Enhancing Pedicle Screw Fixation in the Lumbar Spine Using Allograft Bone Plug Interference Fixation

被引:9
作者
Chrea, Bopha [1 ]
Malempati, Harsha [2 ]
Campbell, Jeffrey R. [1 ]
Khan, Sonja [1 ]
Ching, Randal P. [1 ]
Lee, Michael J. [2 ]
机构
[1] Univ Washington, Appl Biomech Lab, Dept Mech Engn, Seattle, WA 98109 USA
[2] Univ Washington, Dept Orthopaed & Sports Med, Seattle, WA 98109 USA
来源
JOURNAL OF SPINAL DISORDERS & TECHNIQUES | 2014年 / 27卷 / 03期
关键词
pedicle screw; spine fixation; osteoporosis; lumbar spine; VERTEBROPLASTY AUGMENTATION; PULLOUT STRENGTH; MINERAL DENSITY; POLYMETHYLMETHACRYLATE; INSTRUMENTATION;
D O I
10.1097/BSD.0b013e318290fc84
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design:A within-subjects controlled laboratory study.Objective:To examine a biological alternative to cement augmentation for pedicle screw fixation comparing bilateral axial pullout tests of augmented and nonaugmented (controls) pedicle screws.Summary of Background Data:Fixation in the osteoporotic spine remains a difficult challenge with failure by loosening or backout. Pedicle screw augmentation has been attempted using polymethylmethacrylate and bioabsorbable calcium cements; however, the potential for extravasation and embolization of cement are becoming increasingly concerning and merit the search for alternative methods to improve screw-anchoring strength.Methods:Twenty-four (24) fresh human lumbar vertebrae were tested to compare the pullout strength of augmented and nonaugmented pedicle screws. Two different augmentation strategies were employed using allograft bone plugs (ABPs) and evaluated using 12 specimens per group. Bone mineral density of each specimen was obtained using dual-energy x-ray absorptiometry. The augmented versus nonaugmented pedicle was randomized for each vertebra, and bilateral testing enabled paired statistical analyses. Axial pullout tests were performed using an materials testing system servohydraulic test system, and peak force, failure displacement, and stiffness was obtained for each test and correlated with bone mineral density.Results:Augmentation using 6-mm-diameter ABPs with 6.25-mm-diameter pedicle screws resulted in statistically weaker average pullout strength (775455 N) than the nonaugmented controls (1233 +/- 826 N). When using smaller (5 mm diameter) AGPs with the same diameter screws, there was no statistical difference between average pullout strength for the augmented pedicle screws (1772 +/- 652 N) and the nonaugmented screws (1780 +/- 575 N).Conclusions:Preliminary study of pedicle screw augmentation using cannulated ABPs showed no improvement of fixation with pedicles in the spine. This was even true in osteoporotic specimens, where augmentation would seem to be of considerable benefit.
引用
收藏
页码:E81 / E87
页数:7
相关论文
共 24 条
[1]   The Pedicle Screw Fixation With Vertebroplasty Augmentation in the Surgical Treatment of the Severe Osteoporotic Spines [J].
Aydogan, Mehmet ;
Ozturk, Cagatay ;
Karatoprak, Omer ;
Tezer, Mehmet ;
Aksu, Neslihan ;
Hamzaoglu, Azmi .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2009, 22 (06) :444-447
[2]   Assessment of different screw augmentation techniques and screw designs in osteoporotic spines [J].
Becker, S. ;
Chavanne, A. ;
Spitaler, R. ;
Kropik, K. ;
Aigner, N. ;
Ogon, M. ;
Redl, H. .
EUROPEAN SPINE JOURNAL, 2008, 17 (11) :1462-1469
[3]   Biomechanical Analysis of Pedicle Screws in Osteoporotic Bone With Bioactive Cement Augmentation Using Simulated In Vivo Multicomponent Loading [J].
Choma, Theodore J. ;
Frevert, Wesley F. ;
Carson, William L. ;
Waters, Nicole P. ;
Pfeiffer, Ferris M. .
SPINE, 2011, 36 (06) :454-462
[4]  
Cook Stephen D, 2004, Spine J, V4, P402, DOI 10.1016/j.spinee.2003.11.010
[5]   EFFECTS OF BONE-MINERAL DENSITY ON PEDICLE SCREW FIXATION [J].
HALVORSON, TL ;
KELLEY, LA ;
THOMAS, KA ;
WHITECLOUD, TS ;
COOK, SD .
SPINE, 1994, 19 (21) :2415-2420
[6]   Major neurological complications following percutaneous vertebroplasty with polymethylmethacrylate - A case report [J].
Harrington, KD .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A (07) :1070-1073
[7]  
Jang Jee Soo, 2002, Spine (Phila Pa 1976), V27, pE416, DOI 10.1097/00007632-200210010-00021
[8]   PEDICLE SCREWS AND BONE-MINERAL DENSITY [J].
KUMANO, K ;
HIRABAYASHI, S ;
OGAWA, Y ;
AOTA, Y .
SPINE, 1994, 19 (10) :1157-1161
[9]  
Lee Bong-Jin, 2002, Spine (Phila Pa 1976), V27, pE419, DOI 10.1097/00007632-200210010-00022
[10]   Percutaneous Treatment of Vertebral Compression Fractures A Meta-analysis of Complications [J].
Lee, Michael J. ;
Dumonski, Mark ;
Cahill, Patrick ;
Stanley, Tom ;
Park, Daniel ;
Singh, Kern .
SPINE, 2009, 34 (11) :1228-1232