Right pneumothorax resulting from an endocardial screw-in atrial lead

被引:26
作者
Ho, WJ [1 ]
Kuo, CT [1 ]
Lin, KH [1 ]
机构
[1] Chang Gung Univ, Chang Gung Mem Hosp, Dept Cardiol, Taipei, Taiwan
关键词
atrial perforation; permanent pacemaker; pneumothorax; screw-in atrial lead;
D O I
10.1378/chest.116.4.1133
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Right pneumothorax complicated by an endocardial atrial lead has never been reported. Herein, we report on a small-build 79-year-old Taiwanese woman who suffered from complete AV block and underwent dual-chamber permanent pacemaker implantation. An active fixation screw-in atrial lead was chosen. The procedure was complicated by right pneumothorax associated with atrial perforation. Since simple measurements of the implantation parameters could not be used to predict the occurrence of perforation, great caution should be taken in to avoid overscrewing the atrial lead, and in scrutinizing the penetration depth of the helix of the lead under fluoroscopy.
引用
收藏
页码:1133 / 1134
页数:2
相关论文
共 6 条
[1]   CLINICAL SURVEILLANCE OF AN ACTIVE FIXATION, BIPOLAR, POLYURETHANE INSULATED PACING LEAD .1. THE ATRIAL LEAD [J].
GLIKSON, M ;
VONFELDT, LK ;
SUMAN, VJ ;
HAYES, DL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (08) :1399-1404
[2]   ACUTE PERICARDITIS RESULTING FROM AN ENDOCARDIAL ACTIVE FIXATION SCREW-IN ATRIAL LEAD [J].
GREENE, TO ;
PORTNOW, AS ;
HUANG, SKS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (01) :21-25
[3]   RELIABILITY OF ATRIAL SCREW-IN LEADS [J].
MARKEWITZ, A ;
WENKE, K ;
WEINHOLD, C .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1777-1783
[4]  
NOOTEN GV, 1990, AM HEART J, V119, P1439
[5]  
PORSONNET V, 1995, PACE, V18, P955
[6]   RIGHT ATRIAL PERFORATION BY A TEMPORARY EPICARDIAL PACING WIRE [J].
SMITH, JA ;
TATOULIS, J .
ANNALS OF THORACIC SURGERY, 1990, 50 (01) :141-142