Diagnosis of inherited von Willebrand disease: comparison of two methodologies and analysis of the discrepancies

被引:29
作者
Costa-Pinto, J. [1 ]
Perez-Rodriguez, A. [1 ]
Gomez-del-Castillo, M. Del C. [1 ]
Loures, E. [1 ]
Rodriguez-Trillo, A. [1 ]
Batlle, J. [1 ]
Lopez-Fernandez, M. F. [1 ]
机构
[1] Univ Santiago de Compostela, Serv Hematol & Hemoterapia, Xerencia Xest Integrada A Coruna, Dept Med,Sch Med, La Coruna, Spain
关键词
diagnosis management; gene mutations; multimeric analysis; ristocetin cofactor; von Willebrand disease; von Willebrand factor; RISTOCETIN COFACTOR ACTIVITY; IMPROVED PERFORMANCE-CHARACTERISTICS; VONWILLEBRAND-FACTOR MULTIMERS; AUTOMATED-ASSAY; CHEMILUMINESCENCE; CLASSIFICATION; HETEROGENEITY; IMMUNOASSAY; VALIDATION; MANAGEMENT;
D O I
10.1111/hae.12380
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Diagnostics of von Willebrand disease (VWD) includes assessment of factor VIII (FVIII) coagulant activity, von Willebrand factor (VWF) antigen (VWF:Ag) and VWF ristocetin cofactor activity (VWF:RCo), and more specific tests as multimeric and genetic analyses are necessary for the correct VWD classification. The ACL AcuStar analyzer introduces chemiluminescence (CL) technology in detection of VWD with automated VWF:Ag and VWF:RCo assays. Compare VWF:Ag-ELISA and VWF:RCo by aggregometry conventional assays with new CL VWF:Ag-IL and VWF:RCo-IL assays, investigate the ability to make accurate VWD diagnosis and concordance with multimeric and genetic analyses. 146 patients with congenital VWD (51 Type 1; 34 Type2A; 16 Type 2B; 31 Type 2M; 5 Type 2N; 9 Type 3) and 30 healthy normal subjects were included. A comparison was made between CL and conventional methods. Diagnostic evaluation included: VWF:RCo/VWF:Ag ratio, multimeric distribution (sodium dodecyl sulfate [SDS]-agarose gel) of VWF and genetic analysis in 110 of 146 patients. CL and conventional methods revealed good correlation. Kappa test agreement diagnosis was >0.8. CL diagnostic sensitivity was 100% and specificity 97%. Multimeric and genetic analysis were of help in clarifying 13 discrepancies of diagnosis between methods, of which six discrepancies were explained by lack of conventional methods sensibility. CL methodology can detect VWD and discriminate between type 1, 3 and variant forms and offers an automated, faster, sensitive and less cumbersome method when compared to conventional assays, in particular VWF:RCo by aggregometry. In some cases, even with all phenotype and genetic analyses, discrepancies exist in the classification of VWD.
引用
收藏
页码:559 / 567
页数:9
相关论文
共 39 条
[1]  
Batlle J, 2008, J THROMB HAEMOST, V6, P388, DOI 10.1111/j.1538-7836.2008.02853.x
[2]  
BATLLE J, 1986, BLOOD, V68, P1207
[3]   ELECTROBLOTTING OF FACTOR-VIII VONWILLEBRAND-FACTOR MULTIMERS AFTER ELECTROPHORESIS IN SDS-AGAROSE GEL, DISCONTINUOUS BUFFER SYSTEM [J].
BATLLE, J ;
LOPEZFERNANDEZ, MF ;
LOPEZBERGES, C ;
LOPEZBORRASCA, A .
BRITISH JOURNAL OF HAEMATOLOGY, 1983, 55 (01) :195-196
[4]  
Batlle J., 2011, Von Willebrand Disease. Basic and Clinical Aspects, P74
[5]  
Bodo I., 2012, ECAT FDN, P27
[6]  
BUDDE U, 1990, THROMB HAEMOSTASIS, V63, P312
[7]   Comparison of a new chemiluminescent immunoassay for von Willebrand factor activity with the ristocetin cofactor-induced platelet agglutination method [J].
Cabrera, N. ;
Moret, A. ;
Caunedo, P. ;
Cid, A. R. ;
Vila, V. ;
Espana, F. ;
Aznar, J. A. .
HAEMOPHILIA, 2013, 19 (06) :920-925
[8]   Principles of care for the diagnosis and treatment of von Willebrand disease [J].
Castaman, Giancarlo ;
Goodeve, Anne ;
Eikenboom, Jeroen .
HAEMATOLOGICA, 2013, 98 (05) :667-674
[9]  
Chen D, 2008, J THROMB HAEMOST, V6, P323, DOI 10.1111/j.1538-7836.2008.02845.x
[10]   Detection of Botulinum Neurotoxin Serotypes A and B Using a Chemiluminescent versus Electrochemiluminescent Immunoassay in Food and Serum [J].
Cheng, Luisa W. ;
Stanker, Larry H. .
JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 2013, 61 (03) :755-760