Development of a Hemodynamically Optimized Outflow Cannula for Cardiopulmonary Bypass

被引:16
作者
Kaufmann, Tim A. S. [1 ]
Schlanstein, Peter [1 ]
Moritz, Anton [2 ]
Steinseifer, Ulrich [1 ]
机构
[1] Rhein Westfal TH Aachen, Inst Appl Med Engn, Dept Cardiovasc Engn, D-52074 Aachen, Germany
[2] Goethe Univ Frankfurt, D-60054 Frankfurt, Germany
关键词
Computational fluid dynamics; Cannulation; Cardiopulmonary bypass; Cannula design; COMPUTATIONAL FLUID-DYNAMICS; CEREBRAL AUTOREGULATION; EXTRACORPOREAL-CIRCULATION; AORTIC CANNULA; BLOOD DAMAGE; FLOW; PERFUSION; SURGERY; IMPACT; ARCH;
D O I
10.1111/aor.12262
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The jet of the outflow cannula is a potential risk for patients undergoing cardiopulmonary bypass (CPB), because increased jet velocities lead to altered flow conditions and might furthermore mobilize atherosclerotic plaques from calcified aortas. The cannula jet is therefore among the main reasons for cerebral hypoxia and stroke in CPB patients. In the past, we developed a validated computational fluid dynamics (CFD) model to analyze flow conditions during CPB as dependent on cannulation and support modalities. This model is now applied to develop a novel CPB outflow cannula to reduce the jet effect and increase cerebral blood flow. The Multi-Module Cannula (MMC) is based on a generic elbow cannula that was iteratively improved. It features an inner wall to smoothly guide the blood as well as an elliptically shaped outlet diffuser. During standard CPB conditions of 5L/min, the pressure drop over the MMC is 61mmHg, compared with 68mmHg with a standard cannula. The maximum velocities are decreased from 3.7m/s to 3.3m/s. In the cannula jet of the MMC, the velocities are reduced further, down to 1.6m/s. The cerebral blood flow is typically reduced during CPB. Using the MMC, however, it reaches almost physiological values at 715mL/min. These results suggest that the MMC outperforms standard CPB cannulas. Further design improvements and improved insertion techniques are under consideration.
引用
收藏
页码:972 / 978
页数:7
相关论文
共 23 条
[1]   Pulsatile extracorporeal circulation during on-pump cardiac surgery enhances aortic wall shear stress [J].
Assmann, Alexander ;
Benim, Ali Cemal ;
Guel, Fethi ;
Lux, Philipp ;
Akhyari, Payam ;
Boeken, Udo ;
Joos, Franz ;
Feindte, Peter ;
Lichtenberg, Artur .
JOURNAL OF BIOMECHANICS, 2012, 45 (01) :156-163
[2]   Simulation of blood flow in human aorta with emphasis on outlet boundary conditions [J].
Benim, A. C. ;
Nahavandi, A. ;
Assmann, A. ;
Schubert, D. ;
Feindt, P. ;
Suh, S. H. .
APPLIED MATHEMATICAL MODELLING, 2011, 35 (07) :3175-3188
[3]   MODEL FOR A GENERAL MECHANICAL BLOOD DAMAGE PREDICTION [J].
BLUDSZUWEIT, C .
ARTIFICIAL ORGANS, 1995, 19 (07) :583-589
[4]   Experimental Fluid Dynamics of Transventricular Apical Aortic Cannulation [J].
Fukuda, Ikuo ;
Yanaoka, Hideki ;
Inamura, Takao ;
Minakawa, Masahito ;
Daitoku, Kazuyuki ;
Suzuki, Yasuyuki .
ARTIFICIAL ORGANS, 2010, 34 (03) :222-223
[5]   Computer-simulated fluid dynamics of arterial perfusion in extracorporeal circulation: From reality to virtual simulation [J].
Fukuda, Ikuo ;
Osanai, Satoshi ;
Shirota, Minori ;
Inamura, Takao ;
Yanaoka, Hideki ;
Minakawa, Masahito ;
Fukui, Kozo .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2009, 32 (06) :362-370
[6]   On the Use of In Vivo Measured Flow Rates as Boundary Conditions for Image-Based Hemodynamic Models of the Human Aorta: Implications for Indicators of Abnormal Flow [J].
Gallo, D. ;
De Santis, G. ;
Negri, F. ;
Tresoldi, D. ;
Ponzini, R. ;
Massai, D. ;
Deriu, M. A. ;
Segers, P. ;
Verhegghe, B. ;
Rizzo, G. ;
Morbiducci, U. .
ANNALS OF BIOMEDICAL ENGINEERING, 2012, 40 (03) :729-741
[7]  
HEUSER G, 1980, BIORHEOLOGY, V17, P17
[8]   The impact of aortic manipulation on neurologic outcomes after coronary artery bypass surgery: A risk-adjusted study [J].
Kapetanakis, EI ;
Stamou, SC ;
Dullum, MKC ;
Hill, PC ;
Haile, E ;
Boyce, SW ;
Bafi, AS ;
Petro, KR ;
Corso, PJ .
ANNALS OF THORACIC SURGERY, 2004, 78 (05) :1564-1571
[9]   Implementation of Cerebral Autoregulation Into Computational Fluid Dynamics Studies of Cardiopulmonary Bypass [J].
Kaufmann, Tim A. S. ;
Schmitz-Rode, Thomas ;
Steinseifer, Ulrich .
ARTIFICIAL ORGANS, 2012, 36 (08) :754-758
[10]   Mimicking of Cerebral Autoregulation by Flow-Dependent Cerebrovascular Resistance: A Feasibility Study [J].
Kaufmann, Tim A. S. ;
Wong, Kai C. ;
Schmitz-Rode, Thomas ;
Steinseifer, Ulrich .
ARTIFICIAL ORGANS, 2012, 36 (04) :E97-E101