[10]   Upper limits for the residual aberrations of a high-resolution aberration-corrected STEM [J].
Haider, M ;
Uhlemann, S ;
Zach, J .
ULTRAMICROSCOPY, 2000, 81 (3-4) :163-175