A solid iodinated phantom material for use in tomographic x-ray imaging

被引:14
作者
Hill, Melissa L. [1 ,2 ]
Mainprize, James G. [1 ]
Mawdsley, Gordon E. [1 ]
Yaffe, Martin J. [1 ,2 ]
机构
[1] Sunnybrook Hlth Sci Ctr, Toronto, ON M4N 3M5, Canada
[2] Univ Toronto, Dept Med Biophys, Toronto, ON M4N 3M5, Canada
关键词
phantoms; iodine; quantitative imaging; dual energy; computed tomography; quality control; INITIAL CLINICAL-EXPERIENCE; DIGITAL MAMMOGRAPHY SYSTEM; CONTRAST-ENHANCED CT; COMPUTED-TOMOGRAPHY; PHYSICAL-CHARACTERISTICS; CDZNTE DETECTOR; PATIENT SIZE; ANGIOGRAPHY; SPECTRA; PERFORMANCE;
D O I
10.1118/1.3213516
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: Iodinated phantoms are of value in x-ray imaging for quality control measurements, system calibration, and for use in the research setting; however, the liquid phantoms that are most often used have several limitations including variability between repeated dilutions, inhomogeneities from air bubbles or precipitants, and long set up times. Although suitable materials have been investigated for projection radiography, quantitative measurements of contrast enhancement in computed tomography (CT) have become increasingly important in the clinic, and a need exists for a durable and reproducible iodinated phantom material. In this work, the authors describe a solid radiographic phantom material that has an accurately known concentration of iodine distributed uniformly throughout its volume and that has stable properties over time. This material can be molded or machined into a desired shape to create a test object or for use in an anthropomorphic phantom. METHODS: Two sets of calibration phantoms were produced with a clinically relevant range of iodine concentrations. Measurements were made on these phantoms to characterize the material properties in terms of accuracy of iodine concentration, radiographic uniformity, temporal stability of x-ray attenuation, and manufacturing repeatability. Experimentally measured linear x-ray attenuation coefficients were compared to those predicted by a theoretical model. The uniformity of the iodine distribution in the material was assessed by measuring image intensity variation, both in projection images and in reconstructed CT volumes. The reproducibility of manufacture was estimated on a set of independently produced samples. A longitudinal study was performed to assess the stability of the material x-ray characteristics over time by making measurements at 6 month intervals. RESULTS: Good agreement was seen between the experimental measurements of effective attenuation and the theoretically predicted values. It is estimated that a desired iodine concentration could be produced to within 0.04 mg/ml. Comparison of the measured effective linear iodine attenuation coefficients of eight 1.0 mg/ml samples indicated a manufacturing reproducibility of +/- 0.03 mg/ml iodine. Variations in uniformity across each of the samples were on the order of image intensity fluctuations (sigma). No inhomogeneities due to mixing or settling were apparent. An analysis of longitudinal data collected for both calibration sets revealed no perceptible change in radiographic properties over the first 6 months after manufacture, nor over a subsequent 1.5 yr period from 1 yr postmanufacture onward. CONCLUSIONS: The uniformity, stability, and precision of this iodinated material suggest that it is suitable for use in accurate calibration tools for contrast tomographic imaging. (C) 2009 American Association of Physicists in Medicine. [DOI: 10.1118/1.3213516]
引用
收藏
页码:4409 / 4420
页数:12
相关论文
共 49 条
[1]  
Allert KD, 2007, J X-RAY SCI TECHNOL, V15, P9
[2]  
ARNOLD BA, 1987, Patent No. 4649561
[3]  
ARNOLD BA, 1988, Patent No. 4724110
[4]   A new tool for benchmarking cardiovascular fluoroscopes [J].
Balter, S .
RADIATION PROTECTION DOSIMETRY, 2001, 94 (1-2) :161-166
[5]   Dynamic contrast-enhanced CT of head and neck tumors: perfusion measurements using a distributed-parameter tracer kinetic model. Initial results and comparison with deconvolution-based analysis [J].
Bisdas, Sotirios ;
Konstantinou, George N. ;
Lee, Puor Sherng ;
Thng, Choon Hua ;
Wagenblast, Jens ;
Baghi, Mehran ;
Koh, Tong San .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (20) :6181-6196
[6]   Lag and ghosting in a clinical flat-panel selenium digital mammography system [J].
Bloomquist, Aili K. ;
Yaffe, Martin J. ;
Mawdsley, Gordon E. ;
Hunter, David M. ;
Beideck, Daniel J. .
MEDICAL PHYSICS, 2006, 33 (08) :2998-3005
[7]   Radiological interpretation 2020: Toward quantitative image assessment [J].
Boone, John M. .
MEDICAL PHYSICS, 2007, 34 (11) :4173-4179
[8]  
Cosgrove DO, 1999, TXB CONTRAST MEDIA
[9]  
*EP RES, 2004, R3500 H5064 PROD DAT
[10]   DYNAMIC AND STATIC PHANTOMS FOR EVALUATION OF DIGITAL SUBTRACTION ANGIOGRAPHIC SYSTEMS [J].
FUJITA, H ;
DOI, K ;
CHAN, HP ;
GIGER, ML ;
DUDA, EE .
RADIOLOGY, 1985, 155 (03) :799-803