IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry

被引:7
|
作者
Zhang, Yang [1 ]
Zheng, Lin [1 ]
Bian, Ailing [1 ]
Zhou, Qi [1 ]
机构
[1] Chinese Acad Med Sci, Dept Ophthalmol, Peking Union Med Coll Hosp, Peking Union Med Coll, Dongdan Shuaifuyuan 1, Beijing 100730, Peoples R China
关键词
Intraocular pressure; Silicone oil tamponade; Goldmann applanation tonometry; Corvis ST tonometer; Corneal biomechanical properties; INTRAOCULAR-PRESSURE; BIOMECHANICAL PROPERTIES; CORNEAL BIOMECHANICS; GLAUCOMA; DEFORMATION; PARAMETERS; THICKNESS; IMPACT;
D O I
10.1007/s10792-017-0519-1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To compare the postoperative intraocular pressure (IOP) of eyes following pars plana vitrectomy (PPV) combined with intravitreal silicone oil (SO) tamponade by Corneal Visualization Scheimpflug Technology (CST), Goldmann applanation tonometry (GAT) and non-contact tonometry (NCT). Methods Thirty-eight participants who had undergone PPV combined with SO tamponade to treat vitreoretinal diseases were enrolled. Postoperative IOP measurements were obtained using CST, NCT and GAT. Inter-device agreement was assessed by Bland-Altman analysis. The correlation coefficient was used to describe the potential postoperative factors affecting the postoperative IOP differences between each device. Results Bland-Altman analysis revealed the bias between CST and GAT, between CST and NCT, and between GAT and NCT to be -0.2, 2.1 and 2.4 mmHg, respectively. CST and GAT correlated well with each other. NCT values were lower than those of GAT and CST (all p < 0.05), whereas CST values did not differ from the GAT readings. Central corneal thickness, corneal biomechanical properties and age showed significant correlation with the differences of CST-NCT and GAT-NCT. Conclusions In SO tamponade eyes, NCT obtains lower IOP than other tonometry techniques, and CST is highly consistent with GAT. CST offers an optional non-contact method for measuring postoperative IOP in SO tamponade eyes.
引用
收藏
页码:697 / 703
页数:7
相关论文
共 50 条
  • [21] Comparison of Intraocular Pressure Measurements Using Three Different Methods (Goldmann Applanation Tonometry (GAT), Corvis ST, and iCare) Following Penetrating Keratoplasty
    Tziola, Tatiana
    Tzamalis, Argyrios
    Koronis, Spyridon
    Garitsis, Panagiotis
    Tsinopoulos, Ioannis
    Ziakas, Nikolaos
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (23)
  • [22] Influence of Corneal Thickness in Keratoconic Corneas on IOP Measurement with IOPen, iCare, Dynamic Contour Tonometry and Goldmann Applanation Tonometry
    Klamann, M. K. J.
    Maier, A. -K. B.
    Gonnermann, J.
    Ruokonen, P.
    Bertelmann, E.
    Torun, N.
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2013, 230 (07) : 697 - 700
  • [23] Comparison of Goldmann Applanation Tonometry and Dynamic Contour Tonometry in the Measurement of Intraocular Pressure in Eyes with Different Corneal Thicknesses
    Roszkowska, A. M.
    De Grazia, L.
    Cirone, M.
    Ferreri, G.
    OPHTHALMOLOGICA, 2009, 223 (04) : 244 - 249
  • [24] Corneal Modulus and IOP Measurements in Canine Eyes Using Goldmann Applanation Tonometry and Tono-pen
    Tang, Junhua
    Pan, Xueliang
    Weber, Paul A.
    Liu, Jun
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (11) : 7866 - 7871
  • [25] The accuracy of Goldmann applanation tonometry over silicone hydrogel contact lenses in patients with glaucoma
    Gomes, Beatriz A. F.
    Turiel, Paulo
    Cavalcanti, Rachel S.
    Moraes, Haroldo V., Jr.
    Santhiago, Marcony R.
    CONTACT LENS & ANTERIOR EYE, 2014, 37 (04) : 297 - 299
  • [26] Goldmann Applanation Tonometry versus Dynamic Contour Tonometry after Vitrectomy with Silicone Oil Endotamponade
    Fuest, Matthias
    Mamas, Nikolaos
    Walter, Peter
    Mazinani, Babac E.
    Roessler, Gemot
    Plange, Niklas
    CURRENT EYE RESEARCH, 2017, 42 (07) : 1007 - 1012
  • [27] Inter-optometrist variability of IOP measurement for modern tonometers and their agreement with Goldmann Applanation Tonometry
    Campbell, Peter
    Edgar, David F.
    Shah, Rakhee
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2021, 104 (05) : 602 - 610
  • [28] Comparison among Ocular Response Analyzer,Corvis ST and Goldmann applanation tonometry in healthy children
    Ramin Salouti
    Ali agha Alishiri
    Reza Gharebaghi
    Mostafa Naderi
    Khosrow Jadidi
    Ahmad Shojaei-Baghini
    Mohammadreza Talebnejad
    Zahra Nasiri
    Seyedmorteza Hosseini
    Fatemeh Heidary
    International Journal of Ophthalmology, 2018, 11 (08) : 1330 - 1336
  • [29] Goldmann applanation tonometry and dynamic contour tonometry in eyes with elevated intraocular pressure (IOP): comparison in the same eyes after subsequent medical normalization of IOP
    Yoo, Chungkwon
    Eom, Young Sub
    Kim, Yong Yeon
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (11) : 1611 - 1616
  • [30] Difference in intraocular pressure measurements between non-contact tonometry and Goldmann applanation tonometry and the role of central corneal thickness in affecting glaucoma referrals
    Kamel, Khalid
    Dervan, Edward
    Falzon, Kevin
    O'Brien, Colm
    IRISH JOURNAL OF MEDICAL SCIENCE, 2019, 188 (01) : 321 - 325