IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry

被引:7
|
作者
Zhang, Yang [1 ]
Zheng, Lin [1 ]
Bian, Ailing [1 ]
Zhou, Qi [1 ]
机构
[1] Chinese Acad Med Sci, Dept Ophthalmol, Peking Union Med Coll Hosp, Peking Union Med Coll, Dongdan Shuaifuyuan 1, Beijing 100730, Peoples R China
关键词
Intraocular pressure; Silicone oil tamponade; Goldmann applanation tonometry; Corvis ST tonometer; Corneal biomechanical properties; INTRAOCULAR-PRESSURE; BIOMECHANICAL PROPERTIES; CORNEAL BIOMECHANICS; GLAUCOMA; DEFORMATION; PARAMETERS; THICKNESS; IMPACT;
D O I
10.1007/s10792-017-0519-1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To compare the postoperative intraocular pressure (IOP) of eyes following pars plana vitrectomy (PPV) combined with intravitreal silicone oil (SO) tamponade by Corneal Visualization Scheimpflug Technology (CST), Goldmann applanation tonometry (GAT) and non-contact tonometry (NCT). Methods Thirty-eight participants who had undergone PPV combined with SO tamponade to treat vitreoretinal diseases were enrolled. Postoperative IOP measurements were obtained using CST, NCT and GAT. Inter-device agreement was assessed by Bland-Altman analysis. The correlation coefficient was used to describe the potential postoperative factors affecting the postoperative IOP differences between each device. Results Bland-Altman analysis revealed the bias between CST and GAT, between CST and NCT, and between GAT and NCT to be -0.2, 2.1 and 2.4 mmHg, respectively. CST and GAT correlated well with each other. NCT values were lower than those of GAT and CST (all p < 0.05), whereas CST values did not differ from the GAT readings. Central corneal thickness, corneal biomechanical properties and age showed significant correlation with the differences of CST-NCT and GAT-NCT. Conclusions In SO tamponade eyes, NCT obtains lower IOP than other tonometry techniques, and CST is highly consistent with GAT. CST offers an optional non-contact method for measuring postoperative IOP in SO tamponade eyes.
引用
收藏
页码:697 / 703
页数:7
相关论文
共 50 条
  • [41] Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry
    Li, Yuehua
    Shi, Jingming
    Duan, Xuanchu
    Fan, Fang
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (12) : 1765 - 1770
  • [42] Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry
    Yuehua Li
    Jingming Shi
    Xuanchu Duan
    Fang Fan
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2010, 248 : 1765 - 1770
  • [43] A clinical comparison of the Xpert non-contact tonometer with the Goldmann applanation tonometer after penetrating keratoplasty
    Lisle, C
    Ehlers, N
    ACTA OPHTHALMOLOGICA SCANDINAVICA, 2000, 78 (02): : 211 - 215
  • [44] The impact of silicone hydrogel contact lenses on the measurement of intraocular pressure using non-contact tonometry
    Pesic, Snezana
    Jovanovic, Svetlana
    Mitrasevic, Milos
    Vuletic, Biljana
    Jovanovic, Milena
    Jovanovic, Zorica
    VOJNOSANITETSKI PREGLED, 2017, 74 (08) : 763 - 766
  • [45] Non-contact tonometry using Corvis ST: analysis of corneal vibrations and their relation with intraocular pressure
    Boszczyk, Agnieszka
    Kasprzak, Henryk
    Siedlecki, Damian
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2019, 36 (04) : B28 - B34
  • [46] Comparison of Biomechanically Corrected Intraocular Pressure Obtained by Corvis ST and Goldmann Applanation Tonometry in Patients With Open-angle Glaucoma and Ocular Hypertension
    Ye, Yiming
    Yang, Yangfan
    Fan, Yanmei
    Lan, Min
    Yu, Keming
    Yu, Minbin
    JOURNAL OF GLAUCOMA, 2019, 28 (10) : 922 - 928
  • [47] Measurement of intraocular pressure using the Tono-Pen in comparison with Goldmann applanation tonometry - a clinical study in 100 eyes
    Deuter, CME
    Schlote, T
    Hahn, GA
    Bende, T
    Derse, M
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2002, 219 (03) : 138 - 142
  • [48] Transpalpebral measurement of intraocular pressure using the TGDc-01 tonometer versus standard Goldmann applanation tonometry
    Annette Lösch
    Alexander Scheuerle
    Volker Rupp
    Gerd Auffarth
    Matthias Becker
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2005, 243 : 313 - 316
  • [49] Agreement among Goldmann applanation tonometer, iCare, and Icare PRO rebound tonometers; non-contact tonometer; and Tonopen XL in healthy elderly subjects
    Kato, Yoshitake
    Nakakura, Shunsuke
    Matsuo, Naoko
    Yoshitomi, Kayo
    Handa, Marina
    Tabuchi, Hitoshi
    Kiuchi, Yoshiaki
    INTERNATIONAL OPHTHALMOLOGY, 2018, 38 (02) : 687 - 696
  • [50] Comparison of intraocular pressure measurements obtained by icare pro tonometer, non-contact tonometer and Goldmann applanation tonometer in healthy individuals
    Duran, M.
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2023, 46 (10): : 1195 - 1203